  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Raynham",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79117",
      "recalling_firm": "DePuy Mitek, Inc., a Johnson & Johnson Co.",
      "address_1": "325 Paramount Dr",
      "address_2": "N/A",
      "postal_code": "02767-5199",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-0695-2018",
      "product_description": "1113 Gravity Tube Set Pk/24 -SH    Product Usage:  The FMS Gravity Tube Set is intended to deliver irrigation fluid from irrigation containers to the operative site during arthroscopic procedures.",
      "product_quantity": "15,678",
      "reason_for_recall": "Leakage of FMS gravity tub sets (281113) at the bulb and tube connection",
      "recall_initiation_date": "20170831",
      "center_classification_date": "20180223",
      "termination_date": "20190409",
      "report_date": "20180307",
      "code_info": "Lot codes: 138, 194, 633, 1067, 1741, 2576, 2608, and 2645"
    }
  ]
}