      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72943",
      "recalling_firm": "GE Medical Systems, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (nationwide including PR except NH,  VT, & WY.  Internationally to ALBANIA, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BANGLADESH, BELGIUM, BOLIVIA, BOSNIA, BRAZIL, CAMBODIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, ESTONIA, FINLAND, FRANCE, GEORGIA, GERMANY, CHANA, GIBRALTAR, GREECE, GUATEMALA, HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN KENYA, REPUBLIC OF KOREA, KUWAIT, LATVIA, LITHUANIA,MALAYSIA, MALDIVES, MARTINIQUE, MEXICO, MOROCCO, NETHERLANDS, NEW ZEALAND, NICARAGUA, NORWAY, OMAN, PAKISTAN, PERU, PHILIPPINES, POLAND, QATAR, REUNION, ROMANIA, RUSSIAN FEDERATION SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SRI lanka, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, VENEZUELA, VIETNAM.",
      "recall_number": "Z-0681-2016",
      "product_description": "GE Healthcare, Aisys CS2,1011-9050-000;  Avance Amingo,1009-9002-000, & Avance CS2, 1009-9050-000.  The GE Datex-Ohmeda Aisys Anesthesia System is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal, pediatric, adult). The device is intended for volume or pressure control ventilation. The Aisys is not suitable for use in a MRI environment.",
      "product_quantity": "2627",
      "reason_for_recall": "The Drive Gas Check Valve of certain Anesthesia products could become stuck in a fixed open position, which could cause pressure to build in the mechanical ventilation cycle. If this issue is left unresolved, it could result in excessive or prolonged pressure in the patient breathing circuit during ventilation potentially resulting in barotrauma.",
      "recall_initiation_date": "20151215",
      "center_classification_date": "20160121",
      "termination_date": "20160921",
      "report_date": "20160127",
      "code_info": "APWU00723  APWU00724  APWU01057  APWU01058  APWU01059  APWU01785  APWU01827  APWU01988  APWU01989  APWU01990  APWU01991  APWU01992  APWU01993  APWU01994  APWU01995  APWU01996  APWU01997  APWU01576  APWU01577  APWU01578  APWU00707  APWU00708  APWU00709  APWU00710  APWU00711  APWU00712  APWU00713  APWU01486  APWU01487  APWU01488  APWU01489  APWU01490  APWU01491  APWU01492  APWU01493  APWU01494  APWU01495  APWU01825  APWU01909  APWU02017  APWU02018  APWU02019  APWU02020  APWU01054  APWU01055  APWU01056  APWU00840  APWU00841  APWU00842  APWU00843  APWU00844  APWU00845  APWU00846  APWU00847  APWU01608  APWU01609  APWU01610  APWU01611  APWU01612  APWU01613  APWU01614  APWU01615  APWU02066  APWU02067  APWU00728  APWU00729  APWU00730  APWU00751  APWU00752  APWU00753  APWU00754  APWU00755  APWU01595  APWU01240  APWU01241  APWU01242  APWU01243  APWU01244  APWU01359  APWU01949  APWU01410  APWU01411  APWU01412  APWU00767  APWU00768  APWU00769  APWU01131  APWU01132  APWU01133  APWU01340  APWU01341  APWU01342  APWU01343  APWU01344  APWU01616  APWU01617  APWU01618  APWU01619  APWU01620  APWU01621  APWU01622  APWU01623  APWU01699  APWU01700  APWU01701  APWU01702  APWU01703  APWU01704  APWU01705  APWU01706  APWU01707  APWU01708  APWU01709  APWU01710  APWU01652  APWU01653  APWU01654  APWU01655  APWU01897  APWU01096  APWU01097  APWU01098  APWU01099  APWU01100  APWU01101  APWU01102  APWU01103  APWU01104  APWU01105  APWU01106  APWU01107  APWU01108  APWU01109  APWU01496  APWU01497  APWU01498  APWU01499  APWU01500  APWU01501  APWU01502  APWU01503  APWU01504  APWU01505  APWU01736  APWU01737  APWU01579  APWU01580  APWU01581  APWU01582  APWU00864  APWU00865  APWU00866  APWU00867  APWU00868  APWU00869  APWU00870  APWU00871  APWU01149  APWU01150  APWU01910  APWU00913  APWU01265  APWU01266  APWU01911  APWU01912  APWU01913  APWU01914  APWU01922  APWU01923  APWU01929  APWU01930  APWU01931  APWU01932  APWU01933  APWU01934  APWU01935  