  "results": [
    {
      "country": "United States",
      "city": "Waukesha",
      "address_1": "3000 N Grandview Blvd",
      "reason_for_recall": "Issue #1 affects Software Versions 6.0 SP7, SP7.0.1, SP7.0.2, SP7.0.3, SP8, SP8.0.1, SP8.0.a and SP8.0.2.  When an image is flipped horizontally and vertically, the orientation markers appear correctly on the image to which they were applied. However, once new image rendering happens on this image or series (with scrolling; playing cin¿; maximizing or changing viewports/layouts; or synchronizing viewports and resizing the browser), the new image orientation is correctly maintained but the display of orientation markers can become incorrect.    Issue #2 affects Software Versions 6.0 SP7, SP7.0.1, SP7.0.2, SP7.0.3, SP8, SP8.0.1, SP8.0.a, SP8.0.2, SP9 and SP9.0.1.   When an image in a series is flipped and then rotated in Zero Footprint Viewer (ZFP), the orientation markers appear correctly on the image to which they were applied. However, once you navigate to another image in the series and the image you navigated to has a different orientation than the prior image from which you navigated, the new image orientation is correctly maintained but the display of orientation markers can become incorrect.",
      "address_2": "",
      "product_quantity": "508 units",
      "code_info": "Serial Number  UDI  4532-2-Centricity Universal Viewer Zero Footprint / ZFP-01889426 (01)00840682102988(10)6.0SP82093796001D  4532-2-Centricity Universal Viewer Zero Footprint / ZFP-01889426 (01)00840682102988(10)6.0SP82093796001D  10213-1-Centricity Universal Viewer Zero Footprint / ZFP-01815213 (01)00840682102988(10)6.0SP72093796001D  10711-1-Centricity Universal Viewer Zero Footprint / ZFP-01970566 (01)00840682102988(10)6.0SP72093796001D  100620-2-Centricity Universal Viewer Zero Footprint / ZFP-00099694 (01)00840682102988(10)6.0SP9.0.12093796001D  103960-2-Centricity Universal Viewer Zero Footprint / ZFP-00105531 (01)00840682102988(10)6.0SP7.0.22093796001D  3730-1-Centricity Universal Viewer Zero Footprint / ZFP-01803680 (01)00840682102988(10)6.0SP7.0.22093796001D  3730-1-Centricity Universal Viewer Zero Footprint / ZFP-01803680 (01)00840682102988(10)6.0SP7.0.22093796001D  103960-2-Centricity Universal Viewer Zero Footprint / ZFP-00102068 (01)00840682102988(10)6.0SP9.0.12093796001D  103960-2-Centricity Universal Viewer Zero Footprint / ZFP-00105530 (01)00840682102988(10)6.0SP9.0.12093796001D  103960-2-Centricity Universal Viewer Zero Footprint / ZFP-00105531 (01)00840682102988(10)6.0SP9.0.12093796001D  103960-2-Centricity Universal Viewer Zero Footprint / ZFP-00105531 (01)00840682102988(10)6.0SP9.0.12093796001D  103960-2-Centricity Universal Viewer Zero Footprint / ZFP-02025410 (01)00840682102988(10)6.0SP9.0.12093796001D  103960-2-Centricity Universal Viewer Zero Footprint / ZFP-02025461 (01)00840682102988(10)6.0SP9.0.12093796001D  105412-2-Centricity Universal Viewer Zero Footprint / ZFP-01978386 (01)00840682102988(10)6.0SP92093796001D  3846-2-Centricity Universal Viewer Zero Footprint / ZFP-01613236 (01)00840682102988(10)6.0SP72093796001D  2317-3-Centricity Universal Viewer Zero Footprint / ZFP-00165279 (01)00840682102988(10)6.0SP7.0.22093796001D  2317-3-Centricity Universal Viewer Zero Footprint / ZFP-00165278 (01)00840682102988(10)6.0SP7.0.22093796001D  1202-8-Centricity Universal Viewer Zero Footprint / ZFP-00100999 (01)00840682102988(10)6.0SP7.0.22093796001D  3805-8-Centricity Universal Viewer Zero Footprint / ZFP-02098759 (01)00840682102988(10)6.0SP8.0.22093796001D  3805-8-Centricity Universal Viewer Zero Footprint / ZFP-02098758 (01)00840682102988(10)6.0SP8.0.22093796001D  3805-8-Centricity Universal Viewer Zero Footprint / ZFP-02098757 (01)00840682102988(10)6.0SP8.0.22093796001D  3805-8-Centricity Universal Viewer Zero Footprint / ZFP-02098755 (01)00840682102988(10)6.0SP8.0.22093796001D  101764-1-Centricity Universal Viewer Zero Footprint / ZFP-01432060 (01)00840682102988(10)6.0SP7.0.12093796001D  101764-1-Centricity Universal Viewer Zero Footprint / ZFP-01432060 (01)00840682102988(10)6.0SP7.0.12093796001D  101764-1-Centricity Universal Viewer Zero Footprint / ZFP-01432060 (01)00840682102988(10)6.0SP7.0.12093796001D  101764-1-Centricity Universal Viewer Zero Footprint / ZFP-01432060 (01)00840682102988(10)6.0SP72093796001D  101764-1-Centricity Universal Viewer Zero Footprint / ZFP-01432060 (01)00840682102988(10)6.0SP72093796001D  101764-1-Centricity Universal Viewer Zero Footprint / ZFP-01432060 (01)00840682102988(10)6.0SP72093796001D  103944-3-Centricity Universal Viewer Zero Footprint / ZFP-01477454 (01)00840682102988(10)6.0SP8.0.22093796001D  10345-9-Centricity Universal Viewer Zero Footprint / ZFP-01425634 (01)00840682102988(10)6.0SP8.0.22093796001D  10345-9-Centricity Universal Viewer Zero Footprint / ZFP-01425634 (01)00840682102988(10)6.0SP8.0.22093796001D  10345-1-Centricity Universal Viewer Zero Footprint / ZFP-01419025 (01)00840682102988(10)6.0SP8.0.22093796001D  101349-2-Centricity Universal Viewer Zero Footprint / ZFP-00107435 (01)00840682102988(10)6.0SP7.0.22093796001D  101349-2-Centricity Universal Viewer Zero Footprint / ZFP-02098718 (01)00840682102988(10)6.0SP7.0.22093796001D  103944-3-Centricity Universal Viewer Zero Footprint / ZFP-01477454 (01)00840682102988(10)6.0SP9.0.12093796001D  13132-11-Centricity Universal Viewer Zero Footprint / ZFP-01445043 (01)00840682102988(10)6.0SP72093796001D  13132-11-Centricity Universal Viewer Zero Footprint / ZFP-01445043 (01)00840682102988(10)6.0SP72093796001D  13132-11-Centricity Universal Viewer Zero Footprint / ZFP-01445043 (01)00840682102988(10)6.0SP72093796001D  