  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Mountain View",
      "address_1": "685 E Middlefield Rd",
      "reason_for_recall": "Firm received reports about \"acquisition errors\" and the deterioration of material covering the articulating section of the Z6Ms transesophageal transducer.",
      "address_2": "",
      "product_quantity": "33",
      "code_info": "Model number 10436113 -Z6Ms Transesophageal tranducer.  US Serial numbers:  63229007  63229008  63432010  63530518  63322007  63230009  63260584  63530516  63432007  63319012  63332004  63325007  63328003  63328007  63332002  63029015  63329017  63230003.  Foreign serial numbers:  63260585  63230002  63029009  63329018  63229011  63432015  63229012  63430524  63029016  63360611  63360612  63432017.",
      "center_classification_date": "20161128",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) Internationally to AUSTRIA, GERMANY, SOUTH AFRICA, NETHERLANDS, MAURITIUS, MEXICO, and ITALY,  JAPAN, SWEDEN",
      "state": "CA",
      "product_description": "Z6Ms transesophageal transducers used on ACUSON SC2000 ultrasound systems, Z6Ms model 10436113.    Radiology:  The ACUSON SC2000 ultrasound imaging system is intended for the following applications: Cardiac, Neo-natal and Fetal Cardiac, Pediatric, Transesophageal, Adult Cephalic, Peripheral Vessel, Abdominal, Abdominal Intraoperative, Intraoperative Neurological,  Neurological-skeletal Conventional, and Musculo-skeletal Superficial applications. The system  also provides the ability to measure anatomical structures and calculation packages that provide information to the clinician that may be used adjunctive with other medical data obtained by a physician for clinical diagnosis purposes.",
      "report_date": "20161207",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Medical Solutions USA, Inc.",
      "recall_number": "Z-0667-2017",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "75639",
      "termination_date": "20170526",
      "more_code_info": "",
      "recall_initiation_date": "20161107",
      "postal_code": "94043-4045",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}