  "results": [
    {
      "status": "Ongoing",
      "city": "Carlsbad",
      "state": "CA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89490",
      "recalling_firm": "Respironics California, LLC",
      "address_1": "2271 Cosmos Ct",
      "address_2": "",
      "postal_code": "92011-1517",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U.S.: AR, CA, FL, GA, IN, LA, MD, MO, NY, PA, SC, TN and TX.    O.U.S.  N/A",
      "recall_number": "Z-0663-2022",
      "product_description": "Philips Respironics V60 Plus Ventilator Part Number 1138747",
      "product_quantity": "294 systems",
      "reason_for_recall": "A limited number of ventilators were assembled with expired adhesive. If the adhesive fails, a bracket may become loose and potentially damage the capacitors, which could cause the ventilator to stop providing ventilation to the patient. This failure may either activate both visual and audible alarms, or it may not sound or display an alarm (a silent shutdown).",
      "recall_initiation_date": "20220124",
      "center_classification_date": "20220225",
      "report_date": "20220309",
      "code_info": "The Component of the Ventilators that is impacted is: N 1034132 BRACKET,  CAPACITOR,MOTOR DRIVER PCB    Model V60  Part Number: 1138747  Serial Numbers/UDI Code:  MX0016240/   (01)00884838090996(21)MX0016240  MX0016241/   (01)00884838090996(21)MX0016241  MX0016242/   (01)00884838090996(21)MX0016242  MX0016243/   (01)00884838090996(21)MX0016243  MX0016244/   (01)00884838090996(21)MX0016244  MX0016245/   (01)00884838090996(21)MX0016245  MX0016246/ (01)00884838090996(21)MX0016246  MX0016247/ (01)00884838090996(21)MX0016247  MX0016248/ (01)00884838090996(21)MX0016248  MX0016522/ (01)00884838090996(21)MX0016522  MX0016523/ (01)00884838090996(21)MX0016523  MX0016525/ (01)00884838090996(21)MX0016525  MX0016527/ (01)00884838090996(21)MX0016527  MX0016529/ (01)00884838090996(21)MX0016529  MX0016531/ (01)00884838090996(21)MX0016531  MX0016532/ (01)00884838090996(21)MX0016532  MX0016533/ (01)00884838090996(21)MX0016533  MX0016534/ (01)00884838090996(21)MX0016534  MX0016535/ (01)00884838090996(21)MX0016535  MX0016536/ (01)00884838090996(21)MX0016536  MX0016538/ (01)00884838090996(21)MX0016538  MX0016539/ (01)00884838090996(21)MX0016539  MX0016540/ (01)00884838090996(21)MX0016540  MX0016541/ (01)00884838090996(21)MX0016541  MX0016542/ (01)00884838090996(21)MX0016542  MX0016545/ (01)00884838090996(21)MX0016545  MX0016546/ (01)00884838090996(21)MX0016546  MX0016547/ (01)00884838090996(21)MX0016547  MX0016548/ (01)00884838090996(21)MX0016548  MX0016549/ (01)00884838090996(21)MX0016549  MX0016550/ (01)00884838090996(21)MX0016550  MX0016552/ (01)00884838090996(21)MX0016552  MX0016553/ (01)00884838090996(21)MX0016553  MX0016554/ (01)00884838090996(21)MX0016554  MX0016556/ (01)00884838090996(21)MX0016556  MX0016559/ (01)00884838090996(21)MX0016559  MX0016560/ (01)00884838090996(21)MX0016560  MX0016561/ (01)00884838090996(21)MX0016561  MX0016562/ (01)00884838090996(21)MX0016562  MX0016568/ (01)00884838090996(21)MX0016568  MX0016569/ (01)00884838090996(21)MX0016569  MX0016570/ (01)00884838090996(21)MX0016570  MX0016571/ (01)00884838090996(21)MX0016571  MX0016572/ (01)00884838090996(21)MX0016572  MX0016573/ (01)00884838090996(21)MX0016573  MX0016574/ (01)00884838090996(21)MX0016574  MX0016575/ (01)00884838090996(21)MX0016575  MX0016576/ (01)00884838090996(21)MX0016576  MX0016577/ (01)00884838090996(21)MX0016577  MX0016578/ (01)00884838090996(21)MX0016578  MX0016580/ (01)00884838090996(21)MX0016580  MX0016581/ (01)00884838090996(21)MX0016581  MX0016583/ (01)00884838090996(21)MX0016583  MX0016584/ (01)00884838090996(21)MX0016584  MX0016807/ (01)00884838090996(21)MX0016807  MX0016813/ (01)00884838090996(21)MX0016813  MX0016818/ (01)00884838090996(21)MX0016818  MX0016838/ (01)00884838090996(21)MX0016838  MX0016850/ (01)00884838090996(21)MX0016850  MX0016860/ (01)00884838090996(21)MX0016860  MX0016868/ (01)00884838090996(21)MX0016868  MX0016873/ (01)00884838090996(21)MX0016873  MX0016880/ (01)00884838090996(21)MX0016880  MX0016881/ (01)00884838090996(21)MX0016881  MX0016882/ (01)00884838090996(21)MX0016882  MX0016883/ (01)00884838090996(21)MX0016883  MX0016884/ (01)00884838090996(21)MX0016884  MX0016886/ (01)00884838090996(21)MX0016886  MX0016887/ (01)00884838090996(21)MX0016887  MX0016989/ (01)00884838090996(21)MX0016989  MX0016990/ (/01)00884838090996(21)MX0016990  MX0016991/ (01)00884838090996(21)MX0016991  MX0016992/ (01)00884838090996(21)MX0016992  MX0016993/ (01)00884838090996(21)MX0016993  MX0016994/ (01)00884838090996(21)MX0016994  MX0016995/ (01)00884838090996(21)MX0016995  MX0016996/ (01)00884838090996(21)MX0016996  MX0016997/ (01)00884838090996(21)MX0016997  MX0016998/ (01)00884838090996(21)MX0016998  MX0016999/ (01)00884838090996(21)MX0016999  MX0017000/ (01)00884838090996(21)MX0017000  MX0017001/ (01)00884838090996(21)MX0017001  MX0017003/ (01)00884838090996(21)MX0017003  MX0017004/ (01)00884838090996(21)MX0017004  MX0017005/ (01)00884838090996(21)MX0017005  MX0017006/ (01)00884838090996(21)MX0017006  MX0017007/ (01)00884838090996(21)MX0017007  MX0017008/ (01)00884838090996(21)MX0017008  MX0017009/ (01)00884838090996(21)MX0017009  MX0017010/ (01)00884838090996(21)MX0017010  MX0017011/ (01)00884838090996(21)MX0017011  MX0017012/ (01)00884838090996(21)MX0017012  MX0017013/ (01)00884838090996(21)MX0017013  MX0017014/ (01)00884838090996(21)MX0017014  MX0017015/ (01)00884838090996(21)MX0017015  MX0017016/ (01)00884838090996(21)MX0017016  MX0017017/ (01)00884838090996(21)MX0017017  MX0017019/ (01)00884838090996(21)MX0017019  MX0017020/ (01)00884838090996(21)MX0017020  MX0017021/ (01)00884838090996(21)MX0017021  MX0017022/ (01)00884838090996(21)MX0017022  MX0017023/ (01)00884838090996(21)MX0017023  MX0017024/ (01)00884838090996(21)MX0017024  MX0017025/ (01)00884838090996(21)MX0017025  MX0017026/ (01)00884838090996(21)MX0017026  MX0017405/ (01)00884838090996(21)MX0017405  MX0017406/ (01)00884838090996(21)MX0017406  MX0017407/ (01)00884838090996(21)MX0017407  MX0017408/ (01)00884838090996(21)MX0017408  MX0017409/ (01)00884838090996(21)MX0017409  MX0017410/ (01)00884838090996(21)MX0017410  MX0017411/ (01)00884838090996(21)MX0017411  MX0017412/ (01)00884838090996(21)MX0017412  MX0017413/ (01)00884838090996(21)MX0017413  MX0017418/ (01)00884838090996(21)MX0017418  MX0017426/ (01)00884838090996(21)MX0017426  MX0017427/ (01)00884838090996(21)MX0017427  MX0017428/ (01)00884838090996(21)MX0017428  MX0017429/ (01)00884838090996(21)MX0017429  MX0017431/ (01)00884838090996(21)MX0017431  MX0017432/ (01)00884838090996(21)MX0017432  MX0017433/ (01)00884838090996(21)MX0017433  MX0017434/ (01)00884838090996(21)MX0017434  MX0017435/ (01)00884838090996(21)MX0017435  MX0017436/ (01)00884838090996(21)MX0017436  MX0017437/ (01)00884838090996(21)MX0017437  MX0017438/ (01)00884838090996(21)MX0017438  MX0017439/ (01)00884838090996(21)MX0017439  MX0017442/ (01)00884838090996(21)MX0017442  MX0017443/ (01)00884838090996(21)MX0017443  MX0017444/ (01)00884838090996(21)MX0017444  MX0017445/ (01)00884838090996(21)MX0017445  MX0017446/ (01)00884838090996(21)MX0017446  MX0017447/ (01)00884838090996(21)MX0017447  MX0017448/ (01)00884838090996(21)MX0017448  MX0017449/ (01)00884838090996(21)MX0017449  MX0017451/ (01)00884838090996(21)MX0017451  MX0017452/ (01)00884838090996(21)MX0017452  MX0017453/ (01)00884838090996(21)MX0017453  MX0017454/ (01)00884838090996(21)MX0017454  MX0017455/ (01)00884838090996(21)MX0017455  MX0017456/ (01)00884838090996(21)MX0017456  MX0017457/ (01)00884838090996(21)MX0017457  MX0017458/ (01)00884838090996(21)MX0017458  MX0017459/ (01)00884838090996(21)MX0017459  MX0017460/ (01)00884838090996(21)MX0017460  MX0017461/ (01)00884838090996(21)MX0017461  MX0017463/ (01)00884838090996(21)MX0017463  MX0017465/ (01)00884838090996(21)MX0017465  MX0017466/ (01)00884838090996(21)MX0017466  MX0017467/ (01)00884838090996(21)MX0017467  MX0017468/ (01)00884838090996(21)MX0017468  MX0017470/ (01)00884838090996(21)MX0017470  MX0017471/ (01)00884838090996(21)MX0017471  MX0017473/ (01)00884838090996(21)MX0017473  MX0017475/ (01)00884838090996(21)MX0017475  MX0017476/ (01)00884838090996(21)MX0017476  MX0017477/ (01)00884838090996(21)MX0017477  MX0017478/ (01)00884838090996(21)MX0017478  MX0017480/ (01)00884838090996(21)MX0017480  MX0017481/ (01)00884838090996(21)MX0017481  MX0017482/ (01)00884838090996(21)MX0017482  MX0017483/ (01)00884838090996(21)MX0017483  MX0017484/ (01)00884838090996(21)MX0017484  MX0017485/ (01)00884838090996(21)MX0017485  MX0017486/ (01)00884838090996(21)MX0017486  MX0017487/ (01)00884838090996(21)MX0017487  MX0017488/ (01)00884838090996(21)MX0017488  MX0017490/ (01)00884838090996(21)MX0017490  MX0017491/ (01)00884838090996(21)MX0017491  MX0017492/ (01)00884838090996(21)MX0017492  MX0017493/ (01)00884838090996(21)MX0017493  MX0017494/ (01)00884838090996(21)MX0017494  MX0017495/ (01)00884838090996(21)MX0017495  MX0017496/ (01)00884838090996(21)MX0017496  MX0017497/ (01)00884838090996(21)MX0017497  MX0017498/ (01)00884838090996(21)MX0017498  MX0017499/ (01)00884838090996(21)MX0017499  MX0017500/ (01)00884838090996(21)MX0017500  MX0017602/ (01)00884838090996(21)MX0017602  MX0017605/ (01)00884838090996(21)MX0017605  MX0017607/ (01)00884838090996(21)MX0017607  MX0017608/ (01)00884838090996(21)MX0017608  MX0017609/ (01)00884838090996(21)MX0017609  MX0017610/ (01)00884838090996(21)MX0017610  MX0017611/ (01)00884838090996(21)MX0017611  MX0017612/ (01)00884838090996(21)MX0017612  MX0017613/ (01)00884838090996(21)MX0017613  MX0017614/ (01)00884838090996(21)MX0017614  MX0017615/ (01)00884838090996(21)MX0017615  MX0017616/ (01)00884838090996(21)MX0017616  MX0017617/ (01)00884838090996(21)MX0017617  MX0017618/ (01)00884838090996(21)MX0017618  MX0017619/ (01)00884838090996(21)MX0017619  MX0017620/ (01)00884838090996(21)MX0017620  MX0017622/ (01)00884838090996(21)MX0017622  MX0017623/ (01)00884838090996(21)MX0017623  MX0017625/ (01)00884838090996(21)MX0017625  MX0017628/ (01)00884838090996(21)MX0017628  MX0017632/ (01)00884838090996(21)MX0017632  MX0017633/ (01)00884838090996(21)MX0017633  MX0017635/ (01)00884838090996(21)MX0017635  MX0017637/ (01)00884838090996(21)MX0017637  MX0017639/ (01)00884838090996(21)MX0017639  MX0017645/ (01)00884838090996(21)MX0017645  MX0017646/ (01)00884838090996(21)MX0017646  MX0017647/ (01)00884838090996(21)MX0017647  MX0017649/ (01)00884838090996(21)MX0017649  MX0017650/ (01)00884838090996(21)MX0017650  MX0017651/ (01)00884838090996(21)MX0017651  MX0017652/ (01)00884838090996(21)MX0017652  MX0017653/ (01)00884838090996(21)MX0017653  MX0017657/ (01)00884838090996(21)MX0017657  MX0017659/ (01)00884838090996(21)MX0017659  