APWU01936  APWU01937  APWU01938  APWU01422  APWU00764  APWU00765  APWU00766  APWU00889  APWU02061  APWU01443  APWU01444  APWU01445  APWU01446  APWU01447  APWU01448  APWU01449  APWU01450  APWU01451  APWU01452  APWU01774  APWU01775  APWU01095  APWU00725  APWU00726  APWU00727  APWU01245  APWU01788  APWU01789  APWU01790  APWU01791  APWU01246  APWU01247  APWU01248  APWU01267  APWU01383  APWU01384  APWU01632  APWU01633  APWU01634  APWU01665  APWU01666  APWU01667  APWU01668  APWU01669  APWU01670  APWU01671  APWU01672  APWU01673  APWU01674  APWU01675  APWU01684  APWU01685  APWU01686  APWU01687  APWU01688  APWU01689  APWU01690  APWU01691  APWU01692  APWU01693  APWU01694  APWU01540  APWU01541  APWU01542  APWU01543  APWU01544  APWU01545  APWU01546  APWU01547  APWU01548  APWU01549  APWU01550  APWU01551  APWU01552  APWU01553  APWU01554  APWU01555  APWU01556  APWU01557  APWU01558  APWU01559  APWU01560  APWU01561  APWU01562  APWU01563  APWU01564  APWU01565  APWU01566  APWU01567  APWU01568  APWU01569  APWU01570  APWU01641  APWU01114  APWU00937  APWU00938  APWU00667  APWU00668  APWU00669  APWU00670  APWU00671  APWU00672  APWU00673  APWU00674  APWU00675  APWU00676  APWU00677  APWU00678  APWU02068  APWU01345  APWU01360  APWU01361  APWU01362  APWU01364  APWU01365  APWU01366  APWU01367  APWU01368  APWU01369  APWU01370  APWU01088  APWU01089  APWU01090  APWU01346  APWU01347  APWU01426  APWU01794  APWU01795  APWU01796  APWU01826  APWU01348  APWU01349  APWU01350  APWU01351  APWU01352  APWU01353  APWU01354  APWU01355  APWU01356  APWU01357  APWU01358  APWU01898  APWU01899  APWU01900  APWU01901  APWU01902  APWU01903  APWU01296  APWU01297  APWU01298  APWU01299  APWU01300  APWU01301  APWU01302  APWU01303  APWU01304  APWU01305  APWU01306  APWU01307  APWU01308  APWU01309  APWU01310  APWU01311  APWU01312  APWU01313  APWU01314  APWU01315  APWU01316  APWU01317  APWU01318  APWU01319  APWU01320  APWU01321  APWU01322  APWU01323  APWU01324  APWU01325  APWU01326  APWU01327  APWU01328  APWU01329  APWU01330  APWU01331  APWU01332  APWU01333  APWU01334  APWU01335  APWU01336  APWU01337  APWU01338  APWU01083  APWU01155  APWU01156  APWU01196  APWU01197  APWU01198  APWU01199  APWU01200  APWU01201  APWU01202  APWU01203  APWU01204  APWU01205  APWU00756  APWU00757  APWU00758  APWU00759  APWU00760  APWU00761  APWU00762  APWU00763  APWU01161  APWU01162  APWU01163  APWU01249  APWU01681  APWU01682  APWU01683  APWU01924  APWU01925  APWU01926  APWU01927  APWU01168  APWU01169  APWU01170  APWU01171  APWU01172  APWU01173  APWU01174  APWU01175  APWU01176  APWU01177  APWU01178  APWU01179  APWU01180  APWU01181  APWU01182  APWU01183  APWU01184  APWU01185  APWU01186  APWU01187  APWU01188  APWU01189  APWU01190  APWU01939  APWU01940  APWU01941  APWU01942  APWU01943  APWU01781  APWU01782  APWU01783  APWU01571  APWU01572  APWU01573  APWU01574  APWU01575  APWU01885  APWU01886  APWU01676  APWU01828  APWU01829  APWU01157  APWU01158  APWU01159  APWU01160  APWU01418  APWU01797  APWU01798  APWU01638  APWU01639  APWU01640  APWU01520  APWU01521  APWU01522  APWU01191  APWU01192  APWU01999  APWU02000  APWU02001  APWU02002  APWU02003  APWU02004  APWU02005  APWU02006  APWU02007  APWU02008  APWU02009  APWU02010  APWU02011  APWU02012  APWU02013  APWU02014  APWU02015  APWU01624  APWU01625  APWU01626  APWU01627  APWU01628  APWU01629  APWU01630  APWU01631  APWU02064  APWU02065  APWU01421  APWU00990  APWU00991  APWU00992  APWU00993  APWU00994  APWU00995  APWU00996  APWU00997  APWU00998  APWU00999  APWU00714  APWU00715  APWU00716  APWU00717  APWU00718  APWU01193  APWU01194  APWU01195  APWU01529  APWU01413  APWU01414  APWU01415  APWU01153  APWU01154  APWU01915  APWU01916  APWU01917  APWU00830  APWU00831  APWU00832  APWU01950  APWU01951  APWU00972  APWU00974  APWU00977  APWU00978  APWU00979  APWU00981  APWU01835  APWU01836  APWU01837  APWU01842  APWU01843  APWU00912  