  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Arlington",
      "address_1": "5677 Airline Rd",
      "reason_for_recall": "Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing overall trend from 2009 to present and it was found that there was a specific hazard/harm for suspected tissue reaction to metal debris over one percent for modular head CONSERVE(R) systems and DYNASTY(R)  metal acetabular liner systems.",
      "address_2": "",
      "product_quantity": "500 units",
      "code_info": "REF PHA04702  Serial Numbers: PHA04702, 087461467, 089909592, 106368684, 106379196, 106381827, 109949208, 129989153, 0201018636, 0301049040      REF PHA04704  Serial Numbers:  010989155, 027373842, 038461468, 057433172, 077453350, 089924904, 106379197, 106380366, 106381751, 0201018637     REF PHA04706  Serial Numbers:  1414640, 010990534, 017400915, 037415861, 039755749, 047420835, 047424536, 077459275, 087461469, 089906466, 0101034747, 106379879, 106381773, 0301084459, 0401100079, 0501122040, 0801189074     REF PHA04708  Serial Numbers:  010990535, 017400568, 037420836, 047424415, 077455537, 087461470, 089901226, 106380367, 106381752, 0311304891, 0401118999, 0601163086, 0711360334     REF PHA04710  Serial Numbers:  1372615, 1413361, 017400673, 037415862, 037420837, 047423789, 047424535, 077457450, 077460608, 089901146, 098695744, 098695746, 106380368, 106381542, 106381753, 108695745, 109944906, 116380977, 119983782, 129990536, 0201016259, 0301076759, 0311303536, 0311305118, 0401089452, 0401121936, 0501131023, 0801187747, 1291000322     REF PHA04712  Serial Numbers:  029695877, 037414678, 047418090, 057426407, 077457328, 098464129, 099906467, 106373846, 106380571, 116389410, 0201058381, 0301084465, 0311303537, 0401117709, 0601163076,   0801188639     REF PHA04714  Serial Numbers:  1414641, 017400569, 029807399, 037415854, 047423790, 077432883, 077445957, 106368690, 0201025145, 0901196483",
      "center_classification_date": "20161201",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Columbia, Cyprus, Czech Republic, Denmark, Finland, France, Georgia, Germany, Greece, Iran, Italy, Jamaica, Japan, Lithuania, Malaysia, Philippines, Poland, Portugal, Russia, Slovenia, South Africa, South Korea, Spain, Switzerland, Taiwan, Netherlands, Turkey, Arab Emirates, United Kingdom, and Vietnam.",
      "state": "TN",
      "product_description": "Procotyl (Rim- Lock) Liners:  REF PHA04702, REF PHA04704, REF PHA04706, REF PHA04708, REF PHA04710, REF PHA04712, REF PHA04714  For use with the conserve metal head",
      "report_date": "20161207",
      "classification": "Class III",
      "openfda": {},
      "recalling_firm": "MicroPort Orthopedics Inc.",
      "recall_number": "Z-0685-2017",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "75063",
      "termination_date": "20170701",
      "more_code_info": "",
      "recall_initiation_date": "20160621",
      "postal_code": "38002-9501",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}