      "status": "Ongoing",
      "city": "Louisville",
      "state": "CO",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84191",
      "recalling_firm": "Medtronic Navigation, Inc.",
      "address_1": "826 Coal Creek Cir",
      "address_2": "N/A",
      "postal_code": "80027-9710",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: Alabama, Alaska,Arizona,Arkansas,California,Colorado,Connecticut,Delaware,District of Columbia,Florida,Georga, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine,Maryland,Massachusetts,Michigan,Minnesota,Mississippi,Missouri,Montana,Nebraska,Nevada,New Hampshire,New Jersey,New Mexico,New York,North Carolina,North Dakota,Ohio,Oklahoma,Oregon,Pennsylvania,South Carolina,South Dakota,Tennessee,Texas,Utah,Vermont,Virginia,Washington,West Virginia,Wisconsin,Wyoming,    OUS: None: Australia, Austria,Bangladesh,Belgium,Bosnia And Herzegovina,Brazil,Bulgaria,Canada,Canary Islands,China,Colombia,Costa Rica,Croatia,Czech Republic,Denmark,Ecuador,Egypt,Finland,France,Germany,Greece,Hungary,Iceland,India,Indonesia,Ireland,Israel,Italy,Japan,Jordan,Kazakhstan,Kenya,Korea, Republic Of,Latvia,Lebanon,Luxembourg,Malaysia,Martinique,Mexico,Netherlands,New Zealand,Pakistan,Philippines,Poland,Portugal,Puerto Rico,Qatar,Reunion,Romania,Russian Federation,Saudi Arabia,Serbia,Slovakia,Slovenia,South Africa,Spain,Sweden,Switzerland,Thailand,Turkey,United Arab Emirates,United Kingdom",
      "recall_number": "Z-0659-2020",
      "product_description": "Biopsy Needle FPU Kit, Product # 9736075,  UDI: 00763000063108",
      "product_quantity": "1 disposable Biopsy Needle",
      "reason_for_recall": "The firm has become aware that due to a manufacturing issue impacting the biopsy needle depth stop,  there exist the potential that some biopsy needle depth stops, when tightened, may not securely tighten to the biopsy needle.  This could result in a delay in surgery.",
      "recall_initiation_date": "20191030",
      "center_classification_date": "20191206",
      "report_date": "20191218",
      "code_info": "Lot Numbers:  0009754921  0009921274  0009940846  0009921273  0009921276  0009952564",
      "more_code_info": ""
    }
  ]
}