APWU01377  APWU01786  APWU01635  APWU01636  APWU01918  APWU01919  APWU01920  APWU01921  APWU01000  APWU01001  APWU01002  APWU01003  APWU01004  APWU01005  APWU01006  APWU01007  APWU01008  APWU01009  APWU01010  APWU01011  APWU01012  APWU01013  APWU01014  APWU01015  APWU01016  APWU01017  APWU01018  APWU01019  APWU01020  APWU01021  APWU01642  APWU01643  APWU01644  APWU01645  APWU01646  APWU01647  APWU01648  APWU01649  APWU01371  APWU01373  APWU01374  APWU01375  APWU01376  APWU00722  APWU01435  APWU01436  APWU01437  APWU01438  APWU01439  APWU01440  APWU01441  APWU01442  APWU00825  APWU01135  APWU01136  APWU01137  APWU01138  APWU01139  APWU01140  APWU01141  APWU01142  APWU01807  APWU01808  APWU00915  APWU00925  APWU00926  APWU00826  APWU00898  APWU01952  APWU01953  APWU01954  APWU01955  APWU01956  APWU01957  APWU01958  APWU01400  APWU01401  APWU01402  APWU01792  APWU01793  APWU00789  APWU00806  APWU00807  APWU00808  APWU00809  APWU00810  APWU00811  APWU00812  APWU00813  APWU00814  APWU00815  APWU00816  APWU00817  APWU01378  APWU01379  APWU01380  APWU01381  APWU01382  APWU00836  APWU00989  APWU01110  APWU01111  APWU01112  APWU01745  APWU01746  APWU01747  APWU00884  APWU02030  APWU02031  APWU00914  APWU01679  APWU01419  APWU01799  APWU00799  APWU00800  APWU01805  APWU00872  APWU00873  APWU00874  APWU01592  APWU01779  APWU01780  APWU00822  APWU00823  APWU00824  APWU00783  APWU00784  APWU00785  APWU01117  APWU01118  APWU01119  APWU01120  APWU01121  APWU01122  APWU01887  APWU01888  APWU01889  APWU00886  APWU00887  APWU00888  APWU01904  APWU02039  APWU01290  APWU01291  APWU01787  APWU01981  APWU01982  APWU01983  APWU01984  APWU01985  APWU01986  APWU01987  APWU00827  APWU01851  APWU01852  APWU01853  APWU01854  APWU00936  APWU01761  APWU01762  APWU01763  APWU01764  APWU01765  APWU01766  APWU01767  APWU01768  APWU01769  APWU01770  APWU01771  APWU01772  APWU01773  APWU00891  APWU00892  APWU00933  APWU00934  APWU00935  APWU01220  APWU01221  APWU01222  APWU01223  APWU01224  APWU01225  APWU01226  APWU01227  APWU01228  APWU01229  APWU01230  APWU01231  APWU01232  APWU01233  APWU01234  APWU01235  APWU01236  APWU01237  APWU01238  APWU01239  APWU01255  APWU01256  APWU01257  APWU01258  APWU01259  APWU01260  APWU01261  APWU01262  APWU01263  APWU01264  APWU01268  APWU01269  APWU01270  APWU01271  APWU01272  APWU01273  APWU01274  APWU01275  APWU01276  APWU01277  APWU01278  APWU01279  APWU01280  APWU01281  APWU01282  APWU01283  APWU01284  APWU01285  APWU01286  APWU01287  APWU01288  APWU01289  APWU01206  APWU01207  APWU01208  APWU01209  APWU01210  APWU01211  APWU01212  APWU01213  APWU01214  APWU01215  APWU01216  APWU01217  APWU01218  APWU01219  APWU01458  APWU01459  APWU01460  APWU00931  APWU00932  APWU01818  APWU01819  APWU01820  APWU01821  APWU01822  APWU01823  APWU01824  APWU00908  APWU00909  APWU00910  APWU00911  APWU01928  APWU00930  APWU01855  APWU01856  APWU01857  APWU01858  APWU01859  APWU01860  APWU01861  APWU01862  APWU01863  APWU01864  APWU01865  APWU01866  APWU01867  APWU01868  APWU01869  APWU01870  APWU01871  APWU01872  APWU01873  APWU00939  APWU00940  APWU00941  APWU01029  APWU01030  APWU01031  APWU01032  APWU01033  APWU01034  APWU01035  APWU01036  APWU01037  APWU01038  APWU01039  APWU01040  APWU01041  APWU01042  APWU01043  APWU01044  APWU01045  APWU01046  APWU01047  APWU01048  APWU01049  APWU01050  APWU01051  APWU01661  APWU01662  APWU01663  APWU01664  APWU01874  APWU01875  APWU01461  APWU01462  APWU01463  APWU01596  APWU01597  APWU01512  APWU01850  APWU01387  APWU01388  APWU01389  APWU01390  APWU01391  APWU01392  APWU01393  APWU01394  APWU01395  APWU01396  APWU01397  APWU01511  APWU01060  APWU01534  APWU01730  APWU01731  APWU01778  APWU01678  APWU01523  APWU01749  APWU01750  APWU01729  APWU01881  APWU00848  APWU00849  APWU00850  APWU01776  APWU01892  APWU01893  APWU01894  APWU01895  APWU01896  APWU01533  