  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "San Jose",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69838",
      "recalling_firm": "Stryker Endoscopy",
      "address_1": "5900 Optical Ct",
      "address_2": "N/A",
      "postal_code": "95138-1400",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.",
      "recall_number": "Z-0652-2015",
      "product_description": "PKG, MARYLAND FORCEPS, CURVED LEFT, P/N 0250080282.   Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.",
      "product_quantity": "180,573 units",
      "reason_for_recall": "The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods",
      "recall_initiation_date": "20141117",
      "center_classification_date": "20141217",
      "termination_date": "20150717",
      "report_date": "20141224",
      "code_info": "43967  140297  246217  340634  341506  341508  344540  344549  444519  05242012E  542134  641033  641039  646904  646920  646927  646929  647553  741526  747007  747010  747012  747017  747023  747024  846041  846044  846048  846060  846061  846062  846080  941874  941876  941877  941878  941879  941880  941881  941885  941886  941887  941888  941889  941890  941894  941895  941897  941898  941899  941900  941904  941905  941906  942385  945570  945571  945572  945573  945574  945575  945576  945580  945581  945582  945585  945586  945587  945591  945592  945594  945595  945596  945597  945601  945602  945603  1045189  1045364  1045365  1045366  1045675  1045676  1045677  1045678  1045679  1046174  1046612  1046613  1140665  1140666  1140667  1140668  1140669  1141746  1141747  1142170  1142171  1142172  1142185  1142186  1142187  1142631  1142632  1143459  1143460  1143461  1143462  1143463  1143464  1143465  1144713  1144714  1144715  1144716  1144717  1144721  1145476  1145477  1146158  1146159  1146379  1146380  1146381  1146382  1146383  1146705  1146706  1146707  1146708  1146841  1146842  1147049  1147050  1147164  1147165  1147166  1147167  1147168  1147169  1147170  1240961  1240962  1241086  1241087  1241088  1241090  1241401  1241779  1241780  1241781  1241782  1241783  1241784  1241785  1241786  1241787  1241788  1242411  1242412  1243160  1243161  1245033  1245034  1245035  1245036  1245455  1245478  1245479  1245970  1245971  1245972  1340137  1340138  1340139  1340140  1340141  1340142  1340143  1340144  1340718  1340719  1340720  1341826  1341827  1341828  1341829  1341830  1341831  1341832  1342203  1342845  1342846  1343264  1343526  1343528  1343529  1343531  1344420  1440243  1440244  1440286  1442301  1443206  1443602  163699  641047  846044  941880"
    }
  ]
}