  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Saint Charles",
      "state": "MO",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84330",
      "recalling_firm": "Elekta Inc",
      "address_1": "1450 Beale St Ste 205",
      "address_2": "N/A",
      "postal_code": "63303",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "IA, IN, NY, OH, TX, WI.  Internationally distributed to:  Australia, Canada, China, Denmark, Germany, Greece, Hong Kong, Italy, Netherlands, Sweden, Switzerland, United Kingdom.",
      "recall_number": "Z-0670-2020",
      "product_description": "Monaco Radiation Treatment Planning System (RTP) System",
      "product_quantity": "76 (7 US, 69 OUS)",
      "reason_for_recall": "Monaco is using the incorrect energy when optimizing and calculating dose.",
      "recall_initiation_date": "20191122",
      "center_classification_date": "20191212",
      "report_date": "20191218",
      "code_info": "Software Builds:  5.40.00, 5.50.00 and 5.51.00.    UDI GTIN:  (01)00858164002190(10) 5.40.00,  (01)00858164002244(10) 5.50.00, and  (01)00858164002268(10) 5.51.00",
      "more_code_info": ""
    }
  ]
}