      "status": "Terminated",
      "city": "West Chester",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79021",
      "recalling_firm": "Synthes, Inc.",
      "address_1": "1302 Wrights Ln E",
      "address_2": "N/A",
      "postal_code": "19380-3417",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distribution to US, Canada, and Germany.",
      "recall_number": "Z-0669-2018",
      "product_description": "LCP Drill Sleeve 1.5 Drill Bit 1.1mm  Part Number 03.114.001    Used in LCP Modular Mini Fragment System, for fixation of small bones and small bone fragments",
      "product_quantity": "567 units",
      "reason_for_recall": "May have difficulty attaching the drill sleeve to the plate. If the drill sleeve will not connect to the plate intraoperatively, surgical delay could occur. If the pilot hole is drilled without the drill sleeve, the pilot hole could be off-center which could result in screw and/or plate loosening.",
      "recall_initiation_date": "20170807",
      "center_classification_date": "20180221",
      "termination_date": "20180725",
      "report_date": "20180228",
      "code_info": "Lot numbers  FT00227  9972971  H028400  H161606  H288454"
    }
  ]
}