  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Sunnyvale",
      "address_1": "1310 Moffett Park Dr",
      "reason_for_recall": "Potential for unexpected heating of the transformer mounting bolt located within the system chassis and the subsequent melting of a non-slip computer shelf mat, which may manifest an odor and visible smoke.",
      "address_2": "",
      "product_quantity": "78 (39 in US, 39 internationally)",
      "code_info": "All units shipped prior to Octotber 18, 2013.  US Serial numbers:  44055311,  44055611,  44055811,  44056012,  44056312,  44056612,  44056712,   44056812,  44057312,  44057812,  44058012,  44058212,  44058312,  44059212,  44059312,  44059612,  44059712,  44059812,  44059912,  44060012,  44060112,  44060312,  44060412.    International Serial numbers:  44000811,  44055211,  44055411,  44055711,  44055911,  44056112,  44056212,  44056412,  44056512,  44056912,  44057012,  44057112,  44057212,  44057412,  44057512,  44057612,  44057712,  44057912,  44058112,  44058512,  44058612,  44058912,  44059012.  44059112  44059412  44059512  44060212  44060512  44061112  44061212  44061312  44061512  44061712  44061913  44062113  44062313  44062513  44062913  44063013  44060612  44060712  44060912  44061012  44061412  44061612  44061812  44062013  44062213  44062613  44062813  44063313  44063513  44063613  44063713  44063913.",
      "center_classification_date": "20140108",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) including states of: TX, NC, OH, IA, UT, KY, ME, CA, NY, MA. NJ, IL, PA,, TN, FL, MO, GA, HI, SC, NE, MI, and GA and  countries of: Dominican Republic, Germany, Switzerland, Australia, Spain, Japan, United Kingdom, Canada, France, The Netherlands, Austria, and Singapore.",
      "state": "CA",
      "product_description": "Catalys Precision Laser System (United States) - Catalys-U:  Catalys Precision Laser System (International) - Catalys I    200-240V ~, 50/60 Hz, 15A,  (2) 15A Resettable CB:  Tested and complies with FCC Part 15 Class A.    OptiMedica  1310 Moffett Park Drive  Sunnyvale, CA 94089    The OptiMedica Catalys Precision Laser System is indicated for use in patients undergoing cataract surgery for removal of the crystalline lens.",
      "report_date": "20140115",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Optimedica Corporation",
      "recall_number": "Z-0649-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "67150",
      "termination_date": "20140626",
      "more_code_info": "",
      "recall_initiation_date": "20131220",
      "postal_code": "94089-1133",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}