  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "75594",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution.",
      "recall_number": "Z-0692-2017",
      "product_description": "SOMATOM Force, System x-ray, tomography, computed",
      "product_quantity": "48 units distributed in U.S.",
      "reason_for_recall": "Siemens is providing software update version  VA50A_SP3 to address the software bugs that  were identified through normal field monitoring  and the Global Complaint Handling Process.  Correction for the problems are as follows:  1. Correction to volumetric  misrepresentations of high contrast  objects when using ADMIRE.  2. Correction to highly sporadic scan  aborts due to temporarily tube  currents at 0mA.  3. Correction to missing Microsoft  Hotfixes (MS16-001, MS15-135,  MS15-088, MS15-048).  4. Correction to highly sporadic image  artifacts caused by error signals from  high voltage chain at extremely high  tube currents.  5. Correction to highly sporadic  incomplete image reconstruction due  to temporary slow hard disk write  speed.  6. Correction to highly sporadic scan  aborts due to temporarily high disk  load of system disk.  7. Correction to highly sporadic image  artifacts generated by scatter  correction in case of tube arcings.",
      "recall_initiation_date": "20161020",
      "center_classification_date": "20161202",
      "termination_date": "20170530",
      "report_date": "20161214",
      "code_info": "Model Number 10742326, Serial Numbers: 75708,75709,75607,75617,75550,75526  75512,75567,75536,75711,75514,75600,75563,75510,75454,75540,75571,75500  75527,75475,75535,75481,75482,75710,75460,75476,75588,75493,75458,75528  75467,75602,75487,75464,75478,75604,75513,75532,75555,75524,75515,75559  75610,75501,75570,75616,75576,75605"
    }
  ]
}