MX0017660/ (01)00884838090996(21)MX0017660  MX0017661/ (01)00884838090996(21)MX0017661  MX0017662/ (01)00884838090996(21)MX0017662  MX0017663/ (01)00884838090996(21)MX0017663  MX0017665/ (01)00884838090996(21)MX0017665  MX0017670/ (01)00884838090996(21)MX0017670  MX0017672/ (01)00884838090996(21)MX0017672  MX0017675/ (01)00884838090996(21)MX0017675  MX0017676/ (01)00884838090996(21)MX0017676  MX0017677/ (01)00884838090996(21)MX0017677  MX0017678/ (01)00884838090996(21)MX0017678  MX0017682/ (01)00884838090996(21)MX0017682  MX0017683/ (01)00884838090996(21)MX0017683  MX0017684/ (01)00884838090996(21)MX0017684  MX0017685/ (01)00884838090996(21)MX0017685  MX0017686/ (01)00884838090996(21)MX0017686  MX0017687/ (01)00884838090996(21)MX0017687  MX0017688/ (01)00884838090996(21)MX0017688  MX0017689/ (01)00884838090996(21)MX0017689  MX0017690/ (01)00884838090996(21)MX0017690  MX0017691/ (01)00884838090996(21)MX0017691  MX0017692/ (01)00884838090996(21)MX0017692  MX0017822/ (01)00884838090996(21)MX0017822  MX0017823/ (01)00884838090996(21)MX0017823  MX0017824/ (01)00884838090996(21)MX0017824  MX0017826/ (01)00884838090996(21)MX0017826  MX0017827/ (01)00884838090996(21)MX0017827  MX0017830/ (01)00884838090996(21)MX0017830  MX0017831/ (01)00884838090996(21)MX0017831  MX0017833/ (01)00884838090996(21)MX0017833  MX0017837/ (01)00884838090996(21)MX0017837  MX0017843/ (01)00884838090996(21)MX0017843  MX0017845/ (01)00884838090996(21)MX0017845  MX0017848/ (01)00884838090996(21)MX0017848  MX0017850/ (01)00884838090996(21)MX0017850  MX0017856/ (01)00884838090996(21)MX0017856  MX0017857/ (01)00884838090996(21)MX0017857  MX0017858/ (01)00884838090996(21)MX0017858  MX0017859/ (01)00884838090996(21)MX0017859  MX0017860/ (01)00884838090996(21)MX0017860  MX0017862/ (01)00884838090996(21)MX0017862  MX0017863/ (01)00884838090996(21)MX0017863  MX0017864/ (01)00884838090996(21)MX0017864  MX0017865/ (01)00884838090996(21)MX0017865  MX0017866/ (01)00884838090996(21)MX0017866  MX0017867/ (01)00884838090996(21)MX0017867  MX0017868/ (01)00884838090996(21)MX0017868  MX0017870/ (01)00884838090996(21)MX0017870  MX0017874/ (01)00884838090996(21)MX0017874  MX0017885/ (01)00884838090996(21)MX0017885  MX0017891/ (01)00884838090996(21)MX0017891  MX0017895/ (01)00884838090996(21)MX0017895  MX0017896/ (01)00884838090996(21)MX0017896  MX0017897/ (01)00884838090996(21)MX0017897  MX0017899/ (01)00884838090996(21)MX0017899  MX0017901/ (01)00884838090996(21)MX0017901  MX0017910/ (01)00884838090996(21)MX0017910  MX0017912/ (01)00884838090996(21)MX0017912  MX0017914/ (01)00884838090996(21)MX0017914  MX0018141/ (01)00884838090996(21)MX0018141  MX0018143/ (01)00884838090996(21)MX0018143  MX0018144/ (01)00884838090996(21)MX0018144  MX0018148/ (01)00884838090996(21)MX0018148  MX0018150/ (01)00884838090996(21)MX0018150  MX0018151/ (01)00884838090996(21)MX0018151  MX0018154/ (01)00884838090996(21)MX0018154  MX0018155/ (01)00884838090996(21)MX0018155  MX0018159/ (01)00884838090996(21)MX0018159  MX0018179/ (01)00884838090996(21)MX0018179  MX0018185/ (01)00884838090996(21)MX0018185  MX0018192/ (01)00884838090996(21)MX0018192  MX0018193/ (01)00884838090996(21)MX0018193  MX0018197/ (01)00884838090996(21)MX0018197  MX0018199/ (01)00884838090996(21)MX0018199  MX0018202/ (01)00884838090996(21)MX0018202  MX0018204/ (01)00884838090996(21)MX0018204  MX0018209/ (01)00884838090996(21)MX0018209  MX0018210/ (01)00884838090996(21)MX0018210  MX0018222/ (01)00884838090996(21)MX0018222  MX0018224/ (01)00884838090996(21)MX0018224",
      "more_code_info": ""
    }
  ]
}