  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Warsaw",
      "address_1": "56 E Bell Dr",
      "reason_for_recall": "Biomet is recalling Part Number PT-113950 PT Hybrid Glenoid Post,  following an investigation which identified that the male thread of the post may be oversized. This oversized condition can vary in degree and may lead to the following three events:  1) If the PT Hybrid Glenoid Post is not fully seated into the Hybrid Base and the implant construct is  implanted, then a gap of 1-3 mm will be prese",
      "address_2": "",
      "product_quantity": "1050",
      "code_info": "Part 113950  Lot 093540, 093550,093570,093580,093590,093600,093610,093620,093720, 093730,093740,093760,093770,093780,282590,282600,282620,282630,  282680,282700,282710,489160,489180,489200,489210,489770,489780,  489790,714480,714550,714570,714700,714730,722210,722250,722280,  984770,984780,984790,984800,984810,984820,984830,984840, and 984850",
      "center_classification_date": "20130109",
      "distribution_pattern": "Worldwide Distribution -- US (nationwide) including the countries of , Canada, Japan, Argentina, Australia ,Netherlands, Europe and Chile.",
      "state": "IN",
      "product_description": "PT113950 PT HYBRID Glenoid Post  Regenerex porous titanium construct  Sterile R    Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented",
      "report_date": "20130116",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Biomet, Inc.",
      "recall_number": "Z-0660-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "63850",
      "termination_date": "20131108",
      "more_code_info": "",
      "recall_initiation_date": "20061212",
      "postal_code": "46582-6989",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}