{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Palm Beach Gardens",
      "address_1": "4500 Riverside Drive",
      "reason_for_recall": "The firm Anspach Effort of Palm Beach Gardens, FL initiated a voluntary removal of the Pneumatic Drill System, including the motor hand-piece, Auto Lube foot control, without a pressure relief valve (PRV) due to the hand piece's outer hose can rupture without warning if the air flow is blocked during use..",
      "address_2": "",
      "product_quantity": "1,224",
      "code_info": "Serial Number MM900",
      "center_classification_date": "20140109",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) including Puerto Rico and states of: AL, AR, AZ, CA, CO, CT, DC, DE, GA, FL, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, VA, VE, VT, WA, WI, WY, and WV. Internationally to: AT, AU, BR, BS, CA, CH, CL, CO, CZ, CR, DE, DK, DO, EG, ES, FI, FR, GB, GR, HK, HU, ID, IL, IN, IS, JP, KH, KR, LB, LT, MX, MY, NL, NO, NZ, PA,  PE, PH, PL, PT, SA, SG, SY, TH, TR, TW, and ZA.",
      "state": "FL",
      "product_description": "***REF MICROMAX***microMax Handpiece***Rx Only***Manufacturer: The Anspach Effort, Inc.   Intended for cutting and shaping bone.",
      "report_date": "20140115",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "The Anspach Effort, Inc.",
      "recall_number": "Z-0656-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "66323",
      "termination_date": "20150720",
      "more_code_info": "",
      "recall_initiation_date": "20130425",
      "postal_code": "33410-4235",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}