  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Vancouver",
      "state": "WA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72915",
      "recalling_firm": "CompView Medical, LLC",
      "address_1": "2818 E 2nd St",
      "address_2": "Suite 200",
      "postal_code": "98661-7778",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "US nationwide distribution including AZ, CA, CO, GA, OH, and TN.",
      "recall_number": "Z-0644-2016",
      "product_description": "Model NuCARTGN.  The NuCART is a mobile display and image management platform used in Minimally Invasive Surgery (MIS) environments to facilitate the routing and ergonomic display of video signals from source devices to flat-panel monitors.  It is tailored for use with an OEC C-arm and Endo-Cam in MIS surgery suites, Ambulatory Surgery Centers, Urology, GI and other suites.",
      "product_quantity": "10 units",
      "reason_for_recall": "The attachment of the NuCART Spring Arm to the NuCART horizontal boom may come apart if the securing ring (snap ring) was not installed correctly causing the spring arm to release and detach.",
      "recall_initiation_date": "20151204",
      "center_classification_date": "20160113",
      "termination_date": "20160331",
      "report_date": "20160120",
      "code_info": "Lot# T9993-02-01 /15924CARTXXTNNA / DX00357  Lot# T9993-02-02 / 15924CARTXXAZPH / DX00358  Lot# U0328-02-02 / 15A01CARTXXCOEN /DX00360  Lot# U0328-02-01 / 15A26CARTXXCAK1 / DX00359  Lot# U0718-02-01 / 15A26CARTXXCAK2 / DX00361  Lot# U0718-02-02 / 15A26CARTXXCAK3 / DX00362  Lot# U1033-02-01 / 15A19CARTXXOHCA /DX00363  Lot# U1033-02-02 / 15A26CARTXXCASJ / DX00364  Lot# U1126-02-01 / 15A26CARTXXCAK41 /DX00365  Lot# U1126-02-02 / 15806CARTXXGAAT / DX00366"
    }
  ]
}