      "status": "Terminated",
      "city": "Houston",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72896",
      "recalling_firm": "Cyberonics, Inc",
      "address_1": "100 Cyberonics Blvd",
      "address_2": "N/A",
      "postal_code": "77058-2069",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution -- United States, Austria, Belgium, Croatia, Czech Republic, Finland, France, Germany, Iceland, Italy, Netherlands, Norway, Poland, Slovakia, Spain, Sweden, Switzerland, United Kingdom, Cyprus, Israel, Kuwait, Lebanon, Oman, Qatar, Saudi Arabia, and United Arab Emirates.",
      "recall_number": "Z-0659-2016",
      "product_description": "VNS Therapy AspireSR Generator Model 106.    Indicated for use as an adjunctive therapy in reducing the frequency of seizures.",
      "product_quantity": "4,935 units",
      "reason_for_recall": "Recall being initiated in response to three reports of \"Burst Watchdog Timeout\" events occurring with the Model 106 AspireSR Generator, resulting in a device reset condition where stimulation output is disabled.",
      "recall_initiation_date": "20151218",
      "center_classification_date": "20160115",
      "termination_date": "20170302",
      "report_date": "20160127",
      "code_info": "All VNS Therapy AspireSR (Model 106) Generators; Device Identifier - (01)05425025750061",
      "more_code_info": ""
    }
  ]
}