APWU01148  APWU02037  APWU02038  APWU01485  APWU01429  APWU01430  APWU01967  APWU01968  APWU01969  APWU01970  APWU01971  APWU01972  APWU01973  APWU01974  APWU01975  APWU01976  APWU01977  APWU01978  APWU01979  APWU01423  APWU01403  APWU01151  APWU00921  APWU00922  APWU00923  APWU00969  APWU00970  APWU00971  APWU01883  APWU01884  APWU01876  APWU01877  APWU01878  APWU00828  APWU00829  APWU01022  APWU01023  APWU01024  APWU01025  APWU01589  APWU01590  APWU01591  APWU01804  APWU00796  APWU00924  APWU02035  APWU00893  APWU00894  APWU01738  APWU01739  APWU01740  APWU01741  APWU02060  APWU02062  APWU00801  APWU00802  APWU00803  APWU00804  APWU00805  APWU01696  APWU01697  APWU01593  APWU01594  APWU01084  APWU01085  APWU01086  APWU01432  APWU01433  APWU01431  APWU01091  APWU01092  APWU01677  APWU01947  APWU01948  APWU01959  APWU01960  APWU01961  APWU01962  APWU01963  APWU01964  APWU01965  APWU01966  APWU00781  APWU00782  APWU00787  APWU01603  APWU01604  APWU01605  APWU01606  APWU01607  APWU02046  APWU02047  APWU00838  APWU00839  APWU02021  APWU02022  APWU02023  APWU02024  APWU02025  APWU02026  APWU02027  APWU02028  APWU02029  APWU00794  APWU00821  APWU00875  APWU00876  APWU00877  APWU00878  APWU00879  APWU02040  APWU02041  APWU01800  APWU01806  APWU01116  APWU01660  APWU01742  APWU01743  APWU01744  APWU00818  APWU02063  APWU01152  APWU00916  APWU00917  APWU00918  APWU00919  APWU00920  APWU00927  APWU00928  APWU00929  APWU01134  APWU01416  APWU01711  APWU00790  APWU00771  APWU00772  APWU00773  APWU00774  APWU00775  APWU00797  APWU01890  APWU01891  APWU01372  APWU00986  APWU01814  APWU01815  APWU00982  APWU01166  APWU01167  APWU01398  APWU01399  APWU00987  APWU01845  APWU00903  APWU00904  APWU01292  APWU01777  APWU02059  APWU01598  APWU00820  APWU01052  APWU01424  APWU00776  APWU00777  APWU00778  APWU00779  APWU00780  APWU01123  APWU01124  APWU01125  APWU01126  APWU01127  APWU01128  APWU01339  APWU01698  APWU00984  APWU01583  APWU02071  APWU00719  APWU01830  APWU01880  APWU01531  APWU01532  APWU00720  APWU00721  APWU01831  APWU02032  APWU01053  APWU01784  APWU01879  APWU01945  APWU01944  APWU01464  APWU01465  APWU01466  APWU02049  APWU02050  APWU01404  APWU01405  APWU01406  APWU01801  APWU01802  APWU01803  APWU01026  APWU01732  APWU01386  APWU01946  APWU01832  APWU02048  APWU00706  APWU00890  APWU01833  APWU01846  APWU01847  APWU01848  APWU01849  APWU01407  APWU01408  APWU01409  APWU01656  APWU01657  APWU01658  APWU01659  APWU02033  APWU02034  APWU01113  APWU01524  APWU01530  APWU00786  APWU02036  APWU00880  APWU00792  APWU00793  APWU01427  APWU01428  APWU00851  APWU00852  APWU00853  APWU00854  APWU00855  APWU00856  APWU00857  APWU00858  APWU00859  APWU00860  APWU00861  APWU00862  APWU00863  APWU01712  APWU01713  APWU01714  APWU01715  APWU01716  APWU01717  APWU01718  APWU01719  APWU01720  APWU01721  APWU01722  APWU01723  APWU01724  APWU01725  APWU01726  APWU01727  APWU01728  APWU01751  APWU01752  APWU01753  APWU01754  APWU01755  APWU01756  APWU01757  APWU01758  APWU01759  APWU01760  APWU01908  APWU00905  APWU00906  APWU00907  APWU00983  APWU00988  APWU01385  APWU00833  APWU00834  APWU00835  APWU01434  APWU00788  APWU01417  APWU01748  APWU01882  APWU01980  APWU02042  APWU02043  APWU02044  APWU01695  APWU00770  APWU00791  APWU01733  APWU01734  APWU01735  APWU01513  APWU01514  APWU02086  APWU02087  APWU02089  APWU02091  APWU00795  APWU00881  APWU00882  APWU00883  APWU00895  APWU00896  APWU00897  APWU00899  APWU00900  APWU00901  APWU00902  APWU00942  APWU00943  APWU00944  APWU00945  APWU00946  APWU00947  APWU00948  APWU00949  APWU00950  APWU00951  APWU00952  APWU00953  APWU00954  APWU00955  APWU00956  APWU00957  APWU00958  APWU00959  APWU00960  APWU00961  APWU00962  APWU00963  APWU00964  APWU00965  APWU00966  APWU00967  APWU00968  APWU01027  APWU01028  APWU01061  APWU01062  APWU01063  APWU01064  