      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72943",
      "recalling_firm": "GE Medical Systems, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (nationwide including PR except NH,  VT, & WY.  Internationally to ALBANIA, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BANGLADESH, BELGIUM, BOLIVIA, BOSNIA, BRAZIL, CAMBODIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, ESTONIA, FINLAND, FRANCE, GEORGIA, GERMANY, CHANA, GIBRALTAR, GREECE, GUATEMALA, HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN KENYA, REPUBLIC OF KOREA, KUWAIT, LATVIA, LITHUANIA,MALAYSIA, MALDIVES, MARTINIQUE, MEXICO, MOROCCO, NETHERLANDS, NEW ZEALAND, NICARAGUA, NORWAY, OMAN, PAKISTAN, PERU, PHILIPPINES, POLAND, QATAR, REUNION, ROMANIA, RUSSIAN FEDERATION SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SRI lanka, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, VENEZUELA, VIETNAM.",
      "recall_number": "Z-0677-2016",
      "product_description": "GE Healthcare, Aespire 7900, 1009-9012-000 &   Aespire View, 1009-9212-000.  The family of GE Datex-Ohmeda Aespire anesthesia systems with 7900 ventilator (Aespire 7900 and Aespire View) is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients. The devices are intended for volume or pressure control ventilation.",
      "product_quantity": "1081",
      "reason_for_recall": "The Drive Gas Check Valve of certain Anesthesia products could become stuck in a fixed open position, which could cause pressure to build in the mechanical ventilation cycle. If this issue is left unresolved, it could result in excessive or prolonged pressure in the patient breathing circuit during ventilation potentially resulting in barotrauma.",
      "recall_initiation_date": "20151215",
      "center_classification_date": "20160121",
      "termination_date": "20160921",
      "report_date": "20160127",
      "code_info": "AESPIRE 7900    ANCU00662  ANCU00813  ANCU00814  ANCU00815  ANCU00816  ANCU00817  ANCU00818  ANCU00597  ANCU00598  ANCU00599  ANCU00600  ANCU00601  ANCU00671  ANCU00672  ANCU00673  ANCU00676  ANCU00678  ANCU00683  ANCU00830  ANCU00832  ANCU00870  ANCU00871  ANCU00872  ANCU00873  ANCU00874  ANCU00875  ANCU00629  ANCU00630  ANCU00638  ANCU00639  ANCU00640  ANCU00641  ANCU00806  ANCU00604  ANCU00684  ANCU00836  ANCU00837  ANCU00839  ANCU00786  ANCU00680  ANCU00681  ANCU00625  ANCU00626  ANCU00677  ANCU00618  ANCU00619  ANCU00620  ANCU00602  ANCU00603  ANCU00448  ANCU00449  ANCU00770  ANCU00771  ANCU00865  ANCU00866  ANCU00867  ANCU00803  ANCU00605  ANCU00663  ANCU00682  ANCU00685  ANCU00458  ANCU00459  ANCU00831  ANCU00653  ANCU00820  ANCU00591  ANCU00592  ANCU00586  ANCU00583  ANCU00584  ANCU00585  ANCU00825  ANCU00593  ANCU00664  ANCU00838  ANCU00433  ANCU00679  ANCU00555  ANCU00456  ANCU00457  ANCU00590  ANCU00661  ANCU00669  ANCU00670  ANCU00674  ANCU00819  ANCU00807  ANCU00808  ANCU00809  ANCU00810  ANCU00811  ANCU00812  ANCU00588  ANCU00589  ANCU00455  ANCU00587  ANCU00868  ANCU00581  ANCU00582  ANCU00466  ANCU00467  ANCU00468  ANCU00469  ANCU00840  ANCU00434  ANCU00435  ANCU00436  ANCU00437  ANCU00438  ANCU00439  ANCU00440  ANCU00822  ANCU00823  ANCU00824  ANCU00827  ANCU00828  ANCU00654  ANCU00729  ANCU00730  ANCU00496  ANCU00675  ANCU00624  ANCU00826  ANCU00829  ANCU00465  ANCU00441  ANCU00844  ANCU00656  ANCU00876  ANCU00877  ANCU00878  ANCU00426  ANCU00427  ANCU00485  ANCU00470  ANCU00608  ANCU00609  ANCU00610  ANCU00611  ANCU00612  