  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Oakville",
      "state": "",
      "country": "Canada",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89425",
      "recalling_firm": "Natus Neurology DBA Excel Tech., Ltd. (XLTEK)",
      "address_1": "2568 Bristol Cir",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of FL, IL, TX, MI, MA, NE, OK, WV, MO, VA, CT, IN, HI, PA, CA, MD, OH, CO, SC, PR, NV, NJ, OR, LA, DC, NC and the countries of Nepal, Viet Nam, Canada, Saudi Arabia, Indonesia, Germany, Korea, Republic of",
      "recall_number": "Z-0661-2022",
      "product_description": "IPELA FHD PTZ TILE KITS with Axis cameras used in combination with Natus Neuroworks/Sleepworks software",
      "product_quantity": "221",
      "reason_for_recall": "Tile kits that allow one to install and situate a video camera within a patient environment are not fitting properly in some domestic institution ceiling tile grids that have a thinner style grid frame. The ceiling tile may suddenly fall from the ceiling and cause a safety risk to the patient or end user.",
      "recall_initiation_date": "20211202",
      "center_classification_date": "20220222",
      "termination_date": "20240918",
      "report_date": "20220302",
      "code_info": "All serial numbers",
      "more_code_info": ""
    }
  ]
}