APWU01065  APWU01066  APWU01067  APWU01068  APWU01069  APWU01070  APWU01071  APWU01072  APWU01073  APWU01074  APWU01075  APWU01076  APWU01077  APWU01078  APWU01079  APWU01080  APWU01081  APWU01082  APWU01145  APWU01146  APWU01147  APWU01420  APWU01425  APWU01483  APWU01484  APWU01515  APWU01516  APWU01517  APWU01518  APWU01519  APWU01816  APWU01817  APWU01834  APWU02016  APWU02045  APWU01637  APWU01143  APWU01144  APWU01588  APWU02069  APWU01467  APWU01468  APWU01469  APWU01470  APWU01471  APWU01472  APWU01473  APWU01474  APWU01475  APWU01476  APWU01477  APWU01478  APWU01479  APWU01480  APWU01481  APWU01482  APWU00819  APWU01506  APWU01507  APWU00885  APWU00798  APWU00837  APWU01087  APWU01164  APWU01165  APWU01250  APWU01251  APWU01252  APWU01253  APWU01525  APWU01526  APWU01527  APWU01528  APWU01536  APWU01537  APWU01538  APWU01539  APWU01584  APWU01585  APWU01586  APWU01587  APWU01535  APWU01093  APWU01094  APWU01905  APWU01906  APWU01907  APWU01293  APWU01294  APWU01295  APWU01129  APWU01130  APWU01115  APWU01809  APWU01810  APWU01811  APWU01812  APWU01813  APWU01650  APWU01651    AVANCE    ANBU00119  ANBU00124  ANBU00134  ANBU00138  ANBU00139  ANBU00163  ANBU00121  ANBU00122  ANBU00123  ANBU00125  ANBU00126  ANBU00127  ANBU00128  ANBU00129  ANBU00130  ANBU00131  ANBU00120    AVANCE CS2    APKU01696  APKU01697  APKU01698  APKU01699  APKU01872  APKU01237  APKU01238  APKU01239  APKU01240  APKU01241  APKU01242  APKU01243  APKU01569  APKU01570  APKU02074  APKU01701  APKU01427  APKU01694  APKU01053  APKU02163  APKU02164  APKU01200  APKU01201  APKU01202  APKU01203  APKU01204  APKU01686  APKU01687  APKU01688  APKU01904  APKU01810  APKU01447  APKU01742  APKU01743  APKU01656  APKU01657  APKU01658  APKU01682  APKU01683  APKU01684  APKU01685  APKU01729  APKU01730  APKU01731  APKU01454  APKU01455  APKU01456  APKU01457  APKU01458  APKU01459  APKU01460  APKU01461  APKU01462  APKU01463  APKU01464  APKU01465  APKU01466  APKU01467  APKU01468  APKU01469  APKU01470  APKU01473  APKU01378  APKU01379  APKU01380  APKU01381  APKU01382  APKU01383  APKU01384  APKU01385  APKU01021  APKU02049  APKU02050  APKU02051  APKU02052  APKU02053  APKU02054  APKU02055  APKU02056  APKU02057  APKU02058  APKU02059  APKU02060  APKU02061  APKU02062  APKU02063  APKU02064  APKU02065  APKU02066  APKU02067  APKU02068  APKU02069  APKU02070  APKU02071  APKU02072  APKU02073  APKU02075  APKU02076  APKU02077  APKU02078  APKU02079  APKU02080  APKU02081  APKU02082  APKU02083  APKU02084  APKU02085  APKU02086  APKU02087  APKU02088  APKU02089  APKU02090  APKU02091  APKU02092  APKU02093  APKU02094  APKU02095  APKU02096  APKU02097  APKU02098  APKU02099  APKU02234  APKU02235  APKU02236  APKU02237  APKU02238  APKU02239  APKU02240  APKU02241  APKU02242  APKU02243  APKU02268  APKU02269  APKU02270  APKU02271  APKU02272  APKU02273  APKU02274  APKU02275  APKU02276  APKU02277  APKU02295  APKU02296  APKU02297  APKU02298  APKU02299  APKU02300  APKU02301  APKU02302  APKU02303  APKU02304  APKU02307  APKU02308  APKU02309  APKU02310  APKU02311  APKU02312  APKU02313  APKU02314  APKU02315  APKU02316  APKU02366  APKU02367  APKU01178  APKU01268  APKU01269  APKU01270  APKU01617  APKU01618  APKU01619  APKU01326  APKU01780  APKU01164  APKU01183  APKU01361  APKU01362  APKU01363  APKU01719  APKU01720  APKU01721  APKU01722  APKU01723  APKU02267  APKU00988  APKU02036  APKU02037  APKU02038  APKU01959  APKU01960  APKU01784  APKU01785  APKU01924  APKU01925  APKU01124  APKU02244  APKU02245  APKU02246  APKU02247  APKU02248  APKU02249  APKU02250  APKU02251  APKU02118  APKU02119  APKU02120  APKU02121  APKU02122  APKU01961  APKU01167  APKU01168  APKU01169  APKU01170  APKU01171  APKU01172  APKU01180  APKU01181  APKU01182  APKU01364  APKU02103  APKU02104  APKU01224  APKU01444  APKU01744  APKU01174  APKU01365  APKU01366  APKU01367  APKU01368  APKU01369  APKU01370  APKU01430  