  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Franklin Lakes",
      "state": "NJ",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89486",
      "recalling_firm": "Becton Dickinson & Company",
      "address_1": "1 Becton Dr",
      "address_2": "",
      "postal_code": "07417-1815",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "US Nationwide distribution in the state of IL.",
      "recall_number": "Z-0665-2022",
      "product_description": "BD Vacutainer Luer Lok Access Device - intended as a sterile, single use device designed to provide a connection to a catheter port equipped with a threaded female luer during blood or urine sample collection using evacuated BD Vacutainer Blood or Urine Collection Tubes  Catalog Number: 364902, 36490200",
      "product_quantity": "288,152,790 Units",
      "reason_for_recall": "May shed contaminant particles into the urine specimen that may be misidentified as Red Blood Cells when used in conjunction with the Sysmex UF-1000i or UX-2000 Analyzers and lead to false elevations in RBC counts",
      "recall_initiation_date": "20220106",
      "center_classification_date": "20220223",
      "report_date": "20220302",
      "code_info": "All Lots within Expiry  Catalog No: 364902    UDI: (01)50382903649025 Catalog Number: 36490200  UDI:(01)60382903649022",
      "more_code_info": ""
    }
  ]
}