  "results": [
    {
      "country": "United States",
      "city": "Arlington",
      "address_1": "5677 Airline Rd",
      "reason_for_recall": "Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing overall trend from 2009 to present and it was found that there was a specific hazard/harm for suspected tissue reaction to metal debris over one percent for modular head CONSERVE(R) systems and DYNASTY(R)  metal acetabular liner systems.",
      "address_2": "",
      "product_quantity": "68,221 units",
      "code_info": "REF 38BF3642  Serial Numbers:  1380112, 1399025, 1402166, 1340677, 1427276, 1429743      REF 38BF3642SP  Serial Number 1416590    REF 38BF3844  Serial Numbers:  1368260, 1402167, 1399026, 1380113, 1418795     REF 38BF3844SP  Serial Number 1416814    REF 38BF4046  Serial Numbers:  1368261, 1380114, 1402168, 1399027, 1418796, 1429744, 1449113     REF 38BF4046SP  Serial Number 1416950    REF 38BF4248  Serial Numbers: 1397139, 1368262, 1381417, 1399029, 1422026, 1418797, 1433514, 1449114, 1452995     REF 38BF4248SP  Serial Number 1416816    REF 38BF4450  Serial Numbers:  1380115, 1399031, 1399231, 1373711, 1366654, 1418948, 1433515, 1431187, 1449115, 1482461, 1484998    REF 38BF4450SP  Serial number 1416563    REF 38BF4652  Serial Numbers:  1380116, 1395709, 1368263, 1373712, 1399233, 1427277, 1429745, 1433516, 1445264, 1449116, 1460186, 1460187, 1482463, 1481980     REF 38BF4652SP  Serial Numbers:  1417403, 1417616    REF 38BF4854  Serial Numbers:  1381418, 1399234, 1373904, 1368264, 1399235, 1427278, 1431188, 1433517, 1445265, 1460189, 1460188, 1468891, 1481979, 1484999     REF 38BF5056  Serial Numbers:  1381419, 1399236, 1373905, 1368265, 1397140, 1402169, 1413708, 1427279, 1445266, 1449117, 1452997, 1468892, 1468893, 1476014, 1476013, 1482462, 1571773, 1571772, 1650637, 1650638, 1651368     REF 38BF5258  Serial Numbers:  1397141, 1399237, 1368266, 1380117, 1422025, 1431189, 1445267, 1449118, 1452996, 1468894, 1476016, 1482460, 1481456, 1571771, 1571770, 1589362, 1650639     REF 38BF5258SP  Serial numbers:  1416945    REF 38BF5460  Serial Numbers:  1368267, 1380118, 1397142, 1399238, 1402170, 1429746, 1433518, 14335181454410, 14335181454409, 1452478, 1476017, 1481978, 1484997, 1651370      REF 38BF5460SP  Serial Number 1416944    REF 38BF5662  Serial Numbers:  1368268, 1381420, 1399239, 1422021, 1431190, 1452994, 1484996     REF 38BF5662SP  Serial Number 1416957    REF 38BF5864  Serial Numbers:  1368269, 1402171, 1380124, 1399240     REF 38BF5864SP  Serial Number 1416589    REF 38BF6066  Serial Numbers:  1399241, 1402172, 1380125, 1427280     REF 38BF6066SP  Serial Number 1416815    REF 38QH3644  Serial Numbers:  098450454, 098450453    REF 38QH3846  Serial Numbers:  087448320, 087448319, 028455786, 108636368, 039752495, 089843391, 1283757     REF 38QH4048  Serial Numbers: 087448317, 087448318, 078573304, 108636369, 039753686     REF 38QH4250  Serial Numbers:  087447402, 087447838, 038474921, 088573305, 019741984, 039783429     REF 38QH4452  Serial Numbers:  087446899, 087447160, 058552436, 029746761, 069832435     REF 38QH4654  Serial Numbers:  036310979, 036312725, 066322950, 087445485, 087445891, 058533122, 088584084, 029751144, 059832434, 099864589,109864590     REF 38QH4856  Serial Numbers:  036308619, 036312726, 066322951, 087443056, 087444883, 038510025, 058552900, 128691657, 039763773     REF 38QH5058  Serial Numbers:  036303438, 036310069, 066322952, 087424788, 087444884, 038515854, 088584085, 029746762, 059832436, 059832438, 1322641     REF 38QH5260  Serial Numbers:  036310978, 036312727, 087445887, 087446423, 088552437, 128688680, 059831040     REF 38QH5462  Serial Numbers:  036311691, 066322954, 087445888, 087446426, 028469801, 058552901, 128691658, 029763769, 059838299     REF 38QH5664  Serial Numbers:  087446896, 087447161, 058544859, 118691659, 029751973, 059829624     REF 38QH5666  Serial Numbers:  087447403, 087447839, 058544860, 128691660, 059830306     REF 38SC3650  Serial Numbers:    025149290, 025160237, 025174327, 075208859, 085242413, 085246262, 