APKU01431  APKU01432  APKU01153  APKU02126  APKU02223  APKU02224  APKU02225  APKU02226  APKU01681  APKU01689  APKU01690  APKU01691  APKU01692  APKU01693  APKU02128  APKU02129  APKU01438  APKU02100  APKU02101  APKU02102  APKU02116  APKU02117  APKU02112  APKU02230  APKU01160  APKU01396  APKU01108  APKU01048  APKU01049  APKU01050  APKU01404  APKU01405  APKU01878  APKU02016  APKU02017  APKU02018  APKU01391  APKU01754  APKU01755  APKU01756  APKU01757  APKU01704  APKU01705  APKU01706  APKU01707  APKU01708  APKU02130  APKU02131  APKU02132  APKU02133  APKU02134  APKU02135  APKU02136  APKU02137  APKU02138  APKU02139  APKU02140  APKU02141  APKU02142  APKU02143  APKU02144  APKU02145  APKU02146  APKU02147  APKU02221  APKU02222  APKU02260  APKU02261  APKU02262  APKU02263  APKU02264  APKU02265  APKU02031  APKU02032  APKU01550  APKU02215  APKU02216  APKU02217  APKU02218  APKU02219  APKU02220  APKU02305  APKU02306  APKU01342  APKU01343  APKU01344  APKU01345  APKU01346  APKU01347  APKU01348  APKU01349  APKU01350  APKU01351  APKU01352  APKU01353  APKU01354  APKU01355  APKU01356  APKU01357  APKU01358  APKU02041  APKU02042  APKU02043  APKU01324  APKU01909  APKU01910  APKU01911  APKU01912  APKU01913  APKU01914  APKU01915  APKU01916  APKU01917  APKU01918  APKU01919  APKU02109  APKU02110  APKU02111  APKU01265  APKU01266  APKU02291  APKU02292  APKU01069  APKU01070  APKU01071  APKU01072  APKU01584  APKU01051  APKU01772  APKU01773  APKU01774  APKU01775  APKU01745  APKU01746  APKU01747  APKU01983  APKU01984  APKU01985  APKU01424  APKU01425  APKU01426  APKU02331  APKU02336  APKU02357  APKU02360  APKU02361  APKU01951  APKU01952  APKU01953  APKU01954  APKU01955  APKU01019  APKU01195  APKU01647  APKU01648  APKU01649  APKU02169  APKU02170  APKU02171  APKU02172  APKU02173  APKU02174  APKU02175  APKU02176  APKU02177  APKU02178  APKU02179  APKU02180  APKU02181  APKU02182  APKU01998  APKU01997  APKU01398  APKU01399  APKU01400  APKU01371  APKU01372  APKU01373  APKU01374  APKU01375  APKU01376  APKU01377  APKU01975  APKU01976  APKU01968  APKU01969  APKU01970  APKU01971  APKU01972  APKU01973  APKU01974  APKU01977  APKU01978  APKU01979  APKU01980  APKU01981  APKU01982  APKU01401  APKU01402  APKU01551  APKU01552  APKU01711  APKU01712  APKU01713  APKU02021  APKU02022  APKU02023  APKU01962  APKU01963  APKU01964  APKU01965  APKU01966  APKU01841  APKU01842  APKU01359  APKU02048  APKU01718  APKU01849  APKU01850  APKU01851  APKU01852  APKU01853  APKU01854  APKU01855  APKU01856  APKU01857  APKU01858  APKU01859  APKU02157  APKU02158  APKU02159  APKU02160  APKU02161  APKU02162  APKU01967  APKU02105  APKU01926  APKU02024  APKU02025  APKU02026  APKU02027  APKU02028  APKU01740  APKU01613  APKU01614  APKU01615  APKU01948  APKU01929  APKU01930  APKU01931  APKU01406  APKU01407  APKU01408  APKU01409  APKU01410  APKU01411  APKU01412  APKU01413  APKU01414  APKU01415  APKU01416  APKU01417  APKU01418  APKU01419  APKU01026  APKU01027  APKU02002  APKU01620  APKU01621  APKU01165  APKU02029  APKU02030  APKU02106  APKU01587  APKU01588  APKU01589  APKU01579  APKU01580  APKU01581  APKU01582  APKU00993  APKU01726  APKU01717  APKU01820  APKU01811  APKU01812  APKU01813  APKU01814  APKU01634  APKU01635  APKU01636  APKU01637  APKU01638  APKU01639  APKU01640  APKU01641  APKU01079  APKU01943  APKU01136  APKU01765  APKU01766  APKU01433  APKU01434  APKU01435  APKU01436  APKU01137  APKU01138  APKU01139  APKU01140  APKU01141  APKU01142  APKU01905  APKU01906  APKU02228  APKU02341  APKU02342  APKU02343  APKU02346  APKU01830  APKU01831  APKU01832  APKU01833  APKU01834  APKU01835  APKU01986  APKU01987  APKU01988  APKU01325  APKU00983  APKU00984  APKU02229  APKU01724  APKU01725  APKU02378  APKU02019  APKU02020  APKU01583  APKU01152  APKU01502  APKU01503  APKU01504  APKU01505  APKU01506  APKU01507  APKU01508  APKU01509  APKU01510  APKU01511  APKU01512  APKU01513  APKU01514  