  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87026",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355",
      "voluntary_mandated": "FDA Mandated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide-US Nationwide",
      "recall_number": "Z-0656-2021",
      "product_description": "Artis zee/ zeego & Artis Q/ Q.zen  Artis Pheno",
      "product_quantity": "5602",
      "reason_for_recall": "When the user changes frame rates from lower frequency (e.g. 10 f/s) to higher frequency (e.g. 30 f/s) during continuous release of Fluoro \"Fluoro Override\", the measured Air Kerma Rate can exceed the regulatory limits.",
      "recall_initiation_date": "20201130",
      "center_classification_date": "20210111",
      "report_date": "20210120",
      "code_info": "Model number:  Artis zee floor- 10094135  Artis zee ceiling- 10094137  Artis zee multi-purpose- 10094139  Artis zee biplane- 10094141  Artis zeego- 10280959  Artis Pheno- 10849000  Artis Q floor- 10848280  Artis Q ceiling- 10848281  Artis Q biplane- 10848282  Artis Q zeego- 10848283  Artis Q.zen floor- 10848353  Artis Q.zen ceiling- 10848354  Artis Q.zen biplane- 10848355"
    }
  ]
}