      "country": "United States",
      "city": "Arlington",
      "address_1": "5677 Airline Rd",
      "reason_for_recall": "Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing overall trend from 2009 to present and it was found that there was a specific hazard/harm for \u001csuspected tissue reaction to metal debris\u001d over one percent for modular head CONSERVE(R) systems and DYNASTY(R)  metal acetabular liner systems.",
      "address_2": "",
      "product_quantity": "6,250 units",
      "code_info": "Serial Numbers:    REF DLCOGA28  Seral Numbers:  1509498, 1509897, 107461555, 117476083, 127477625, 0411341968, 0611341969, 0901199714    REF DLCOGB32  Serial Numbers:  1428321, 027403682, 027403684, 027412214, 027412216, 027412218, 027412220, 027412222, 029789268, 048571448, 058605178, 068625411, 069837383, 078628910, 079859721, 087467577, 097474918, 097476396, 097477626, 098657540, 099905615, 099927666, 0101018635,   107480309, 107486655, 109968495, 0111248666, 0111254637, 117486656, 117501983, 119982392, 127504540, 127508429, 0211298850, 0211298851, 0401072608, 0501123237, 0601150684, 0711370347, 0711370832, 0711371855, 0711388872, 0801184970, 0801185465, 0901196484, 0901199209, 0901200444, 0901200704     REF DLCOGC36  Serial Numbers:  1389336, 1411596, 1412300, 1419309, 1421706, 1421872, 1421873, 1425040, 1425041, 1426635, 1441425, 1447128, 1448009, 017395680, 017396168, 017398843, 019752394, 027406078, 027406080, 027406549, 027408437, 027408506, 027412667, 029797073,038565179,   038571449, 039808007, 039823319, 039830311, 058576611, 059835277, 059843767, 059859139, 059869515, 067444752, 068608657, 069876613, 069885333, 077449136, 078628911, 078648401, 078648402, 086356142, 086364960, 087462096, 089905617, 089920491, 089924903, 097475888,   097475891, 097475892, 099932177, 099937595, 0101009548, 108705038, 109966090, 0111273375, 0111274786, 118728271, 119975127, 127506074, 127508430, 127514978, 129979634, 129990530, 0201044377, 0201044378, 0201048252, 0201071222, 0211301143, 0301068674, 0301084452, 0401085769, 0501125124, 0611343376, 0611343377, 0611365354, 0611366004, 0611366646, 0701164679, 0701165562, 0701178783, 0801190509, 0901189893, 0901191386, 0901200880, 0901201441, 0901201663     REF DLCOGD38  Serial Numbers:  1390011, 1422255, 1422256, 017395681, 017396169, 019751147, 019752153, 027404999, 027405000, 027408507, 027408521, 027411898, 027412100, 029793006, 038565180, 038571452, 039808008, 039823320, 039831045, 048576612, 049839752, 059846062, 059846677,   059868799, 059869517, 067445955, 067447299, 068608658, 069884531, 069888788, 077459273, 078628912, 078648403, 087464200, 087466701, 089901144, 089911489, 089922646, 096362335, 096374885, 097472362, 097474113, 097476397, 099932178, 099933792, 099937596, 0101038419, 0101038420, 107482012, 107486663, 108706223, 108716643, 108720178, 109941938, 109970863, 0111274868, 0111274869, 117486664, 117489826, 117501984, 118731716, 119980968, 127466702, 127504541, 127508431, 127512298, 128748672, 129990531, 0201062908, 0301062907, 0301067280, 0301067281, 0301081677, 0401096840, 0401098166, 0401109395, 0501127076, 0511344345, 0601147758, 0601148281, 0601148932, 0611344346, 0611346090, 0611346455, 0701172549, 0801192624, 0801193584, 0801193987, 0801194817, 0901192045, 0901203167, 0901206932, 1001206425, 1291016257, 1291016258, 1291019953, 1291019954, 08991148910     REF DLCOGE40  Serial Numbers:  1383898, 1422893, 017395849, 017396101, 017397513, 017398320, 019747416, 019749028, 019752393, 027404377, 027404964, 027406550, 027406551, 027407780, 039802355, 039809788, 039826200, 048571453, 048571454, 049834730, 049839160, 067436735, 069859138,   069868803, 069869518, 076357782, 077440680, 077453681, 078595478, 078595479, 078648404, 078648405, 079888789, 086362336, 087460119, 087460601, 089904051, 089920496, 096372317, 097472371, 097476398, 099931104, 099933793, 0101022064, 0101022065, 0101030547, 0101030548, 0101042594, 0101042595, 107482014, 107486667, 107486668, 107489825, 109941939, 109944905, 0111271873, 0111272250, 117489816, 117501985, 118722756, 118731717, 119980969, 127504542,   127508432, 127512297, 129990532, 0201057245, 0201064292, 0211295490, 0211301165, 0211302281, 0301064291, 0301081699, 0301083868, 0301084453, 0301095292, 0401095291, 