  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Telford",
      "state": "PA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72914",
      "recalling_firm": "Draeger Medical, Inc.",
      "address_1": "3135 Quarry Rd",
      "address_2": "N/A",
      "postal_code": "18969-1042",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution -- including states of AR, AZ, CA, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY.",
      "recall_number": "Z-0692-2016",
      "product_description": "Emergency Transport Ventilator  Oxylog 3000 Ventilator and Oxylog 3000 plus Ventilator  Catalog Numbers: 2M86300/2M86965 and 5704813/5704831.    Volume controlled and pressure-controlled emergency and transport ventilators.",
      "product_quantity": "402 units (314 - Oxylog 3000; 88 - Oxylog 3000 Plus)",
      "reason_for_recall": "Drager became aware of situations where the error message \"Poti unplugged\" was generated causing acoustic and visual alarms, the breathing system releases pressure and the ventilation function stops operating.",
      "recall_initiation_date": "20151222",
      "center_classification_date": "20160203",
      "termination_date": "20170405",
      "report_date": "20160210",
      "code_info": "All Oxylog 3000 Ventilators and Oxylog 3000 Plus Ventilators with the identified catalog numbers distributed between April 2007 - December 2015.",
      "more_code_info": ""
    }
  ]
}