{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Brooklyn Park",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79058",
      "recalling_firm": "Medtronic Perfusion Systems",
      "address_1": "7611 Northland Dr N",
      "address_2": "N/A",
      "postal_code": "55428-1088",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Distribution and to the countries of : Australia, Canada, Colombia, Denmark, Germany, Guyana, Israel, Italy, Japan, Netherlands, Norway, Saudi Arabia and  United Kingdom.",
      "recall_number": "Z-0646-2018",
      "product_description": "Medtronic TRILLIUM AFFINITY NT, Cardiotomy/Venous Reservoir (CVR) with Filter,with Trillium Biosurface, 540T  The AFFINITY NT Cardiotomy/Venous Reservoir (CVR) with Filter is a single used device designed to collect and store blood during extracorporeal circulation. Venous blood is collected and de foamed while cardiotomy blood is collected, de foamed and filtered before mixing with the venous blood. The AFFINITY NT Cardiotomy/Venous Reservoir (CVR) with filter is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures for up to 6 hours in duration.",
      "product_quantity": "3864 units",
      "reason_for_recall": "Possible sterile barrier breach in the pouch of Affinity NT Cardiotomy Venous Reservoir stand alone uncoated and Affinity NT Cardiotomy Venous Reservoir stand alone Trillium coated products.",
      "recall_initiation_date": "20180116",
      "center_classification_date": "20180220",
      "termination_date": "20200821",
      "report_date": "20180228",
      "code_info": "All lots"
    }
  ]
}