0401096841, 0601165833, 0601165834, 0611361690, 0611361691, 0611364735, 0701170121, 0701170775, 0701171294, 0801194833, 0901194835, 0901207002, 0901208793, 0901208953      REF DLCOGF42  Serial Numbers:  1199809, 1366660, 1422894, 1423547, 017396170, 017397514, 017397515, 017398844, 027406424, 027406553, 027411519, 027411895, 027411896, 027413499, 029789333, 029789334, 029798203, 029809789, 029814314, 029819342, 037413500, 039819343, 049833519,    049839753, 058608659, 059846676, 059857910, 068595480, 069883241, 077449093, 078634682, 078634683, 078648406, 079860580, 079887540, 079894027, 086370088, 087462468, 089914628,  096372318, 097472373, 097476399, 099904052, 099931103, 0101022066, 0101030549, 0101030550, 107482016, 107486669, 107486670, 108705039, 108716644, 109922647, 109937597, 109943610, 109966091, 0111276618, 117496631, 117501986, 118728903, 118744506, 119972083, 127504543,   127509428, 127514975, 127514976, 128749026, 128751278, 129990533, 0201053787, 0201053788, 0201062909, 0201062910, 0201068675, 0211299901, 0301071223, 0301081700, 0301083869, 0401104095, 0401104096, 0501127409, 0611346456, 0701167385, 0701168088, 0701168230, 0701169430, 0701170051, 0711367085, 0901208997, 0901209918, 0901209977, 1101247355, 1101248443, 010103054910     REF DLCOGG44  Serial Numbers:  1383128, 1383366, 1392621, 1428789, 017396171, 017397398, 027404375, 027404965, 027406552, 027408531, 027412079, 027412080, 027412086, 029792981, 029793005, 029799054, 029799055, 038565181, 039828320, 039832455, 048571456, 048576613, 049828319,   049830312, 049839159, 057438911,058608660, 059855232, 059857911, 059860406, 059869519, 067445525, 069888790, 077447460, 077459517, 078608662, 078628913, 078648407, 078657546, 079894757, 087464126, 089915880, 089922080, 096362338, 096374886, 097476400, 097476401,   097480312, 098694868, 098706224, 099932179, 099934263, 099939751, 0101025139, 0101025144, 0101027410, 0101038421, 0101038422, 0101046891, 107481184, 107486653, 107487346, 108705037, 108716645, 108723278, 109966092, 117495711, 117495712, 117505591, 118730735, 118744505, 119975128, 119975129, 127504544, 127513338, 127518295, 127522391, 128747417, 128748669, 129998770, 129998777, 0201057246, 0201059716, 0201059717, 0201076758, 0211275710, 0301076757, 0301085876, 0301085877, 0301099035, 0301099036, 0311294091, 0311295489, 0401099037, 0401100077, 0401100078, 0401106374, 0401112729, 0411306527, 0411307069, 0601153393, 0601154045, 0601155591, 0801181073, 0801181695, 0801181920, 0801183924,   0801199319, 0901210490, 1001213377, 1001214138, 1101232813, 1201256582, 1201256583, 1201257832, 1291000319, 1291000321, 1291025140     REF DLCOGH48  Serial Numbers:  027403685, 027403686, 027410682, 027412223, 027412224, 027412225, 027412226, 078608663, 078648408, 078648409, 089904664, 097471988, 097476402, 097476542, 097477627, 0101042593, 107481183, 109967195, 0111277921, 117486671, 117496632, 127504545, 127508400, 127518297, 0201072609, 0401089450, 0601147117, 0901187425, 1101231165, 1291014716     REF DLCOGJ52  Serial Numbers:  099898795, 0101023111, 0201049895, 0201049896    REF DLCOGK56  Serial Numbers:  099898794, 0201023112, 0201049903, 0201052602, 0201052603",
      "center_classification_date": "20161201",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Columbia, Cyprus, Czech Republic, Denmark, Finland, France, Georgia, Germany, Greece, Iran, Italy, Jamaica, Japan, Lithuania, Malaysia, Philippines, Poland, Portugal, Russia, Slovenia, South Africa, South Korea, Spain, Switzerland, Taiwan, Netherlands, Turkey, Arab Emirates, United Kingdom, and Vietnam.",
      "state": "TN",
      "product_description": "Dynasty Liners:  REF DLCOGA28, REF DLCOGB32, REF DLCOGC36, REF DLCOGD38, REF DLCOGE40, REF DLCOGF42, REF DLCOGG44, REF DLCOGH48, REF DLCOGJ52, REF DLCOGK56,   For use with the conserve metal head",
      "report_date": "20161207",
      "classification": "Class III",
      "openfda": {},
      "recalling_firm": "MicroPort Orthopedics Inc.",
      "recall_number": "Z-0684-2017",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "75063",
      "termination_date": "20170701",
      "more_code_info": "",
      "recall_initiation_date": "20160621",
      "postal_code": "38002-9501",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}