ANCU00613  ANCU00789  ANCU00617  ANCU00445  ANCU00446  ANCU00490  ANCU00491  ANCU00492  ANCU00493  ANCU00494  ANCU00495  ANCU00497  ANCU00498  ANCU00518  ANCU00519  ANCU00521  ANCU00522  ANCU00665  ANCU00666  ANCU00709  ANCU00710  ANCU00711  ANCU00772  ANCU00773  ANCU00845  ANCU00846  ANCU00687  ANCU00688  ANCU00737  ANCU00738  ANCU00859  ANCU00860  ANCU00861  ANCU00693  ANCU00694  ANCU00695  ANCU00851  ANCU00852  ANCU00853  ANCU00854  ANCU00855  ANCU00856  ANCU00857  ANCU00858  ANCU00500  ANCU00501  ANCU00505  ANCU00506  ANCU00507  ANCU00508  ANCU00509  ANCU00510  ANCU00511  ANCU00516  ANCU00517  ANCU00696  ANCU00697  ANCU00698  ANCU00699  ANCU00705  ANCU00706  ANCU00707  ANCU00708  ANCU00715  ANCU00712  ANCU00713  ANCU00714  ANCU00569  ANCU00570  ANCU00571  ANCU00572  ANCU00450  ANCU00512  ANCU00513  ANCU00514  ANCU00515  ANCU00648  ANCU00649  ANCU00650  ANCU00735  ANCU00736  ANCU00863  ANCU00442  ANCU00444  ANCU00781  ANCU00464  ANCU00499  ANCU00701  ANCU00523  ANCU00524  ANCU00525  ANCU00526  ANCU00527  ANCU00530  ANCU00531  ANCU00532  ANCU00533  ANCU00534  ANCU00535  ANCU00536  ANCU00537  ANCU00538  ANCU00539  ANCU00540  ANCU00541  ANCU00542  ANCU00543  ANCU00544  ANCU00545  ANCU00546  ANCU00547  ANCU00548  ANCU00549  ANCU00550  ANCU00551  ANCU00552  ANCU00553  ANCU00554  ANCU00594  ANCU00595  ANCU00596  ANCU00731  ANCU00732  ANCU00733  ANCU00739  ANCU00740  ANCU00741  ANCU00742  ANCU00743  ANCU00744  ANCU00745  ANCU00746  ANCU00747  ANCU00758  ANCU00759  ANCU00760  ANCU00761  ANCU00762  ANCU00763  ANCU00764  ANCU00765  ANCU00717  ANCU00718  ANCU00719  ANCU00720  ANCU00721  ANCU00722  ANCU00723  ANCU00724  ANCU00725  ANCU00726  ANCU00727  ANCU00728  ANCU00489  ANCU00528  ANCU00529  ANCU00716  ANCU00484  ANCU00700  ANCU00463  ANCU00606  ANCU00692  ANCU00864  ANCU00821  ANCU00504  ANCU00623  ANCU00502  ANCU00503  ANCU00785  ANCU00520  ANCU00556  ANCU00557  ANCU00558  ANCU00559  ANCU00560  ANCU00804  ANCU00805  ANCU00686  ANCU00488  ANCU00447  ANCU00647  ANCU00842  ANCU00454  ANCU00862  ANCU00451  ANCU00452  ANCU00453  ANCU00689  ANCU00690  ANCU00691  ANCU00425  ANCU00428  ANCU00429  ANCU00430  ANCU00431  ANCU00487  ANCU00561  ANCU00567  ANCU00568  ANCU00607  ANCU00615  ANCU00616  ANCU00621  ANCU00627  ANCU00642  ANCU00643  ANCU00645  ANCU00655  ANCU00657  ANCU00658  ANCU00659  ANCU00660  ANCU00667  ANCU00668  ANCU00782  ANCU00783  ANCU00784  ANCU00790  ANCU00791  ANCU00792  ANCU00797  ANCU00799  ANCU00802  ANCU00833  ANCU00834  ANCU00835  ANCU00841  ANCU00869  ANCU00879  ANCU00880  ANCU00881  ANCU00882  ANCU00883  ANCU00884  ANCU00646  ANCU00766  ANCU00767  ANCU00768  ANCU00769  ANCU00847  ANCU00848  ANCU00486  ANCU00432  ANCU00651  ANCU00652  ANCU00614  ANCU00843  ANCU00562  ANCU00563  ANCU00564  ANCU00565  ANCU00566  ANCU00573  ANCU00574  ANCU00575  ANCU00576  ANCU00577  ANCU00578  ANCU00579  ANCU00580  ANCU00622  ANCU00702  ANCU00703  ANCU00704  ANCU00787  ANCU00788  ANCU00793  ANCU00794  ANCU00795  ANCU00796  ANCU00798      AESPIRE VIEW    APHU01097  APHU00884  APHU00885  APHU00886  APHU00887  APHU00888  APHU01044  APHU01045  APHU01046  APHU01047  APHU01048  APHU01049  APHU01109  APHU00641  APHU00560  APHU01035  APHU00923  APHU00924  APHU00969  APHU00970  APHU00971  APHU00895  APHU00972  APHU00742  APHU01133  APHU01134  APHU00737  APHU00738  APHU00739  APHU00740  APHU01180  APHU00554  APHU00555  APHU00500  APHU00731  APHU00732  APHU00733  APHU00734  APHU00735  APHU00736  APHU00777  APHU01087  