10691-2-Centricity Universal Viewer Zero Footprint / ZFP-01817128 (01)00840682102988(10)6.0SP7.0.22093796001D  10691-2-Centricity Universal Viewer Zero Footprint / ZFP-02095550 (01)00840682102988(10)6.0SP7.0.22093796001D  1215-20-Centricity Universal Viewer Zero Footprint / ZFP-00105026 (01)00840682102988(10)6.0SP72093796001D  1215-20-Centricity Universal Viewer Zero Footprint / ZFP-00105026 (01)00840682102988(10)6.0SP72093796001D  1215-20-Centricity Universal Viewer Zero Footprint / ZFP-00105026 (01)00840682102988(10)6.0SP72093796001D  11619-2-Centricity Universal Viewer Zero Footprint / ZFP-01810676 (01)00840682102988(10)6.0SP72093796001D  11619-2-Centricity Universal Viewer Zero Footprint / ZFP-01810676 (01)00840682102988(10)6.0SP72093796001D  11619-2-Centricity Universal Viewer Zero Footprint / ZFP-01810676 (01)00840682102988(10)6.0SP72093796001D  101499-1-Centricity Universal Viewer Zero Footprint / ZFP-01797031 (01)00840682102988(10)6.0SP7.0.22093796001D  101499-1-Centricity Universal Viewer Zero Footprint / ZFP-02091471 (01)00840682102988(10)6.0SP72093796001D  102742-1-Centricity Universal Viewer Zero Footprint / ZFP-01804953 (01)00840682102988(10)6.0SP7.0.12093796001D  102742-1-Centricity Universal Viewer Zero Footprint / ZFP-01804952 (01)00840682102988(10)6.0SP7.0.12093796001D  102742-1-Centricity Universal Viewer Zero Footprint / ZFP-01804952 (01)00840682102988(10)6.0SP7.0.12093796001D  3212-8-Centricity Universal Viewer Zero Footprint / ZFP-01740249 (01)00840682102988(10)6.0SP72093796001D  3212-8-Centricity Universal Viewer Zero Footprint / ZFP-01740249 (01)00840682102988(10)6.0SP72093796001D  3212-8-Centricity Universal Viewer Zero Footprint / ZFP-01740249 (01)00840682102988(10)6.0SP72093796001D  1049-5-Centricity Universal Viewer Zero Footprint / ZFP-01477460 (01)00840682102988(10)6.0SP8.0.22093796001D  1049-5-Centricity Universal Viewer Zero Footprint / ZFP-01477460 (01)00840682102988(10)6.0SP7.0.12093796001D  1049-5-Centricity Universal Viewer Zero Footprint / ZFP-01477460 (01)00840682102988(10)6.0SP7.0.12093796001D  1049-5-Centricity Universal Viewer Zero Footprint / ZFP-01477460 (01)00840682102988(10)6.0SP7.0.12093796001D  1049-5-Centricity Universal Viewer Zero Footprint / ZFP-01477460 (01)00840682102988(10)6.0SP7.0.12093796001D  10547-2-Centricity Universal Viewer Zero Footprint / ZFP-01294055 (01)00840682102988(10)6.0SP72093796001D  10547-2-Centricity Universal Viewer Zero Footprint / ZFP-01294055 (01)00840682102988(10)6.0SP72093796001D  10547-2-Centricity Universal Viewer Zero Footprint / ZFP-01294055 (01)00840682102988(10)6.0SP72093796001D  9915-2-Centricity Universal Viewer Zero Footprint / ZFP-01459010 (01)00840682102988(10)6.0SP7.0.22093796001D  3228-12-Centricity Universal Viewer Zero Footprint / ZFP-01451967 (01)00840682102988(10)6.0SP7.0.12093796001D  3228-12-Centricity Universal Viewer Zero Footprint / ZFP-01451967 (01)00840682102988(10)6.0SP72093796001D  3228-12-Centricity Universal Viewer Zero Footprint / ZFP-01451967 (01)00840682102988(10)6.0SP7.0.22093796001D  Entitlement ID 710d-5e4e-c28d-4be1-a48f-52ca-60cd-c333 (01)00840682102988(10)6.0SP9.0.12093796001D  4220-1-Centricity Universal Viewer Zero Footprint / ZFP-01404916 (01)00840682102988(10)6.0SP82093796001D  4220-1-Centricity Universal Viewer Zero Footprint / ZFP-01404916 (01)00840682102988(10)6.0SP82093796001D  4220-1-Centricity Universal Viewer Zero Footprint / ZFP-01404916 (01)00840682102988(10)6.0SP82093796001D  4220-1-Centricity Universal Viewer Zero Footprint / ZFP-01404916 (01)00840682102988(10)6.0SP82093796001D  4220-1-Centricity Universal Viewer Zero Footprint / ZFP-01404916 (01)00840682102988(10)6.0SP82093796001D  4220-1-Centricity Universal Viewer Zero Footprint / ZFP-01404916 (01)00840682102988(10)6.0SP82093796001D  4220-1-Centricity Universal Viewer Zero Footprint / ZFP-01404916 (01)00840682102988(10)6.0SP82093796001D  3480-7-Centricity Universal Viewer Zero Footprint / ZFP-00555367 (01)00840682102988(10)6.0SP8.0.22093796001D  3480-7-Centricity Universal Viewer Zero Footprint / ZFP-00555366 (01)00840682102988(10)6.0SP8.0.22093796001D  3480-7-Centricity Universal Viewer Zero Footprint / ZFP-00555365 (01)00840682102988(10)6.0SP8.0.22093796001D  3480-7-Centricity Universal Viewer Zero Footprint / ZFP-00555364 (01)00840682102988(10)6.0SP8.0.22093796001D  3480-7-Centricity Universal Viewer Zero Footprint / ZFP-00555363 (01)00840682102988(10)6.0SP8.0.22093796001D  3480-7-Centricity Universal Viewer Zero Footprint / ZFP-00555362 (01)00840682102988(10)6.0SP8.0.22093796001D  3480-7-Centricity Universal Viewer Zero Footprint / ZFP-00555361 (01)00840682102988(10)6.0SP8.0.22093796001D  3876-1-Centricity Universal Viewer Zero Footprint / ZFP-00103824 (01)00840682102988(10)6.0SP8.0.22093796001D  1198-18-Centricity Universal Viewer Zero Footprint / ZFP-01370311 (01)00840682102988(10)6.0SP9.0.12093796001D  3480-7-Centricity Universal Viewer Zero Footprint / ZFP-00555361 (01)00840682102988(10)6.0SP9.0.12093796001D  3480-7-Centricity Universal Viewer Zero Footprint / ZFP-00555362 (01)00840682102988(10)6.0SP9.0.12093796001D  3480-7-Centricity Universal Viewer Zero Footprint / ZFP-00555363 (01)00840682102988(10)6.0SP9.0.12093796001D  3480-7-Centricity Universal Viewer Zero Footprint / ZFP-00555364 (01)00840682102988(10)6.0SP9.0.12093796001D  3480-7-Centricity Universal Viewer Zero Footprint / ZFP-00555365 (01)00840682102988(10)6.0SP9.0.12093796001D  3480-7-Centricity Universal Viewer Zero Footprint / ZFP-00555366 (01)00840682102988(10)6.0SP9.0.12093796001D  3480-7-Centricity Universal Viewer Zero Footprint / ZFP-00555367 (01)00840682102988(10)6.0SP9.0.12093796001D  To be provided (01)00840682102988(10)6.0SP9012093796001D  3455-3-Centricity Universal Viewer Zero Footprint / ZFP-01813654 (01)00840682102988(10)6.0SP7.0.12093796001D  3455-3-Centricity Universal Viewer Zero Footprint / ZFP-01813647 (01)00840682102988(10)6.0SP7.0.12093796001D  3455-3-Centricity Universal Viewer Zero Footprint / ZFP-01813644 (01)00840682102988(10)6.0SP7.0.12093796001D  2387-17-Centricity Universal Viewer Zero Footprint / ZFP-01982831 (01)00840682102988(10)6.0SP72093796001D  2387-15-Centricity Universal Viewer Zero Footprint / ZFP-01270499 (01)00840682102988(10)6.0SP7.0.22093796001D  2387-15-Centricity Universal Viewer Zero Footprint / ZFP-01270499 (01)00840682102988(10)6.0SP92093796001D  2219-7-Centricity Universal Viewer Zero Footprint / ZFP-01796537 (01)00840682102988(10)6.0SP8.0.12093796001D  2219-7-Centricity Universal Viewer Zero Footprint / ZFP-01796537 (01)00840682102988(10)6.0SP7.0.12093796001D  2219-7-Centricity Universal Viewer Zero Footprint / ZFP-01796537 (01)00840682102988(10)6.0SP7.0.12093796001D  2219-7-Centricity Universal Viewer Zero Footprint / ZFP-00347513 (01)00840682102988(10)6.0SP8.0.12093796001D  2219-7-Centricity Universal Viewer Zero Footprint / ZFP-00347507 (01)00840682102988(10)6.0SP8.0.12093796001D  2219-7-Centricity Universal Viewer Zero Footprint / ZFP-00347506 (01)00840682102988(10)6.0SP8.0.12093796001D  2219-7-Centricity Universal Viewer Zero Footprint / ZFP-00347505 (01)00840682102988(10)6.0SP8.0.12093796001D  2219-7-Centricity Universal Viewer Zero Footprint / ZFP-00347503 (01)00840682102988(10)6.0SP8.0.12093796001D  1542-5-Centricity Universal Viewer Zero Footprint / ZFP-01970562 (01)00840682102988(10)6.0SP7.0.12093796001D  1542-5-Centricity Universal Viewer Zero Footprint / ZFP-01970562 (01)00840682102988(10)6.0SP7.0.12093796001D  1207-13-Centricity Universal Viewer Zero Footprint / ZFP-01980284 (01)00840682102988(10)6.0SP72093796001D  1207-13-Centricity Universal Viewer Zero Footprint / ZFP-00111071 (01)00840682102988(10)6.0SP7.0.22093796001D  1609-1-Centricity Universal Viewer Zero Footprint / ZFP-00104759 (01)00840682102988(10)6.0SP72093796001D  1609-1-Centricity Universal Viewer Zero Footprint / ZFP-00104758 (01)00840682102988(10)6.0SP72093796001D  Entitlement ID e7d7-883f-4607-4906-a34b-fa41-5e3a-210a (01)00840682102988(10)6.0SP82093796001D  3066-1-Centricity Universal Viewer Zero Footprint / ZFP-00101769 (01)00840682102988(10)6.0SP7.0.12093796001D  3066-1-Centricity Universal Viewer Zero Footprint / ZFP-00101769 (01)00840682102988(10)6.0SP7.0.12093796001D  1331-24-Centricity Universal Viewer Zero Footprint / ZFP-02041554 (01)00840682102988(10)6.0SP82093796001D  1331-24-Centricity Universal Viewer Zero Footprint / ZFP-02041554 (01)00840682102988(10)6.0SP82093796001D  1331-24-Centricity Universal Viewer Zero Footprint / ZFP-02041554 (01)00840682102988(10)6.0SP82093796001D  3922-17-Centricity Universal Viewer Zero Footprint / ZFP-00098227 (01)00840682102988(10)6.0SP7.0.22093796001D  3808-1-Centricity Universal Viewer Zero Footprint / ZFP-00101964 (01)00840682102988(10)6.0SP7.0.22093796001D  3808-1-Centricity Universal Viewer Zero Footprint / ZFP-00101964 (01)00840682102988(10)6.0SP9.0.12093796001D  9862-1-Centricity Universal Viewer Zero Footprint / ZFP-00100263 (01)00840682102988(10)6.0SP7.0.22093796001D  103918-1-Centricity Universal Viewer Zero Footprint / ZFP-00100288 (01)00840682102988(10)6.0SP7.0.22093796001D  103918-1-Centricity Universal Viewer Zero Footprint / ZFP-00100287 (01)00840682102988(10)6.0SP7.0.22093796001D  4204-1-Centricity Universal Viewer Zero Footprint / ZFP-01459008 (01)00840682102988(10)6.0SP72093796001D  4204-1-Centricity Universal Viewer Zero Footprint / ZFP-01459008 (01)00840682102988(10)6.0SP72093796001D  3475-2-Centricity Universal Viewer Zero Footprint / ZFP-01284910 (01)00840682102988(10)6.0SP7.0.22093796001D  3475-2-Centricity Universal Viewer Zero Footprint / ZFP-01284909 (01)00840682102988(10)6.0SP7.0.22093796001D  Entitlement ID 9907-5584-a777-4b07-8de5-64c5-a98e-3bf8 (01)00840682102988(10)6.0SP9.0.12093796001D  101647-1-Centricty Universal Viewer Zero Footprint / ZFP01810709 To be provided  6901-6-Centricity Universal Viewer Zero Footprint / ZFP-01974088 (01)00840682102988(10)6.0SP9.0.12093796001D  6901-6-Centricity Universal Viewer Zero Footprint / ZFP-01974088 (01)00840682102988(10)6.0SP9.0.12093796001D  6901-6-Centricity Universal Viewer Zero Footprint / ZFP-01974088 (01)00840682102988(10)6.0SP9.0.12093796001D  3677-6-Centricity Universal Viewer Zero Footprint / ZFP-01893388 (01)00840682102988(10)6.0SP82093796001D  4424-2-Centricity Universal Viewer Zero Footprint / ZFP-01474436 (01)00840682102988(10)6.0SP72093796001D  1472-15-Centricity Universal Viewer Zero Footprint / ZFP-01372922 (01)00840682102988(10)6.0SP7.0.22093796001D  1472-15-Centricity Universal Viewer Zero Footprint / ZFP-01372921 (01)00840682102988(10)6.0SP7.0.22093796001D  1472-15-Centricity Universal Viewer Zero Footprint / ZFP-01372920 (01)00840682102988(10)6.0SP7.0.22093796001D  1472-15-Centricity Universal Viewer Zero Footprint / ZFP-01372919 (01)00840682102988(10)6.0SP7.0.22093796001D  1472-15-Centricity Universal Viewer Zero Footprint / ZFP-01372918 (01)00840682102988(10)6.0SP7.0.22093796001D  1472-15-Centricity Universal Viewer Zero Footprint / ZFP-01367875 (01)00840682102988(10)6.0SP72093796001D  1472-15-Centricity Universal Viewer Zero Footprint / ZFP-01367875 (01)00840682102988(10)6.0SP72093796001D  1472-15-Centricity Universal Viewer Zero Footprint / ZFP-01367875 (01)00840682102988(10)6.0SP72093796001D  1472-15-Centricity Universal Viewer Zero Footprint / ZFP-01367875 (01)00840682102988(10)6.0SP7.0.22093796001D  