105249039, 016300068, 026305920, 096368471, 126385423, 027393904, 087458988, 068573306, 108602432, 029744514     REF 38SC3852  Serial Numbers:  025149144, 025170078, 025160238, 035179216, 075232801, 075242414, 085246264, 105267353, 115277853, 026301520, 026317024, 036319642, 056329110, 106333747, 027385922, 107460700, 107465345, 118703700     REF 38SC4054  Serial Numbers:  025149111, 025160239, 035179223, 075236081, 085242415, 085246267, 115276025, 016277854, 036318111, 036318361, 066320283, 097438943, 127465344, 108566756, 029753065     REF 38SC4256  Serial Numbers:  025149838, 025146436, 025179217, 035188989, 075233865, 085242416, 095257911, 105268093, 105269854, 125277855, 125283868, 036318819, 066320284, 106368472, 097460701, 097480974, 038466319, 019746760     REF 38SC4458  Serial Numbers:  025144180, 025146460, 035179222, 035188990, 075233866, 085242417, 085249038, 105269855, 115276026, 115283869, 026305908, 056320285, 106336785, 027385923, 077437212,   107460702, 107461472, 029566757      REF 38SC4660  Serial Numbers:  025142838, 025146459, 025174328, 035188992, 035179221, 075233867, 085242418, 095262956, 105269857, 016283870, 036319135, 046320286, 066341722, 106368473, 116371016, 037389856, 037387396, 077432335, 107465343, 108492001, 019745503      REF 38SC4862  Serial Numbers:  025144201, 025140464, 025190407, 025170075, 025170076, 075233868, 085242419, 105262954, 105269861, 016283871, 036319136, 066320287, 087459243, 108605237, 029745502     REF 38SC5064  Serial Numbers:  025128517, 025150758, 025160240, 025170072, 025170074, 035188993, 035190419, 085242420, 085248657, 105263392, 105269867, 016283872, 066320288, 077438092, 068493738, 108605238     REF 38SC5266  Serial numbers:  025190387, 025128518, 025160241, 035179218, 035188994, 035192254, 085242421, 085248658, 105269869, 036319137, 027389086, 087438091, 107465342     REF 38SC5468  Serial Numbers:  025190390, 025128519, 025160242, 025174329, 025187586, 035190750, 075188995, 075192255, 085242394, 095248659, 105274072, 106331324, 077436606, 018495671     REF 38SC5670  Serial Numbers:  025179948, 025190082, 025179949, 025128520, 025155250, 025160243, 035179219, 085242422, 085248660, 105269872, 116331326, 107460703     REF 38SC5672  Serial Numbers:  025190391, 025190397, 025190398, 025155817, 025158548, 025160244, 035188996, 035179220, 075204364, 085242423, 095248661, 026302015, 027394748, 107460704, 028486699",
      "center_classification_date": "20161201",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Columbia, Cyprus, Czech Republic, Denmark, Finland, France, Georgia, Germany, Greece, Iran, Italy, Jamaica, Japan, Lithuania, Malaysia, Philippines, Poland, Portugal, Russia, Slovenia, South Africa, South Korea, Spain, Switzerland, Taiwan, Netherlands, Turkey, Arab Emirates, United Kingdom, and Vietnam.",
      "state": "TN",
      "product_description": "Conserve Cups:  REF 38BF3642, REF 38BF3642SP, REF 38BF3844, REF 38BF3844SP, REF 38BF4046, REF 38BF4046SP, REF 38BF4248, REF 38BF4248SP, REF 38BF4450, REF 38BF4450SP,   REF 38BF4652, REF 38BF4652SP, REF 38BF4854, REF 38BF5056, REF 38BF5258, REF 38BF5258SP, REF 38BF5460, REF 38BF5460SP, REF 38BF5662, REF 38BF5662SP, REF 38BF5864,   REF 38BF5864SP, REF 38BF6066, REF 38BF6066SP, REF 38QH3644, REF 38QH3846, REF 38QH4048, REF 38QH4250, REF 38QH4452, REF 38QH4654, REF 38QH4856, REF 38QH5058,   REF 38QH5260, REF 38QH5462, REF 38QH5664, REF 38QH5666, REF 38SC3650, REF 38SC3852,   REF 38SC4054, REF 38SC4256, REF 38SC4458, REF 38SC4660, REF 38SC4862, REF 38SC5064, REF 38SC5266, REF 38SC5468, REF 38SC5670, REF 38SC5672  For use with the conserve metal head",
      "report_date": "20161207",
      "classification": "Class III",
      "openfda": {},
      "recalling_firm": "MicroPort Orthopedics Inc.",
      "recall_number": "Z-0688-2017",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "75063",
      "termination_date": "20170701",
      "more_code_info": "",
      "recall_initiation_date": "20160621",
      "postal_code": "38002-9501",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}