      "country": "United States",
      "city": "Memphis",
      "address_1": "1450 E Brooks Rd",
      "reason_for_recall": "There is a a potential failure mode associated with the use of the mallet, wherein the poly tip of the mallet may become loose and/or come off during use. If the potential failure mode occurs, particles generated in the poly thread potentially could enter the surgical field or a delay in surgery could occur if the tip disengages during the use of the mallet.",
      "address_2": "",
      "product_quantity": "1371 devices",
      "code_info": "Lot Numbers:  17AM17766 17AM17767 17CM01798 17CM01799 17CM01800   17CM01801 17CM01802 17CM01803 17CM01804 17CM01805 17CM01806   17CM01807 17CM02636 17CM02637 17CM02638 17CM02639 17CM02640 17CM02641 17CM02641A 17CM02642 17CM02643 17CM02644 17CM02645   17CM02646 17CM02647 17CM02648 17CM05047 17CM05048 17CM05049   17CM05050 17CM05050A 17CM05051 17CM05052 17CM05053 17CM05054   17CM05055 17CM05056 17CM05057 17CM05058 17CM05059 17CM07985   17CM07986 17CM07987 17CM07988 17CM10164 17CM10166 17CM10168   17CM15488 17CM15490 17CM15492 17CM15493 17CM15494 17CM15496   17CM15499 17CM19463 17CM19464 17CM19466 17CM19468 17CM19469   17CM19470 17CM19471 17CM19473 17DM03540 17DM03542 17DM03544   17DM03545 17DM03547 17DM03548 17DM03549 17DM03550 17DM03551   17DM03552 17DM03553 17DM10064 17DM10066 17DM10069 17DM10072   17DM14585 17DM14586 17DM14587 17DM14588 17DM14589 17DM14590   17EM00408 17EM00409 17EM00410 17EM00412 17EM00413 17EM00414   17EM00415 17EM00416 17EM00417 17EM00419 17EM00420 17EM00421   17EM00422 17EM00423 17EM00424 17EM05703 17EM05704 17EM05705   17EM05706 17EM05712 17EM10828 17EM10828A 17EM10829 17EM10830   17EM15105 17EM15106 17EM15107 17EM15108 17EM15109 17EM15110   17EM21229 17EM21230 17EM21233 17EM21236 17EM21238 17EM21239   17EM21242 17EM21245 17FM09954 17FM09955 17FM09956 17FM09957   17FM09958 17FM09959 17FM09959A 17GM07654 17GM07655 17GM07656   17GM07657 17GM07658 17GM07659 17GM07660 17GM07661 17GM07662   17GM07663 17GM07664 17GM07665 17GM07677 17GM07678 17GM07679   17GM07680 17GM07681 17GM07682 17GM07683 17GM07684 17GM07685   17GM07686 17GM07687 17GM07688 17GM10245 17GM12933 17GM12934   17GM12935 17GM15687 17GM15688 17GM15689 17GM15690 17JM01118   17JM04860P 17JM09449 17JM10724 17JM11956 17JM11957 17KM04646   17KM08204 17KM08207 17LM10125 17LM11933 17LM14786 17LM15874   17LM15875 17MM05227 17MM05228 17MM10461 18KK00010 18KK00070   18LK00128",
      "center_classification_date": "20191217",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution and countries of Belgium, Canada, China, Columbia, Costa Rica, Finland, France, Germany, Great Britain, Hong Kong, Italy, Netherlands, Norway, Panama, Poland, Switzerland, United Arab Emirates, Venezuela.",
      "state": "TN",
      "product_description": "smith&nephew PROFIX Mallet, 750 GRAMS, REF 71512452 - Product Usage: The PROFIX Mallet is a reusable surgical instrument used to tap instruments or implants during knee or hip surgery.",
      "report_date": "20191225",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Smith & Nephew, Inc.",
      "recall_number": "Z-0696-2020",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "84325",
      "termination_date": "20210507",
      "more_code_info": "",
      "recall_initiation_date": "20191111",
      "postal_code": "38116-1804",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}