3760-1-Centricity Universal Viewer Zero Footprint / ZFP-00101629 (01)00840682102988(10)6.0SP7.0.22093796001D  1847-2-Centricity Universal Viewer Zero Footprint / ZFP-00169967 (01)00840682102988(10)6.0SP7.0.22093796001D  3726-1-Centricity Universal Viewer Zero Footprint / ZFP-00599126 (01)00840682102988(10)6.0SP7.0.22093796001D  3726-1-Centricity Universal Viewer Zero Footprint / ZFP-00599124 (01)00840682102988(10)6.0SP7.0.22093796001D  3726-1-Centricity Universal Viewer Zero Footprint / ZFP-00598186 (01)00840682102988(10)6.0SP7.0.22093796001D  1495-3-Centricity Universal Viewer Zero Footprint / ZFP-00096091 (01)00840682102988(10)6.0SP8.0.22093796001D  1495-3-Centricity Universal Viewer Zero Footprint / ZFP-00096090 (01)00840682102988(10)6.0SP8.0.22093796001D  105452-1-GE Health Cloud Platform-02093367 To be provided  DZ_KPMA-Centricity Universal Viewer Zero Footprint / ZFP-00165343 (01)00840682102988(10)6.0  SiteID-Centricity Universal Viewer Zero Footprint / ZFP-02154606 (01)00840682102988(10)6.0SP72093796001D  1698348_Alfred-Centricity Universal Viewer Zero Footprint / ZFP-02032817 (01)00840682102988(10)6.0SP9.0.12093796001D  110020-Centricity Universal Viewer Zero Footprint / ZFP-00044290 (01)00840682102988(10)6.0  110020-Centricity Universal Viewer Zero Footprint / ZFP-00044290 (01)00840682102988(10)6.0  Entitlement ID 8196-4472-3601-46d9-9a11-52b4-80fb-fd27 (01)00840682102988(10)6.0SP9.0.12093796001D  349104-Centricity Universal Viewer Zero Footprint / ZFP-01984551 (01)00840682102988(10)6.0SP7.0.12093796001D  283060-Centricity Universal Viewer Zero Footprint / ZFP-02043239 (01)00840682102988(10)6.0SP72093796001D  283060-Centricity Universal Viewer Zero Footprint / ZFP-02043239 (01)00840682102988(10)6.0SP72093796001D  283060-Centricity Universal Viewer Zero Footprint / ZFP-02043239 (01)00840682102988(10)6.0SP9.0.12093796001D  104995-1-Centricity Universal Viewer Zero Footprint / ZFP-01819807 (01)00840682102988(10)6.0SP72093796001D  104995-1-Centricity Universal Viewer Zero Footprint / ZFP-01819807 (01)00840682102988(10)6.0SP7.0.12093796001D  104995-1-Centricity Universal Viewer Zero Footprint / ZFP-01819807 (01)00840682102988(10)6.0SP9.0.12093796001D  282984-Centricity Universal Viewer Zero Footprint / ZFP-01373559 (01)00840682102988(10)6.0SP7.0.12093796001D  282984-Centricity Universal Viewer Zero Footprint / ZFP-01373559 (01)00840682102988(10)6.0SP7.0.22093796001D  282984-Centricity Universal Viewer Zero Footprint / ZFP-01373558 (01)00840682102988(10)6.0SP7.0.22093796001D  282984-Centricity Universal Viewer Zero Footprint / ZFP-01373557 (01)00840682102988(10)6.0SP7.0.22093796001D  282984-Centricity Universal Viewer Zero Footprint / ZFP-01373560 (01)00840682102988(10)6.0SP9.0.12093796001D  92317-Centricity Universal Viewer Zero Footprint / ZFP-01763825 (01)00840682102988(10)6.0SP72093796001D  92317-Centricity Universal Viewer Zero Footprint / ZFP-01300703 (01)00840682102988(10)6.0SP72093796001D  92317-Centricity Universal Viewer Zero Footprint / ZFP-01300703 (01)00840682102988(10)6.0SP92093796001D  92317-Centricity Universal Viewer Zero Footprint / ZFP-01763825 (01)00840682102988(10)6.0SP92093796001D  282927-Centricity Universal Viewer Zero Footprint / ZFP-01425173 (01)00840682102988(10)6.0SP82093796001D  282927-Centricity Universal Viewer Zero Footprint / ZFP-01452465 (01)00840682102988(10)6.0SP72093796001D  282927-Centricity Universal Viewer Zero Footprint / ZFP-01815401 (01)00840682102988(10)6.0SP82093796001D  282927-Centricity Universal Viewer Zero Footprint / ZFP-00605617 (01)00840682102988(10)6.0SP7.0.22093796001D  GDZH-JINWAN-Centricity Universal Viewer Zero Footprint / ZFP-02171477 (01)00840682102988(10)6.0SP72093796001D  GX-NO2-Centricity Universal Viewer Zero Footprint / ZFP-02171485 (01)00840682102988(10)6.0SP72093796001D  NANJING-Centricity Universal Viewer Zero Footprint / ZFP-02170933 (01)00840682102988(10)6.0SP72093796001D  JS-TUMOUR-Centricity Universal Viewer Zero Footprint / ZFP-02171453 (01)00840682102988(10)6.0SP72093796001D  SUZHOU-Centricity Universal Viewer Zero Footprint / ZFP-02170934 (01)00840682102988(10)6.0SP72093796001D  WJPEOPLE-Centricity Universal Viewer Zero Footprint / ZFP-02171481 (01)00840682102988(10)6.0SP72093796001D  BEIJING302-Centricity Universal Viewer Zero Footprint / ZFP-02170935 (01)00840682102988(10)6.0SP72093796001D  SH-CH-Centricity Universal Viewer Zero Footprint / ZFP-02171452 (01)00840682102988(10)6.0SP72093796001D  FUDAN-TUMOUR-Centricity Universal Viewer Zero Footprint / ZFP-02171473 (01)00840682102988(10)6.0SP72093796001D  082455130335-Centricity Universal Viewer Zero Footprint / ZFP-02171476 (01)00840682102988(10)6.0SP72093796001D  YN-HAOBANG-Centricity Universal Viewer Zero Footprint / ZFP-02171461 (01)00840682102988(10)6.0SP72093796001D  828550020-Centricity Universal Viewer Zero Footprint / ZFP-01287260 (01)00840682102988(10)6.0SP7.0.12093796001D  0828550068-Centricity Universal Viewer Zero Footprint / ZFP-02094418 (01)00840682102988(10)6.0SP82093796001D  PM-RT-Centricity Universal Viewer Zero Footprint / ZFP-02094419 (01)00840682102988(10)6.0SP7.0.12093796001D  40549-1-Centricity Universal Viewer Zero Footprint / ZFP-01817483 (01)00840682102988(10)6.0SP72093796001D  EG1017-Centricity Universal Viewer Zero Footprint / ZFP-01987374 (01)00840682102988(10)6.0SP72093796001D  A44986-Centricity Universal Viewer Zero Footprint / ZFP-01797479 (01)00840682102988(10)6.0SP7.0.22093796001D  M91961-Centricity Universal Viewer Zero Footprint / ZFP-01301183 (01)00840682102988(10)6.0  SiteID-Centricity Universal Viewer Zero Footprint / ZFP-01811336 (01)00840682102988(10)6.0SP72093796001D  