  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Carlsbad",
      "state": "CA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89490",
      "recalling_firm": "Respironics California, LLC",
      "address_1": "2271 Cosmos Ct",
      "address_2": "",
      "postal_code": "92011-1517",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U.S.: AR, CA, FL, GA, IN, LA, MD, MO, NY, PA, SC, TN and TX.    O.U.S.  N/A",
      "recall_number": "Z-0662-2022",
      "product_description": "Philips Respironics V60 Ventilator Part Number 1053617",
      "product_quantity": "16 systems",
      "reason_for_recall": "A limited number of ventilators were assembled with expired adhesive. If the adhesive fails, a bracket may become loose and potentially damage the capacitors, which could cause the ventilator to stop providing ventilation to the patient. This failure may either activate both visual and audible alarms, or it may not sound or display an alarm (a silent shutdown).",
      "recall_initiation_date": "20220124",
      "center_classification_date": "20220225",
      "report_date": "20220309",
      "code_info": "The Component of the Ventilators that is impacted is: N 1034132 BRACKET,  CAPACITOR,MOTOR DRIVER PCB    Model V60  Part Number: 1053617  Serial Numbers/UDI Code:  MX0016593 / (01)00884838020054(21)MX0016593  MX0016797 / (01)00884838020054(21)MX0016797  MX0016824 / (01)00884838020054(21)MX0016824  MX0016827 / (01)00884838020054(21)MX0016827  MX0016830 / (01)00884838020054(21)MX0016830  MX0016839 / (01)00884838020054(21)MX0016839  MX0016840 / (01)00884838020054(21)MX0016840  MX0016841 / (01)00884838020054(21)MX0016841  MX0016845 (01)00884838020054(21)MX0016845  MX0016853 (01)00884838020054(21)MX0016853  MX0016867 (01)00884838020054(21)MX0016867  MX0016870 (01)00884838020054(21)MX0016870  MX0016871 (01)00884838020054(21)MX0016871  MX0016872 (01)00884838020054(21)MX0016872  MX0016875 (01)00884838020054(21)MX0016875  MX0016878 (01)00884838020054(21)MX0016878",
      "more_code_info": ""
    }
  ]
}