  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Somerville",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79086",
      "recalling_firm": "Ethicon, Inc.",
      "address_1": "Us Highway 22 West",
      "address_2": "N/A",
      "postal_code": "08876",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationally",
      "recall_number": "Z-0693-2018",
      "product_description": "Coated VICRYL RAPIDETM (Polyglactin 910) Suture",
      "product_quantity": "129,372 Eaches (Internationally)",
      "reason_for_recall": "On August 17, 2017, Ethicon had decided to initiate a medical device recall  (removal) of the twenty-five (25) lots of VICRYL RAPIDETM (Polyglactin 910) Suture  that were distributed to the field. During finished goods testing, a suture damage was  identified on a lot of VICRYL RAPIDETM Suture. During the course of the subsequent  investigation, the suture also failed knot tensile strength testing.",
      "recall_initiation_date": "20170817",
      "center_classification_date": "20180223",
      "termination_date": "20190508",
      "report_date": "20180307",
      "code_info": "Coated VICRYL RAPIDETM  (Polyglactin 910) Suture MPVR4260H Size 4-0 VICRYL RAPIDE  Undyed 75cm PS-2 Needle 10705031149168  Coated VICRYL RAPIDETM  (Polyglactin 910) Suture V2140H Size 4-0 VICRYL RAPIDE  Undyed 70cm RB-1 Needle 10705031106048  Coated VICRYL RAPIDETM  (Polyglactin 910) Suture VR2217 Size 2-0 VICRYL RAPIDE  Undyed 90cm V-34 Needle 10705031119925  Coated VICRYL RAPIDETM  (Polyglactin 910) Suture VR2253 Size 2-0 VICRYL RAPIDE  Undyed 75cm FS-1 Needle 10705031119956  Coated VICRYL RAPIDETM  (Polyglactin 910) Suture VR2279 Size 4-0 VICRYL RAPIDE  Undyed 75cm V-4 Needle 10705031086524  Coated VICRYL RAPIDETM  (Polyglactin 910) Suture VR2294 Size 4-0 VICRYL RAPIDE  Undyed 75cm FS-3 Needle 10705031086548  Coated VICRYL RAPIDETM  (Polyglactin 910) Suture VR2297 Size 4-0 VICRYL RAPIDE  Undyed 75cm FS-2 Needle 10705031119970  Coated VICRYL RAPIDETM  (Polyglactin 910) Suture VR496 Size 4-0 VICRYL RAPIDE  Undyed 18'' PS-2 Needle 10705031134690  Coated VICRYL RAPIDETM  (Polyglactin 910) Suture VR945 Size 2-0 VICRYL RAPIDE  Undyed 36'' CT-1 Needle 10705031133358  Coated VICRYL RAPIDETM  (Polyglactin 910) Suture VRI253H Size 2-0 VICRYL RAPIDE  Undyed 75cm FS-1 Needle 10705031111172  Coated VICRYL RAPIDETM  (Polyglactin 910) Suture W9941 Size 2-0 VICRYL RAPIDE  Undyed 75cm FS Needle 10705031130999  Coated VICRYL RAPIDETM  (Polyglactin 910) Suture VR426 Size 4-0 VICRYL RAPIDE  Undyed 75cm PS-2 Needle 10705031134669  Coated VICRYL RAPIDETM  (Polyglactin 910) Suture VR494 Size 4-0 VICRYL RAPIDE  Undyed 45cm P-3 Needle 10705031134683  Coated VICRYL RAPIDETM  (Polyglactin 910) Suture VR845 Size 4-0 VICRYL RAPIDE  Undyed 45cm PC-3 Needle 10705031134744  Coated VICRYL RAPIDETM  (Polyglactin 910) Suture VR917 Size 2-0 VICRYL RAPIDE  Undyed 90cm V-34 Needle 10705031134768  Coated VICRYL RAPIDETM  (Polyglactin 910) Suture VR2286 Size 2-0 VICRYL RAPIDE  Undyed 90cm FSL Needle 10705031080256  Coated VICRYL RAPIDETM  (Polyglactin 910) Suture VR2299 Size 2-0 VICRYL RAPIDE  Undyed 75cm FS-2 Needle 10705031080294  Coated VICRYL RAPIDETM  (Polyglactin 910) Suture W9918 Size 4-0 VICRYL RAPIDE  Undyed 75cm PC-3 Needle 10705031151970  Coated VICRYL RAPIDETM  (Polyglactin 910) Suture W9930 Size 4-0 VICRYL RAPIDE  Undyed 75cm FS-2 Needle 10705031152007  Coated VICRYL RAPIDETM  (Polyglactin 910) Suture FH1642H Size 4-0 VICRYL RAPIDE  Undyed 45cm FS-2 Needle 10705031099968  Coated VICRYL RAPIDETM  (Polyglactin 910) Suture VRI214H Size 4-0 VICRYL RAPIDE  Undyed 75cm RB-1 Plus Needle 10705031111127  Coated VICRYL RAPIDETM  (Polyglactin 910) Suture VRI312H Size 2-0 VICRYL RAPIDE  Undyed 75cm SH-1 Plus Needle 10705031111264  Coated VICRYL RAPIDETM  (Polyglactin 910) Suture VRI417H Size 2-0 VICRY RAPIDE  Undyed 75cm SH Plus Needle 10705031111332  Coated VICRYL RAPIDETM  (Polyglactin 910) Suture W9922 Size 4-0 VICRYL RAPIDE  Undyed 75cm PS-2 Needle 10705031130845  Coated VICRYL RAPIDETM  (Polyglactin 910) Suture W9933 Size 2-0 VICRYL RAPIDE  Undyed 75cm PS Needle 10705031130937"
    }
  ]
}