  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Uppsala",
      "state": "N/A",
      "country": "Sweden",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72969",
      "recalling_firm": "Phadia Ab",
      "address_1": "Rapsgatan 7",
      "address_2": "Rapsgatan Plant",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide in the states of NC, GA, NJ, NY, PA, TX, IL, MI, MT, NM, and OR.",
      "recall_number": "Z-0697-2016",
      "product_description": "EliA SmDP Well, Article Number 14-5624-01    Product Usage:  EliA SmDP is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Sm in human serum and plasma (EDTA, citrate) as an aid in the clinical diagnosis of systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. EliA SmDP uses the EliA IgG method on the instruments Phadia 100 and Phadia 250.",
      "product_quantity": "76",
      "reason_for_recall": "All equivocal and positive results (greater or equal 7 EliA U/ml) on EliA SmDP well lot 0018 may be incorrect and must be considered invalid. Patient samples in complaint investigations caused unspecific signals up to 22 U/ml on EliA SmDP well lot 0018. The nonspecific signals are not caused by anti-Sm antibodies, nor by streptavidin antibodies.",
      "recall_initiation_date": "20151228",
      "center_classification_date": "20160127",
      "termination_date": "20160816",
      "report_date": "20160203",
      "code_info": "Product number: 14-5624-01  Lot: 0018  Exp date: 11/30/2016"
    }
  ]
}