  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Grand Rapids",
      "state": "MI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72897",
      "recalling_firm": "Skytron, Div. The KMW Group, Inc",
      "address_1": "5085 Corporate Exchange Blvd Se",
      "address_2": "N/A",
      "postal_code": "49512-5515",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US including WI, TX, FL, NC, MD, OH, AZ, CA, UT, MT, OK, HI, CO, WY, MO, NJ, IL, NY, MA, IA, DE, PA, MI, IN, and WV  and Internationally to Canada, and Japan.",
      "recall_number": "Z-0655-2016",
      "product_description": "Ergon 2 Articulating Equipment Boom Monitor Brackets    mounting point for display interfaces utilized in clinical applications for viewing purposes and to provide an articulating point of movement on the ceiling mounted radial arm suspension system of the Ergon 2 equipment pendant / booms",
      "product_quantity": "286",
      "reason_for_recall": "The hardware that holds the VESA plate and the display interface / monitor to the bracket may fail causing the display interface to fall from the bracket. Over time and with use the threaded rod hardware material can weaken and detach from the bracket.  A display interface falling from its mounting location may result in injury to the patient or provider.",
      "recall_initiation_date": "20151202",
      "center_classification_date": "20160114",
      "termination_date": "20160719",
      "report_date": "20160120",
      "code_info": "Model Number(s): H8-010-50-FC2-LH and H8-010-50-FC2-RH.  VESA plates distributed from 09/26/08 to 04/30/13."
    }
  ]
}