  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Torrance",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89253",
      "recalling_firm": "Shimadzu Medical Systems",
      "address_1": "20101 S Vermont Ave",
      "address_2": "",
      "postal_code": "90502-1328",
      "voluntary_mandated": "FDA Mandated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U.S.: CA, FL, GA, IL, IN, KY, LA, MI, MS, OH, PA, SC, TN, TX and WA     O.U.S.: Not provided.",
      "recall_number": "Z-0654-2022",
      "product_description": "Digital Angiography System Bransist safire    Generator Model D150GC-40",
      "product_quantity": "18",
      "reason_for_recall": "Due to the inadequate adjusting criteria in installation, the system generator on two Digital Angiography Systems, the irradiated x-ray may exceed the x-ray radiation dose rate in fluoroscopic mode.",
      "recall_initiation_date": "20211216",
      "center_classification_date": "20220227",
      "report_date": "20220309",
      "code_info": "Generator Model: D150GC-40    Serial Number Range - 0161Q81103 to 3M99D3A55001  Specific Generator Serial Number/SMS ID:  3M99D3A24001/SMS20120038  3M99D3A4B001/SMS20140084  3M99D3A55001/SMS20150030  0161Q83701/SMS20080542  0161Q82204/SMS20080045  0161Q82302/SMS20080043  0161Q82304/SMS20080046  0261Q80901/SMS20100011  3M7A1600B001/SMS20080286  3M99D2622001/SMSMA110079  0161Q84501/SMS20090038  3M7A1600C001/SMS20200285  3M7A16A1A001/SMSMA110013  3M99D3A1C001/SMS20110202  0161Q84103/SMS20090099  0161Q83603/SMS20080505  0161Q83702/SMS20080580  0161Q83703/SMS20080581",
      "more_code_info": ""
    }
  ]
}