SiteID-Centricity Universal Viewer Zero Footprint / ZFP-01811336 (01)00840682102988(10)6.0SP72093796001D  M51725-Centricity Universal Viewer Zero Footprint / ZFP-01300873 (01)00840682102988(10)6.0SP7.0.22093796001D  P41728-Centricity Universal Viewer Zero Footprint / ZFP-00044506 (01)00840682102988(10)6.0SP7.0.22093796001D  MQ12345-Centricity Universal Viewer Zero Footprint / ZFP-00162347 (01)00840682102988(10)6.0SP7.0.22093796001D  M44850-Centricity Universal Viewer Zero Footprint / ZFP-01817813 (01)00840682102988(10)6.0SP7.0.22093796001D  NC1075-Centricity Universal Viewer Zero Footprint / ZFP-01987930 (01)00840682102988(10)6.0SP7.0.22093796001D  M41441-Centricity Universal Viewer Zero Footprint / ZFP-01463298 (01)00840682102988(10)6.0SP7.0.12093796001D  M41441-Centricity Universal Viewer Zero Footprint / ZFP-01463299 (01)00840682102988(10)6.0SP72093796001D  U2512164-Centricity Universal Viewer Zero Footprint / ZFP-01811011 (01)00840682102988(10)6.0SP72093796001D  U2512164-Centricity Universal Viewer Zero Footprint / ZFP-01811011 (01)00840682102988(10)6.0SP72093796001D  M40455-Centricity Universal Viewer Zero Footprint / ZFP-00044391 (01)00840682102988(10)6.0SP7.0.22093796001D  M60364-Centricity Universal Viewer Zero Footprint / ZFP-01801170 (01)00840682102988(10)6.0SP7.0.12093796001D  M60364-Centricity Universal Viewer Zero Footprint / ZFP-01977877 (01)00840682102988(10)6.0SP72093796001D  C41917-Centricity Universal Viewer Zero Footprint / ZFP-00117236 (01)00840682102988(10)6.0SP7.0.22093796001D  D41917-Centricity Universal Viewer Zero Footprint / ZFP-00117239 (01)00840682102988(10)6.0SP7.0.22093796001D  M15587-Centricity Universal Viewer Zero Footprint / ZFP-01301185 (01)00840682102988(10)6.0  M54812-Centricity Universal Viewer Zero Footprint / ZFP-00436707 (01)00840682102988(10)6.0  M41980-Centricity Universal Viewer Zero Footprint / ZFP-00044516 (01)00840682102988(10)6.0SP7.0.22093796001D  A2044938-Centricity Universal Viewer Zero Footprint / ZFP-01977789 (01)00840682102988(10)6.0SP72093796001D  A2044938-Centricity Universal Viewer Zero Footprint / ZFP-01977790 (01)00840682102988(10)6.0SP72093796001D  D48589-Centricity Universal Viewer Zero Footprint / ZFP-00044452 (01)00840682102988(10)6.0SP7.0.22093796001D  D48589-Centricity Universal Viewer Zero Footprint / ZFP-00044452 (01)00840682102988(10)6.0SP7.0.22093796001D  M91647-Centricity Universal Viewer Zero Footprint / ZFP-00110957 (01)00840682102988(10)6.0SP7.0.22093796001D  M3488569-Centricity Universal Viewer Zero Footprint / ZFP-00071697 (01)00840682102988(10)6.0SP7.0.22093796001D  M3488569-Centricity Universal Viewer Zero Footprint / ZFP-00071697 (01)00840682102988(10)6.0  M3488569-Centricity Universal Viewer Zero Footprint / ZFP-00071708 (01)00840682102988(10)6.0  SiteID-Centricity Universal Viewer Zero Footprint / ZFP-01815997 (01)00840682102988(10)6.0SP72093796001D  HC0399-Centricity Universal Viewer Zero Footprint / ZFP-01282975 (01)00840682102988(10)6.0SP72093796001D  HC0399-Centricity Universal Viewer Zero Footprint / ZFP-01282975 (01)00840682102988(10)6.0SP7.0.22093796001D  682243-Centricity Universal Viewer Zero Footprint / ZFP-02158351 (01)00840682102988(10)6.0SP72093796001D  PC1695-Centricity Universal Viewer Zero Footprint / ZFP-01447702 (01)00840682102988(10)6.0SP72093796001D  CC0618-Centricity Universal Viewer Zero Footprint / ZFP-00000895 (01)00840682102988(10)6.0SP7.0.12093796001D  530019-Centricity Universal Viewer Zero Footprint / ZFP-01282264 (01)00840682102988(10)6.0SP72093796001D  530019-Centricity Universal Viewer Zero Footprint / ZFP-01282264 (01)00840682102988(10)6.0SP72093796001D  530019-Centricity Universal Viewer Zero Footprint / ZFP-01979729 (01)00840682102988(10)6.0SP72093796001D  530019-Centricity Universal Viewer Zero Footprint / ZFP-01979731 (01)00840682102988(10)6.0SP72093796001D  518086-Centricity Universal Viewer Zero Footprint / ZFP-00158637 (01)00840682102988(10)6.0SP7.0.22093796001D  HC2720-Centricity Universal Viewer Zero Footprint / ZFP-00163499 (01)00840682102988(10)6.0SP7.0.22093796001D  HC2720-Centricity Universal Viewer Zero Footprint / ZFP-00163499 (01)00840682102988(10)6.0SP7.0.22093796001D  HC2720-Centricity Universal Viewer Zero Footprint / ZFP-00163499 (01)00840682102988(10)6.0SP7.0.22093796001D  SiteID-Centricity Universal Viewer Zero Footprint / ZFP-01816023 (01)00840682102988(10)6.0SP72093796001D  HC0074-Centricity Universal Viewer Zero Footprint / ZFP-01463865 (01)00840682102988(10)6.0SP7.0.22093796001D  HC1627-Centricity Universal Viewer Zero Footprint / ZFP-00260658 (01)00840682102988(10)6.0SP7.0.22093796001D  PC0114-Centricity Universal Viewer Zero Footprint / ZFP-00261054 (01)00840682102988(10)6.0SP7.0.22093796001D  HC0881-Centricity Universal Viewer Zero Footprint / ZFP-01423579 (01)00840682102988(10)6.0SP7.0.22093796001D  UC2847-Centricity Universal Viewer Zero Footprint / ZFP-00260806 (01)00840682102988(10)6.0SP7.0.22093796001D  UC2847-Centricity Universal Viewer Zero Footprint / ZFP-00260806 (01)00840682102988(10)6.0  UESS01-Centricity Universal Viewer Zero Footprint / ZFP-02041830 (01)00840682102988(10)6.0SP72093796001D  UESS01-Centricity Universal Viewer Zero Footprint / ZFP-02041830 (01)00840682102988(10)6.0SP72093796001D  UESS01-Centricity Universal Viewer Zero Footprint / ZFP-02041830 (01)00840682102988(10)6.0SP72093796001D  UJJF01-Centricity Universal Viewer Zero Footprint / ZFP-01461578 (01)00840682102988(10)6.0SP72093796001D  600791-Centricity Universal Viewer Zero Footprint / ZFP-02099050 (01)00840682102988(10)6.0SP9.0.12093796001D  HC0021-Centricity Universal Viewer Zero Footprint / ZFP-00007551 (01)00840682102988(10)6.0  HC0727-Centricity