APHU01126  APHU01127  APHU01128  APHU00493  APHU00881  APHU00882  APHU00995  APHU00553  APHU00559  APHU00528  APHU01031  APHU00516  APHU00580  APHU00581  APHU00582  APHU00583  APHU00610  APHU01132  APHU01146  APHU00597  APHU00598  APHU01129  APHU01130  APHU01101  APHU00889  APHU00741  APHU00729  APHU00730  APHU00796  APHU00797  APHU00706  APHU00716  APHU00717  APHU00718  APHU00719  APHU00720  APHU00721  APHU00722  APHU00723  APHU01099  APHU00491  APHU00495  APHU00700  APHU00701  APHU00702  APHU00703  APHU00704  APHU00705  APHU00767  APHU00798  APHU00800  APHU00806  APHU00807  APHU00808  APHU00809  APHU00810  APHU00811  APHU00812  APHU00813  APHU00814  APHU00815  APHU00816  APHU00817  APHU00890  APHU00517  APHU00518  APHU00519  APHU00520  APHU00820  APHU00821  APHU00859  APHU01094  APHU01095  APHU01096  APHU00556  APHU00492  APHU00463  APHU00464  APHU00465  APHU00466  APHU00467  APHU01118  APHU01119  APHU01120  APHU01121  APHU00490  APHU00496  APHU00758  APHU00759  APHU00760  APHU00761  APHU00762  APHU00763  APHU00764  APHU00765  APHU00766  APHU00489  APHU00757  APHU01125  APHU01175  APHU01176  APHU00602  APHU00603  APHU00839  APHU00991  APHU00992  APHU00521  APHU00514  APHU00501  APHU00966  APHU00967  APHU01023  APHU00914  APHU00917  APHU00557  APHU00918  APHU00922  APHU00529  APHU00530  APHU01064  APHU01065  APHU01077  APHU01100  APHU00645  APHU00646  APHU00647  APHU00915  APHU00916  APHU01147  APHU00851  APHU00903  APHU00904  APHU00905  APHU00906  APHU00907  APHU00573  APHU00574  APHU00575  APHU00576  APHU00577  APHU00578  APHU01110  APHU00842  APHU00843  APHU00844  APHU00845  APHU00846  APHU00847  APHU00689  APHU00690  APHU00691  APHU00935  APHU01152  APHU01148  APHU01149  APHU01150  APHU01151  APHU01153  APHU01154  APHU01155  APHU01158  APHU01160  APHU00932  APHU01159  APHU01156  APHU01157  APHU01192  APHU01195  APHU00552  APHU00751  APHU00752  APHU00801  APHU00802  APHU00803  APHU00531  APHU00532  APHU00533  APHU00534  APHU00535  APHU00536  APHU01188  APHU01189  APHU00596  APHU00612  APHU00613  APHU00614  APHU00615  APHU00616  APHU00617  APHU00618  APHU00619  APHU00620  APHU00678  APHU00679  APHU00959  APHU01003  APHU01005  APHU01006  APHU01197  APHU00649  APHU00920  APHU00938  APHU00939  APHU00726  APHU00609  APHU00650  APHU00651  APHU00652  APHU00931  APHU00947  APHU00948  APHU00949  APHU00950  APHU00957  APHU00958  APHU00999  APHU01001  APHU01002  APHU00960  APHU00728  APHU00484  APHU00487  APHU00590  APHU00591  APHU00579  APHU01000  APHU01032  APHU01033  APHU01034  APHU01209  APHU00599  APHU00515  APHU00653  APHU00654  APHU00822  APHU00961  APHU01024  APHU00611  APHU00522  APHU00523  APHU00936  APHU00937  APHU00946  APHU01004  APHU01090  APHU00819  APHU00633  APHU01009  APHU00823  APHU00824  APHU00825  APHU00826  APHU00827  APHU00828  APHU00829  APHU00830  APHU00831  APHU00832  APHU00871  APHU00872  APHU00873  APHU00874  APHU00875  APHU00876  APHU00877  APHU00648  APHU00745  APHU00746  APHU00747  APHU00748  APHU00749  APHU00750  APHU00524  APHU00525  APHU00795  APHU01036  APHU01037  APHU01038  APHU01039  APHU01040  APHU01041  APHU01042  APHU01043  APHU01083  APHU00933  APHU00934  APHU00635  APHU00592  APHU00593  APHU00594  APHU00622  APHU00623  APHU00624  APHU00625  APHU00627  APHU00628  APHU00629  APHU00630  APHU00639  APHU00506  APHU01056  APHU01057  APHU01058  APHU00595  APHU00621  APHU00626  APHU00631  APHU00634  APHU00636  APHU00637  APHU00638  APHU00470  APHU00471  APHU00472  APHU00474  APHU00475  