  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Aliso Viejo",
      "state": "CA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63698",
      "recalling_firm": "Bausch and Lomb, Incorporated",
      "address_1": "30 Enterprise Ste 450",
      "address_2": "N/A",
      "postal_code": "92656-7115",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide distribution: USA including DC and PR; and countries of: Canada and Mexico accounts not available at this time.",
      "recall_number": "Z-0672-2013",
      "product_description": "Bausch and Lomb 27G sterile cannula packed in the Bausch and Lomb Amvisc 1.2% Sodium Hyaluronate ( Model 59051, 59081, 59051L, 59081L) and Amvisc Plus 1.6% Sodium Hyaluronate (Model 60081, 60051, 60051L, 60081L)  ophthalmic viscosurgical device.    The 27G disposable cannula is attached to the syringe standard luer fitting to deliver the Amvis or Amvis Plus during use as a surgical aid in opthalmic anterior or posterior segment procedures.",
      "product_quantity": "2,515,760",
      "reason_for_recall": "Some disposable cannulas (with 4-digit lot numbers) provided with the Amvisc or Amvisc Plus OVD are not properly engaging to the Luer-Lock on the Amvisc and Amvisc Plus sterile glass syringes.  The cannulas may leak viscoelastic material or detach from the syringe during injection.  In rare incidents, detachment has resulted in serious patient injury.",
      "recall_initiation_date": "20121115",
      "center_classification_date": "20130117",
      "termination_date": "20131204",
      "report_date": "20130123",
      "code_info": "Lot Numbers:  23632, 23633, 23639, 23640, 23647, 23801, 23811, 23863, 23886, 23897, 23920, 23933, 23972, 24022, 24025, 24028, 24257, 24323, 24436, B100003C, B100003E, B100003G, B100015A, B100015D, B100023A, B100023B, B100023C, B100023D, B100023E, B100039C, B100039D, B100039F, B100046A, B100046B, B100046C, B100058A, B100069A, B100069E, B110027D, B110028C, B110048A, B110048D, B110056A, B120011D, B120016A, B120046C, C100039A, C100039B, C10003B, C100046A, C100046B, C100046C, D090095B, D090096A, D090096C, D090097C, D090097D, D090097E, D090097G, D090105B, D090105C, D090105D, D100007B, D100007C, D100008A, D100008B, D100009A, D100011A, D100011B, D100011C, D100016A, D100016B, D100016C, D100016D, D100022A, D100022B, D100022C, D100022D, D100026A, D100026B, D100026C, D100026D, D100026E, D100030A, D100030B, D100030C, D100035A, D100035B, D100035C, D100035D, D100043A, D100043B, D100043C, D100044A, D100044B, D100044C, D100044D, D100044E, D100045A, D100045B, D100045C, D100045D, D100059A, D100059B, D100059C, D100060A, D100060B, D100061A, D100061B, D100080A, D100081A, D100081AR, D100081AS, D110015A, D110015B, D110016A, D110045A, D110046A, D110047A, D110070A, D110071A, D110072A, D110090A, D110090B, D110091A, D110092A, D110108A, D110108B, D110108BR, D120015A, D120015B, D120027A, D120028A, D120029A, D120029B, D120038B, D120039A, D120056B, D120068A, D120068B, E090105A, E100016A, E100016B, E100035A, E100035B, E100035C, E100045A, E100045B, E100059A, E100059B, E100060A, E100060B, E100060C, E100061A,     Amvisc 1.2% Sodium Hyaluronate ( Model 59051, 59081, 59051L, 59081L)   Amvisc Plus 1.6% Sodium Hyaluronate (Model 60081, 60051, 60051L, 60081L)"
    }
  ]
}