APKU01515  APKU01516  APKU01517  APKU01518  APKU01519  APKU01520  APKU01521  APKU01437  APKU01471  APKU01472  APKU01568  APKU01727  APKU01163  APKU01360  APKU01395  APKU01080  APKU01081  APKU01082  APKU01083  APKU01084  APKU01732  APKU01733  APKU01734  APKU01735  APKU01736  APKU01737  APKU01738  APKU01776  APKU01777  APKU01778  APKU01779  APKU02044  APKU02045  APKU02046  APKU02047  APKU01932  APKU01933  APKU01934  APKU01935  APKU01936  APKU01937  APKU01938  APKU01527  APKU01528  APKU01529  APKU01530  APKU01531  APKU01532  APKU01533  APKU01534  APKU01535  APKU01536  APKU01537  APKU01538  APKU01539  APKU01540  APKU01541  APKU01542  APKU01543  APKU01544  APKU01545  APKU01546  APKU01547  APKU01548  APKU01403  APKU01062  APKU01063  APKU01064  APKU01065  APKU01066  APKU01067  APKU01068  APKU01327  APKU01328  APKU01329  APKU01330  APKU01331  APKU01332  APKU01333  APKU01334  APKU01335  APKU01336  APKU01386  APKU01387  APKU01388  APKU01389  APKU01390  APKU01392  APKU01393  APKU01394  APKU01158  APKU01159  APKU01397  APKU00977  APKU00978  APKU00979  APKU00980  APKU00981  APKU00982  APKU00985  APKU00986  APKU00987  APKU00989  APKU00990  APKU00991  APKU00992  APKU02039  APKU02040  APKU01090  APKU01091  APKU01092  APKU01093  APKU01094  APKU01095  APKU01096  APKU01097  APKU01098  APKU01099  APKU01100  APKU01101  APKU01210  APKU01218  APKU01646  APKU01112  APKU01113  APKU02167  APKU02168  APKU01023  APKU01024  APKU01025  APKU02227  APKU01817  APKU01029  APKU01030  APKU01146  APKU01819  APKU01247  APKU01448  APKU01449  APKU01450  APKU01451  APKU01452  APKU01453  APKU01556  APKU01557  APKU01558  APKU01559  APKU01560  APKU01561  APKU01562  APKU01563  APKU01564  APKU01565  APKU01566  APKU02001  APKU01290  APKU01291  APKU01292  APKU01293  APKU01702  APKU01703  APKU01157  APKU01222  APKU01223  APKU01233  APKU01314  APKU01315  APKU01316  APKU01317  APKU01318  APKU01816  APKU02015  APKU01089  APKU01881  APKU01882  APKU01883  APKU01884  APKU01885  APKU01886  APKU01610  APKU00994  APKU01606  APKU02152  APKU02350  APKU01497  APKU01498  APKU01499  APKU01500  APKU01788  APKU01145  APKU01211  APKU01212  APKU01213  APKU01214  APKU01319  APKU01555  APKU01791  APKU01795  APKU01796  APKU01797  APKU01423  APKU01028  APKU01143  APKU01144  APKU01173  APKU01874  APKU01875  APKU01876  APKU01877  APKU02107  APKU02108  APKU01622  APKU01626  APKU01627  APKU01799  APKU01815  APKU01150  APKU01151  APKU01219  APKU01220  APKU01770  APKU01771  APKU01787  APKU01287  APKU01337  APKU01794  APKU01871  APKU01215  APKU01216  APKU01217  APKU01764  APKU01903  APKU01184  APKU01198  APKU01768  APKU01286  APKU01792  APKU02148  APKU01230  APKU01767  APKU01205  APKU01879  APKU01728  APKU01323  APKU01821  APKU01822  APKU01823  APKU01824  APKU01843  APKU01844  APKU01845  APKU01846  APKU01847  APKU01848  APKU01956  APKU01957  APKU01958  APKU02127  APKU02183  APKU02184  APKU02185  APKU02186  APKU02187  APKU02188  APKU02189  APKU02190  APKU02191  APKU02192  APKU02193  APKU02194  APKU02195  APKU02196  APKU02197  APKU02198  APKU02199  APKU02200  APKU02201  APKU01102  APKU01103  APKU01104  APKU01105  APKU01236  APKU01420  APKU01421  APKU01422  APKU02003  APKU02004  APKU02005  APKU01313  APKU01052  APKU01054  APKU01177  APKU02323  APKU01252  APKU01253  APKU01254  APKU01250  APKU01251  APKU01623  APKU01624  APKU01625  APKU01055  APKU01056  APKU01057  APKU01255  APKU01256  APKU01257  APKU01259  APKU01260  APKU01261  APKU01262  APKU01263  APKU01264  APKU01175  APKU01176  APKU01297  APKU01298  APKU01710  APKU01321  APKU02231  APKU02232  APKU01950  APKU01741  APKU02214  APKU01301  APKU01302  APKU01303  APKU01304  APKU01299  APKU01300  APKU01781  APKU02325  APKU02326  APKU01338  APKU01234  APKU01235  APKU01793  APKU01758  APKU01654  APKU01655  APKU01075  APKU01322  APKU02123  APKU02124  APKU02125  APKU01786  APKU01644  APKU01645  APKU01154  APKU01873  