APHU00640  APHU00709  APHU00710  APHU00711  APHU00655  APHU00455  APHU00456  APHU00457  APHU00458  APHU00459  APHU00460  APHU00461  APHU00462  APHU00504  APHU01102  APHU00779  APHU01141  APHU00908  APHU00778  APHU00963  APHU00964  APHU00965  APHU00600  APHU01107  APHU01108  APHU00744  APHU00940  APHU00941  APHU00942  APHU00943  APHU00944  APHU00953  APHU00954  APHU00955  APHU00956  APHU01181  APHU01182  APHU01183  APHU01184  APHU01185  APHU01186  APHU01187  APHU00694  APHU00695  APHU00696  APHU00697  APHU00698  APHU00699  APHU00537  APHU00538  APHU00539  APHU00540  APHU00541  APHU00542  APHU00543  APHU00544  APHU00545  APHU00546  APHU00547  APHU00548  APHU00549  APHU00550  APHU00551  APHU01190  APHU01081  APHU00743  APHU00604  APHU00605  APHU00606  APHU00607  APHU00608  APHU00677  APHU00712  APHU00713  APHU00714  APHU00768  APHU00769  APHU00770  APHU00771  APHU00727  APHU00753  APHU00754  APHU00755  APHU00756  APHU01135  APHU01136  APHU01137  APHU00818  APHU00670  APHU00503  APHU00962  APHU01078  APHU01079  APHU00502  APHU00668  APHU00891  APHU01179  APHU00512  APHU00588  APHU00589  APHU00666  APHU00772  APHU00780  APHU00788  APHU00789  APHU00790  APHU00854  APHU00855  APHU00856  APHU00857  APHU00858  APHU00870  APHU00912  APHU00913  APHU00919  APHU00925  APHU00926  APHU00927  APHU00928  APHU00929  APHU00930  APHU00968  APHU00973  APHU00974  APHU01059  APHU01060  APHU01061  APHU01062  APHU01063  APHU01074  APHU01075  APHU01131  APHU01139  APHU01140  APHU01142  APHU01144  APHU01145  APHU01173  APHU01177  APHU01178  APHU00508  APHU00509  APHU00510  APHU00511  APHU00863  APHU00864  APHU00865  APHU00866  APHU00867  APHU00868  APHU00869  APHU00693  APHU00835  APHU00836  APHU00837  APHU00838  APHU00898  APHU00899  APHU00900  APHU00901  APHU00902  APHU01020  APHU01021  APHU01022  APHU01072  APHU01073  APHU01088  APHU00505  APHU00680  APHU00681  APHU00852  APHU00853  APHU00996  APHU00682  APHU00683  APHU00684  APHU00997  APHU00897  APHU00998  APHU01098  APHU01143  APHU00656  APHU00657  APHU00658  APHU00659  APHU00660  APHU00661  APHU00707  APHU00708  APHU00883  APHU01091  APHU01092  APHU01093  APHU00840  APHU00841  APHU01084  APHU01085  APHU01086  APHU00773  APHU00774  APHU00775  APHU00776  APHU00892  APHU00893  APHU00894  APHU00685  APHU00686  APHU00834  APHU00804  APHU00805  APHU00791  APHU00792  APHU00497  APHU00498  APHU00499  APHU00672  APHU00673  APHU00674  APHU00675  APHU00676  APHU00513  APHU00558  APHU01080  APHU01174  APHU00669  APHU01082  APHU00601  APHU01089  APHU00692  APHU00632  APHU00793  APHU00794  APHU00921  APHU00850  APHU00860  APHU00861  APHU00862  APHU01010  APHU01011  APHU01012  APHU01013  APHU01014  APHU01015  APHU01016  APHU01017  APHU01018  APHU01019  APHU00848  APHU00849  APHU01138  APHU01168  APHU01169  APHU01170  APHU01171  APHU01172  APHU00945  APHU00724  APHU00725  APHU01026  APHU01027  APHU01028  APHU01052  APHU00687  APHU00688  APHU01122  APHU01123  APHU01124  APHU01161  APHU01162  APHU01163  APHU01164  APHU01165  APHU01166  APHU01167  APHU00878  APHU00879  APHU00880  APHU00951  APHU00952  APHU01103  APHU01104  APHU01105  APHU01106  APHU00671  APHU00833  APHU00662  APHU00663  APHU00664  APHU00665  APHU00667  APHU01025  APHU01029  APHU00526  APHU00527  APHU00642  APHU00643  APHU00644  APHU01066  APHU01067  APHU01030  APHU01068  APHU01069  APHU01070  APHU01071  APHU00896  APHU01050  APHU01051  APHU01076  APHU01053  APHU01054  APHU01055",
      "more_code_info": ""
    }
  ]
}