Universal Viewer Zero Footprint / ZFP-00111408 (01)00840682102988(10)6.0SP7.0.22093796001D  HC0727-Centricity Universal Viewer Zero Footprint / ZFP-00111408 (01)00840682102988(10)6.0SP7.0.22093796001D  UHAM02-Centricity Universal Viewer Zero Footprint / ZFP-00059329 (01)00840682102988(10)6.0SP7.0.22093796001D  UHAM02-Centricity Universal Viewer Zero Footprint / ZFP-00059329 (01)00840682102988(10)6.0  362520-Centricity Universal Viewer Zero Footprint / ZFP-00393290 (01)00840682102988(10)6.0SP7.0.22093796001D  HHEI01-Centricity Universal Viewer Zero Footprint / ZFP-00545566 (01)00840682102988(10)6.0SP7.0.22093796001D  HC9020-Centricity Universal Viewer Zero Footprint / ZFP-00158636 (01)00840682102988(10)6.0  SiteID-Centricity Universal Viewer Zero Footprint / ZFP-01296263 (01)00840682102988(10)6.0SP7.0.22093796001D  SiteID-Centricity Universal Viewer Zero Footprint / ZFP-01296263 (01)00840682102988(10)6.0  624751-Centricity Universal Viewer Zero Footprint / ZFP-01265404 (01)00840682102988(10)6.0SP8.0.12093796001D  560496-Centricity Universal Viewer Zero Footprint / ZFP-00157858 (01)00840682102988(10)6.0SP7.0.22093796001D  560496-Centricity Universal Viewer Zero Footprint / ZFP-00157858 (01)00840682102988(10)6.0SP9.0.12093796001D  560496-Centricity Universal Viewer Zero Footprint / ZFP-00157858 (01)00840682102988(10)6.0SP92093796001D  145774-Centricity Universal Viewer Zero Footprint / ZFP-01814045 (01)00840682102988(10)6.0SP72093796001D  145774-Centricity Universal Viewer Zero Footprint / ZFP-01814045 (01)00840682102988(10)6.0SP72093796001D  145774-Centricity Universal Viewer Zero Footprint / ZFP-01814045 (01)00840682102988(10)6.0SP72093796001D  145774-Centricity Universal Viewer Zero Footprint / ZFP-01814045 (01)00840682102988(10)6.0SP72093796001D  145774-Centricity Universal Viewer Zero Footprint / ZFP-01814045 (01)00840682102988(10)6.0SP72093796001D  HNUE04-Centricity Universal Viewer Zero Footprint / ZFP-01825781 (01)00840682102988(10)6.0SP7.0.22093796001D  HC0094-Centricity Universal Viewer Zero Footprint / ZFP-00111203 (01)00840682102988(10)6.0SP7.0.22093796001D  PC0070-Centricity Universal Viewer Zero Footprint / ZFP-02042361 (01)00840682102988(10)6.0SP72093796001D  434487-Centricity Universal Viewer Zero Footprint / ZFP-00260666 (01)00840682102988(10)6.0SP7.0.22093796001D  434487-Centricity Universal Viewer Zero Footprint / ZFP-00260666 (01)00840682102988(10)6.0SP7.0.22093796001D  434487-Centricity Universal Viewer Zero Footprint / ZFP-00260666 (01)00840682102988(10)6.0SP7.0.22093796001D  609334-Centricity Universal Viewer Zero Footprint / ZFP-01282257 (01)00840682102988(10)6.0SP72093796001D  609334-Centricity Universal Viewer Zero Footprint / ZFP-01282257 (01)00840682102988(10)6.0SP7.0.22093796001D  HC0623-Centricity Universal Viewer Zero Footprint / ZFP-01282262 (01)00840682102988(10)6.0SP7.0.22093796001D  UC4372-Centricity Universal Viewer Zero Footprint / ZFP-01299640 (01)00840682102988(10)6.0SP7.0.22093796001D  HC4357-Centricity Universal Viewer Zero Footprint / ZFP-02093420 (01)00840682102988(10)6.0SP72093796001D  Entitlement ID c5ef-076c-246d-4e17-a0d1-638d-c807-7ac1 (01)00840682102988(10)6.0SP9.0.12093796001D  TP0575-Centricity Universal Viewer Zero Footprint / ZFP-01987866 (01)00840682102988(10)6.0SP72093796001D  HC1282-Centricity Universal Viewer Zero Footprint / ZFP-00001140 (01)00840682102988(10)6.0SP7.0.22093796001D  GH1022-Centricity Universal Viewer Zero Footprint / ZFP-01971012 (01)00840682102988(10)6.0SP72093796001D  VEDANTAMRF-Centricity Universal Viewer Zero Footprint / ZFP-01973242 (01)00840682102988(10)6.0SP82093796001D  MAX-DC-Centricity Universal Viewer Zero Footprint / ZFP-01892432 (01)00840682102988(10)6.0SP8022093796001D  NIMHANS-Centricity Universal Viewer Zero Footprint / ZFP-01890910 (01)00840682102988(10)6.0SP7.0.22093796001D  NUECLEAR-Centricity Universal Viewer Zero Footprint / ZFP-02003124 (01)00840682102988(10)6.0SP82093796001D  AIIMS-CN-Centricity Universal Viewer Zero Footprint / ZFP-01892392 (01)00840682102988(10)6.0SP7.0.22093796001D  CMC-Centricity Universal Viewer Zero Footprint / ZFP-01820241 (01)00840682102988(10)6.0SP7.0.22093796001D  IL1059-Centricity Universal Viewer Zero Footprint / ZFP-01446762 (01)00840682102988(10)6.0SP7.0.22093796001D  IL1028-Centricity Universal Viewer Zero Footprint / ZFP-00436752 (01)00840682102988(10)6.0SP82093796001D  IL1484-Centricity Universal Viewer Zero Footprint / ZFP-00040948 (01)00840682102988(10)6.0SP7.0.12093796001D  IL1469-Centricity Universal Viewer Zero Footprint / ZFP-01451992 (01)00840682102988(10)6.0SP7.0.12093796001D  IL1469-Centricity Universal Viewer Zero Footprint / ZFP-02156638 (01)00840682102988(10)6.0SP8.0.12093796001D  IL1098-Centricity Universal Viewer Zero Footprint / ZFP-01451983 (01)00840682102988(10)6.0SP7.0.22093796001D  B51673-Centricity Universal Viewer Zero Footprint / ZFP-01423797 (01)00840682102988(10)6.0SP7.0.22093796001D  B53686-Centricity Universal Viewer Zero Footprint / ZFP-01271448 (01)00840682102988(10)6.0SP7.0.22093796001D  SiteID-Centricity Universal Viewer Zero Footprint / ZFP-01452221 (01)00840682102988(10)6.0SP7.0.22093796001D  B53818-Centricity Universal Viewer Zero Footprint / ZFP-01804985 (01)00840682102988(10)6.0SP72093796001D  A52511-Centricity Universal Viewer Zero Footprint / ZFP-01817696 (01)00840682102988(10)6.0SP72093796001D  A3431341-Centricity Universal Viewer Zero Footprint / ZFP-01980190 (01)00840682102988(10)6.0SP7.0.22093796001D  A1892038-Centricity Universal Viewer Zero Footprint / ZFP-02093306 (01)00840682102988(10)6.0SP72093796001D  SiteID-Centricity Universal Viewer Zero Footprint / ZFP-0",
