  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Atlanta",
      "state": "GA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81705",
      "recalling_firm": "Elekta, Inc.",
      "address_1": "400 Perimeter Center Ter NE Ste 50",
      "address_2": "",
      "postal_code": "30346-1227",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distributed in the United States and outside of the Unites States: Algeria; Antigua and Barbuda; Argentina; Australia; Belgium; Botswana; Brazil; Canada; China; Colombia; Denmark; France; Germany; India; Indonesia; Iran; Ireland; Israel; Italy; Japan; Latvia; Lebanon; Malaysia; Mexico; Morocco; Myanmar; Netherlands; New Zealand; Norway; Peru; Philippines; Poland; Portugal; Russian Federation; Rwanda; Saudi Arabia; Senegal; Singapore; Slovenia; South Africa; Spain; Sweden; Switzerland; Taiwan; Thailand; Turkey; United Kingdom; Viet Nam",
      "recall_number": "Z-0686-2019",
      "product_description": "MOSAIQ Oncology Information System Version 2.64.      MOSAIQ is an oncology information system used to manage workflows for treatment planning and delivery",
      "product_quantity": "551 units",
      "reason_for_recall": "If a Setup field (CT, kV, MV or MVCT) is created by COPYING a treatment field or another setup field, then the  copied field will have the same isocenter as the parent and that isocenter may not be the correct isocenter for  imaging when it is sent to the machine.",
      "recall_initiation_date": "20181204",
      "center_classification_date": "20181227",
      "termination_date": "20260223",
      "report_date": "20190102",
      "code_info": "Software Build(s): 2.64 SP9 and higher;      UDI (01)00858164002091(10)2.64.235",