  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Brea",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89178",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "250 S Kraemer Blvd",
      "address_2": "",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "US: Alabama,  Alaska,  Arizona,  Arkansas,  California,  Colorado,  Connecticut,  Delaware,  District of Columbia,  Florida,  Georgia,  Hawaii,  Idaho,  Illinois,  Indiana,  Iowa,  Kansas,  Kentucky,  Louisiana,  Maine,  Maryland,  Massachusetts,  Michigan,  Minnesota,  Mississippi,  Missouri,  Montana,  Nebraska,  Nevada,  New Hampshire,  New Jersey,  New Mexico,  New York,  North Carolina,  North Dakota,  Ohio,  Oklahoma,  Oregon,  Pennsylvania,  Rhode Island,  South Carolina,  South Dakota,  Tennessee,  Texas,  Utah,  Virginia,  Washington,  West Virginia,  Wisconsin,  Wyoming; and      OUS: Albania, Algeria, American Samoa, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Canada, China, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Ethiopia, Finland ,France, Georgia, Germany, Ghana, Gibraltar, Greece, Guam, Guatemala, Guyana, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Lesotho, Libya, Lithuania, Macau, Macedonia, Malawi, Malaysia, Mayotte, Mexico, Morocco, Mozambique, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Niger, Nigeria, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Saint Pierre and Miquelon, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Swaziland, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, United States, Uruguay, Vietnam, Yemen, Zambia, Zimbabwe.",
      "recall_number": "Z-0692-2022",
      "product_description": "AU/DxC AU Magnesium, Catalog/REF Number OSR6189, UDI 15099590010850",
      "product_quantity": "181,776 Kits",
      "reason_for_recall": "Lipemic interference for the Magnesium serum application failed to meet the performance claim as defined in the IFU. Use may cause a maximum positive bias up to 30.38% in low magnesium patient samples. The impact is only to the clinical interpretation of magnesium results in the presence of lipemia.",
      "recall_initiation_date": "20211208",
      "center_classification_date": "20220228",
      "termination_date": "20250128",
      "report_date": "20220309",
      "code_info": "All lot numbers.",
      "more_code_info": ""
    }
  ]
}