      "center_classification_date": "20181219",
      "distribution_pattern": "Worldwide",
      "state": "WI",
      "product_description": "Centricity Universal Viewer Zero Footprint Client, Model Number 2090258-001, Software Versions 6.0 SP7, SP7.0.1, SP7.0.2, SP7.0.3, SP8, SP8.0.1, SP8.0.a and SP8.0.2, SP9 and SP9.0.1 display o orientation markers can become is incorrect.",
      "report_date": "20181226",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "GE Healthcare, LLC",
      "recall_number": "Z-0641-2019",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "81635",
      "more_code_info": "1452217 (01)00840682102988(10)6.0SP7.0.22093796001D  SiteID-Centricity Universal Viewer Zero Footprint / ZFP-01452219 (01)00840682102988(10)6.0SP7.0.22093796001D  A57219-Centricity Universal Viewer Zero Footprint / ZFP-01270577 (01)00840682102988(10)6.0SP7.0.22093796001D  A57219-Centricity Universal Viewer Zero Footprint / ZFP-01270577 (01)00840682102988(10)6.0SP7.0.22093796001D  A76809-Centricity Universal Viewer Zero Footprint / ZFP-02040801 (01)00840682102988(10)6.0SP72093796001D  Entitlement ID e13c-05cf-1fe2-4855-adeb-f5cb-36fd-359c (01)00840682102988(10)6.0SP82093796001D  A54039-Centricity Universal Viewer Zero Footprint / ZFP-01452222 (01)00840682102988(10)6.0SP7.0.22093796001D  Entitlement ID 4ddf-aba0-e5f8-4f45-bb0b-45bc-deff-4a48 (01)00840682102988(10)6.0SP9.0.12093796001D  JP-130527-Centricity Universal Viewer Zero Footprint / ZFP-02093919  (01)00840682102988(10)6.0SP72093796001D  JP-340177-Centricity Universal Viewer Zero Footprint / ZFP-01275424 (01)00840682102988(10)6.0SP7.0.22093796001D  JP-340206-Centricity Universal Viewer Zero Footprint / ZFP-01275380 (01)00840682102988(10)6.0SP7.0.22093796001D  JP-020002-Centricity Universal Viewer Zero Footprint / ZFP-01275422 (01)00840682102988(10)6.0SP8022093796001D  JP-020002-Centricity Universal Viewer Zero Footprint / ZFP-01275423 (01)00840682102988(10)6.0SP8022093796001D  JP-070167-Centricity Universal Viewer Zero Footprint / ZFP-01275409 (01)00840682102988(10)6.0SP7.0.22093796001D  JP-140140-Centricity Universal Viewer Zero Footprint / ZFP-02154910 (01)00840682102988(10)6.0SP9012093796001D  JP-140051-Centricity Universal Viewer Zero Footprint / ZFP-02099295 (01)00840682102988(10)6.0SP9012093796001D  JP-110145-Centricity Universal Viewer Zero Footprint / ZFP-01275393 (01)00840682102988(10)6.0SP7.0.22093796001D  JP-100014-Centricity Universal Viewer Zero Footprint / ZFP-01890709 (01)00840682102988(10)6.0SP72093796001D  JP-240130-Centricity Universal Viewer Zero Footprint / ZFP-02040340 (01)00840682102988(10)6.0SP9012093796001D  JP-150017-Centricity Universal Viewer Zero Footprint / ZFP-02088882 (01)00840682102988(10)6.0SP9012093796001D  JP-272900-Centricity Universal Viewer Zero Footprint / ZFP-02041843 (01)00840682102988(10)6.0SP9012093796001D  JP-110088-Centricity Universal Viewer Zero Footprint / ZFP-01275336 (01)00840682102988(10)6.0SP9012093796001D  JP-110088-Centricity Universal Viewer Zero Footprint / ZFP-01275427 (01)00840682102988(10)6.0SP9012093796001D  JP-290033-Centricity Universal Viewer Zero Footprint / ZFP-01275435 (01)00840682102988(10)6.0SP7.0.22093796001D  JP-290033-Centricity Universal Viewer Zero Footprint / ZFP-01275437 (01)00840682102988(10)6.0SP7.0.22093796001D  JP-290033-Centricity Universal Viewer Zero Footprint / ZFP-01275438 (01)00840682102988(10)6.0SP7.0.22093796001D  JP-130422-Centricity Universal Viewer Zero Footprint / ZFP-01980318 (01)00840682102988(10)6.0SP72093796001D  JP-130440-Centricity Universal Viewer Zero Footprint / ZFP-01979557 (01)00840682102988(10)6.0SP72093796001D  JP-130440-Centricity Universal Viewer Zero Footprint / ZFP-01979560 (01)00840682102988(10)6.0SP72093796001D  JP-130467-Centricity Universal Viewer Zero Footprint / ZFP-01275430 (01)00840682102988(10)6.0SP7.0.22093796001D  JP-130626-Centricity Universal Viewer Zero Footprint / ZFP-02043333 (01)00840682102988(10)6.0SP9012093796001D  JP-160102-Centricity Universal Viewer Zero Footprint / ZFP-01978243 (01)00840682102988(10)6.0SP72093796001D  JP-080215-Centricity Universal Viewer Zero Footprint / ZFP-01825286 (01)00840682102988(10)6.0SP72093796001D  JP-275353-Centricity Universal Viewer Zero Footprint / ZFP-02042636 (01)00840682102988(10)6.0SP8022093796001D  Entitlement ID f3f6-d8c4-4f9e-4052-80d5-ef4b-0812-4d84 (01)00840682102988(10)6.0SP82093796001D  SiteID-Centricity Universal Viewer Zero Footprint / ZFP-02156484 (01)00840682102988(10)6.0SP72093796001D  SiteID-Centricity Universal Viewer Zero Footprint / ZFP-02156484 (01)00840682102988(10)6.0SP72093796001D  SiteID-Centricity Universal Viewer Zero Footprint / ZFP-02156499 (01)00840682102988(10)6.0SP72093796001D  SiteID-Centricity Universal Viewer Zero Footprint / ZFP-02156499 (01)00840682102988(10)6.0SP72093796001D  SiteID-Centricity Universal Viewer Zero Footprint / ZFP-02170617 (01)00840682102988(10)6.0SP72093796001D  RU0472-Centricity Universal Viewer Zero Footprint / ZFP-02092676 (01)00840682102988(10)6.0SP72093796001D  RU1112-Centricity Universal Viewer Zero Footprint / ZFP-01977816 (01)00840682102988(10)6.0SP72093796001D  RU1107-Centricity Universal Viewer Zero Footprint / ZFP-01620728 (01)00840682102988(10)6.0SP72093796001D  tobecreated-Centricity Universal Viewer Zero Footprint / ZFP-01813329 (01)00840682102988(10)6.0SP72093796001D  RU9940-Centricity Universal Viewer Zero Footprint / ZFP-01812753 (01)00840682102988(10)6.0SP72093796001D  SA1039-Centricity Universal Viewer Zero Footprint / ZFP-00065810 (01)00840682102988(10)6.0  SA1005-Centricity Universal Viewer Zero Footprint / ZFP-00065849 (01)00840682102988(10)6.0SP7.0.22093796001D  SA1209-Centricity Universal