APKU01616  APKU01993  APKU01553  APKU01199  APKU01305  APKU01306  APKU01307  APKU01308  APKU01309  APKU01310  APKU01311  APKU01312  APKU01085  APKU01086  APKU01087  APKU01088  APKU01110  APKU01267  APKU01898  APKU01899  APKU01900  APKU01901  APKU01902  APKU01428  APKU01429  APKU01492  APKU01908  APKU01487  APKU01488  APKU01489  APKU01490  APKU01491  APKU01592  APKU01628  APKU01629  APKU01630  APKU01632  APKU01633  APKU01045  APKU01058  APKU01585  APKU01586  APKU01046  APKU01155  APKU01196  APKU01197  APKU02290  APKU01147  APKU01148  APKU01149  APKU01493  APKU01494  APKU02233  APKU01179  APKU01474  APKU01475  APKU01476  APKU01769  APKU01232  APKU01642  APKU02330  APKU01248  APKU01484  APKU02153  APKU01156  APKU01495  APKU01161  APKU01501  APKU01496  APKU01078  APKU02149  APKU02150  APKU02151  APKU01061  APKU01486  APKU01700  APKU01483  APKU01818  APKU02155  APKU02156  APKU01880  APKU01249  APKU01949  APKU01567  APKU01477  APKU01478  APKU01479  APKU01320  APKU02165  APKU02166  APKU01059  APKU01944  APKU02113  APKU01166  APKU01612  APKU01762  APKU01763  APKU01272  APKU01273  APKU01485  APKU01947  APKU01132  APKU01643  APKU01111  APKU01125  APKU01126  APKU01127  APKU01129  APKU01130  APKU01131  APKU01133  APKU01134  APKU01135  APKU01825  APKU01826  APKU01827  APKU01828  APKU01829  APKU01945  APKU01946  APKU02033  APKU02034  APKU02035  APKU02293  APKU02294  APKU01022  APKU01128  APKU01274  APKU01275  APKU01276  APKU01277  APKU01278  APKU01279  APKU01280  APKU01281  APKU01282  APKU01283  APKU01284  APKU01714  APKU01715  APKU01716  APKU01923  APKU02114  APKU02115  APKU02329  APKU00973  APKU01221  APKU01229  APKU01605  APKU01185  APKU01186  APKU01187  APKU01188  APKU01189  APKU01190  APKU01191  APKU01192  APKU01193  APKU01194  APKU01074  APKU01789  APKU01790  APKU02007  APKU02008  APKU01860  APKU01861  APKU01862  APKU01863  APKU01864  APKU01865  APKU01866  APKU01867  APKU01868  APKU01869  APKU01870  APKU01549  APKU01554  APKU01162  APKU01887  APKU01888  APKU01481  APKU01482  APKU01651  APKU01652  APKU02209  APKU02210  APKU02211  APKU02212  APKU02213  APKU01590  APKU01591  APKU01695  APKU01808  APKU01809  APKU02324  APKU02327  APKU02328  APKU01341  APKU01445  APKU01446  APKU02317  APKU02318  APKU02319  APKU02320  APKU01047  APKU02009  APKU02010  APKU01800  APKU01801  APKU01802  APKU02351  APKU02356  APKU00974  APKU00975  APKU00976  APKU01578  APKU01060  APKU02287  APKU02288  APKU02289  APKU01073  APKU02006  APKU01020  APKU01114  APKU01115  APKU01116  APKU01117  APKU01118  APKU01119  APKU01120  APKU01121  APKU01122  APKU01123  APKU01798  APKU01803  APKU01804  APKU01761  APKU01759  APKU01760  APKU01994  APKU01995  APKU01996  APKU02321  APKU02322  APKU01593  APKU01594  APKU01595  APKU01596  APKU01597  APKU01598  APKU01076  APKU01077  APKU01339  APKU01340  APKU01805  APKU01806  APKU01922  APKU01999  APKU02000  APKU02202  APKU02203  APKU02204  APKU02205  APKU02206  APKU02207  APKU02208  APKU01206  APKU01207  APKU01208  APKU01209  APKU01225  APKU01226  APKU01227  APKU01228  APKU01288  APKU01289  APKU01889  APKU01890  APKU01891  APKU01892  APKU01893  APKU01894  APKU01895  APKU01896  APKU01897  APKU01523  APKU01524  APKU01525  APKU01526  APKU01522  APKU01653  APKU00868  APKU01439  APKU01440  APKU01441  APKU01442  APKU01443  APKU02154  APKU01285  APKU01599  APKU01600  APKU01601  APKU01602  APKU01603  APKU01604  APKU01031  APKU01032  APKU01807  APKU01106  APKU01107  APKU01611  APKU02011  APKU02012  APKU02013  APKU02014  APKU02278  APKU02279  APKU02280  APKU02281  APKU02282  APKU02283  APKU02284  APKU02285  APKU02286  APKU01782  APKU01783  APKU01258  APKU01271  APKU01907  APKU01989  APKU01990  APKU01991  APKU01992  APKU01109  APKU01294  APKU01295  APKU01296  APKU01748  APKU01749  APKU01750  APKU01751  APKU01752  APKU01753  APKU01920  APKU01921  APKU01927  APKU01928  APKU01939  APKU01940  APKU01941  APKU01942",