  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Kontich",
      "state": "",
      "country": "Belgium",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81780",
      "recalling_firm": "Agfa-Gevaert, N.V.",
      "address_1": "Prins Boudewijnlaan",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide Distribution in the states:  IN, KS, MA, SC, and VA.",
      "recall_number": "Z-0692-2019",
      "product_description": "DX-D 600 DIRECT RADIOGRAPHY SYSTEM    Product Usage:  The DX-D 600 system is a ceiling mounted General Radiography X-Ray imaging system used in hospitals, clinics and medical practices by physicists, radiographers and radiologists to make, process and view static X-Ray radiographic images of the skeleton (including skull, spinal column and extremities), chest, abdomen and other body parts on adult, pediatric or neonatal patients.",
      "product_quantity": "10",
      "reason_for_recall": "After an upgrade of the software of the Overhead Tube Crane, there were isolated cases in which the Overhead Tube Crane movement does not stop when the movement button is released. Instead of stopping while the movement button is released, the Tube Head Crane  moves to the intended position.",
      "recall_initiation_date": "20181204",
      "center_classification_date": "20190107",
      "termination_date": "20200424",
      "report_date": "20190116",
      "code_info": "Serial Numbers: A5430000015  A5430000020  A5430000030  A5430000042  A5430000052  A5430000061  A5430000066  A5430000067  A5430000079  A5430000081",