  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Irvine",
      "address_1": "33 Technology Dr",
      "reason_for_recall": "Advanced Sterilization Products is recalling three lots of STERRAD NX System Cassettes because it does not have adequate data to support the entire duration of the product's labeled shelf-life.",
      "address_2": "",
      "product_quantity": "12,280 units",
      "code_info": "Lot # 11J037, 11L038, 11J033",
      "center_classification_date": "20130109",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and internationally to the following countries:  AR, CO, TW, SI, AU, MX, MY, HU, PH, BE, FR, DE, IE, IT, ES, NO, CH, and GB.",
      "state": "CA",
      "product_description": "STERRAD NX Cassettes, P/N 10133.    The STERRAD NX sterilizations systems are low-temperature general purpose sterilizers used to sterilize heat and moisture sensitive reusable medical devices. The sterilant used is hydrogen peroxide supplied in cassette form as a separate accessory. The STERRAD NX Cassettes are only used in conjunction with the STERRAD NX Sterilization system.",
      "report_date": "20130116",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Advanced Sterilization Products",
      "recall_number": "Z-0661-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "63877",
      "termination_date": "20131017",
      "more_code_info": "",
      "recall_initiation_date": "20120920",
      "postal_code": "92618-2346",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}