Viewer Zero Footprint / ZFP-00066545 (01)00840682102988(10)6.0SP7.0.22093796001D  SA1209-Centricity Universal Viewer Zero Footprint / ZFP-00066545 (01)00840682102988(10)6.0  SA 1054-Centricity Universal Viewer Zero Footprint / ZFP-00065863 (01)00840682102988(10)6.0  SAU1433-Centricity Universal Viewer Zero Footprint / ZFP-01457278 (01)00840682102988(10)6.0SP7.0.22093796001D  SA1001-Centricity Universal Viewer Zero Footprint / ZFP-01891616 (01)00840682102988(10)6.0SP72093796001D  SA2110-Centricity Universal Viewer Zero Footprint / ZFP-01805103 (01)00840682102988(10)6.0SP72093796001D  SA1008-Centricity Universal Viewer Zero Footprint / ZFP-00602579 (01)00840682102988(10)6.0SP7.0.22093796001D  NNI-Centricity Universal Viewer Zero Footprint / ZFP-02041820 (01)00840682102988(10)6.0SP82093796001D  KTPH-Centricity Universal Viewer Zero Footprint / ZFP-01452994 (01)00840682102988(10)6.0SP72093796001D  SiteID-Centricity Universal Viewer Zero Footprint / ZFP-01476439 (01)00840682102988(10)6.0SP7.0.22093796001D  TTSH-RAD-Centricity Universal Viewer Zero Footprint / ZFP-01977001 (01)00840682102988(10)6.0SP82093796001D  TTSH-CVIS-Centricity Universal Viewer Zero Footprint / ZFP-02041523 (01)00840682102988(10)6.0SP72093796001D  ZA2103-Centricity Universal Viewer Zero Footprint / ZFP-00117174 (01)00840682102988(10)6.0  ZA2541-Centricity Universal Viewer Zero Footprint / ZFP-01979759 (01)00840682102988(10)6.0  ZA2241-Centricity Universal Viewer Zero Footprint / ZFP-00111314 (01)00840682102988(10)6.0  ZA2474-Centricity Universal Viewer Zero Footprint / ZFP-01818412 (01)00840682102988(10)6.0SP7.0.22093796001D  ZA2343-Centricity Universal Viewer Zero Footprint / ZFP-01891392 (01)00840682102988(10)6.0SP7.0.22093796001D  SiteID-Centricity Universal Viewer Zero Footprint / ZFP-01641033 (01)00840682102988(10)6.0  MPX431-Centricity Universal Viewer Zero Footprint / ZFP-01813942 (01)00840682102988(10)6.0  CPZ550-Centricity Universal Viewer Zero Footprint / ZFP-02042135 (01)00840682102988(10)6.0SP92093796001D  GPX021-Centricity Universal Viewer Zero Footprint / ZFP-02041108 (01)00840682102988(10)6.0SP72093796001D  GPX021-Centricity Universal Viewer Zero Footprint / ZFP-02041108 (01)00840682102988(10)6.0SP72093796001D  GPX459-Centricity Universal Viewer Zero Footprint / ZFP-01980453 (01)00840682102988(10)6.0SP72093796001D  GPX459-Centricity Universal Viewer Zero Footprint / ZFP-01980458 (01)00840682102988(10)6.0SP72093796001D  GPX459-Centricity Universal Viewer Zero Footprint / ZFP-01980459 (01)00840682102988(10)6.0SP72093796001D  GPX459-Centricity Universal Viewer Zero Footprint / ZFP-01980460 (01)00840682102988(10)6.0SP72093796001D  GPX459-Centricity Universal Viewer Zero Footprint / ZFP-01980471 (01)00840682102988(10)6.0SP72093796001D  GPX459-Centricity Universal Viewer Zero Footprint / ZFP-01980471 (01)00840682102988(10)6.0SP82093796001D  GPX459-Centricity Universal Viewer Zero Footprint / ZFP-01980472 (01)00840682102988(10)6.0SP72093796001D  GPX459-Centricity Universal Viewer Zero Footprint / ZFP-01980473 (01)00840682102988(10)6.0SP72093796001D  GPX458-Centricity Universal Viewer Zero Footprint / ZFP-01980391 (01)00840682102988(10)6.0SP72093796001D  GPX458-Centricity Universal Viewer Zero Footprint / ZFP-01980391 (01)00840682102988(10)6.0SP72093796001D  GPX458-Centricity Universal Viewer Zero Footprint / ZFP-01980392 (01)00840682102988(10)6.0SP72093796001D  GPX458-Centricity Universal Viewer Zero Footprint / ZFP-01980393 (01)00840682102988(10)6.0SP72093796001D  GPX458-Centricity Universal Viewer Zero Footprint / ZFP-01980394 (01)00840682102988(10)6.0SP72093796001D  GPX458-Centricity Universal Viewer Zero Footprint / ZFP-01980399 (01)00840682102988(10)6.0SP72093796001D  GPX458-Centricity Universal Viewer Zero Footprint / ZFP-01980400 (01)00840682102988(10)6.0SP72093796001D  GPX458-Centricity Universal Viewer Zero Footprint / ZFP-01980401 (01)00840682102988(10)6.0SP72093796001D  GPX458-Centricity Universal Viewer Zero Footprint / ZFP-01980412 (01)00840682102988(10)6.0SP72093796001D  GPX458-Centricity Universal Viewer Zero Footprint / ZFP-01980413 (01)00840682102988(10)6.0SP72093796001D  GPX458-Centricity Universal Viewer Zero Footprint / ZFP-01980414 (01)00840682102988(10)6.0SP72093796001D  GPX458-Centricity Universal Viewer Zero Footprint / ZFP-01980436 (01)00840682102988(10)6.0SP72093796001D  GPX458-Centricity Universal Viewer Zero Footprint / ZFP-01980437 (01)00840682102988(10)6.0SP72093796001D  GPX458-Centricity Universal Viewer Zero Footprint / ZFP-01980438 (01)00840682102988(10)6.0SP72093796001D  GPX458-Centricity Universal Viewer Zero Footprint / ZFP-01980439 (01)00840682102988(10)6.0SP72093796001D  GPX458-Centricity Universal Viewer Zero Footprint / ZFP-01980767 (01)00840682102988(10)6.0SP72093796001D  GPX458-Centricity Universal Viewer Zero Footprint / ZFP-01980772 (01)00840682102988(10)6.0SP72093796001D  GPX458-Centricity Universal Viewer Zero Footprint / ZFP-01980773 (01)00840682102988(10)6.0SP72093796001D  GPX458-Centricity Universal Viewer Zero Footprint / ZFP-01980774 (01)00840682102988(10)6.0SP72093796001D  GPX458-Centricity Universal Viewer Zero Footprint / ZFP-01980785 (01)00840682102988(10)6.0SP72093796001D  GPX458-Centricity Universal Viewer Zero Footprint / ZFP-01980786 (01)00840682102988(10)6.0SP72093796001D  GPX458-Centricity Universal Viewer Zero Footprint / ZFP-01980787 (01)00840682102988(10)6.0SP72093796001D  GPX458-Centricity Universal Viewer Zero Footprint / ZFP-01980808 (01)00840682102988(10)6.0SP72093796001D  GPX458-Centricity Universal Viewer Zero Footprint / ZFP-01980809 (01)00840682102988(10)6.0SP72093796001D  GPX458-Centricity Universal Viewer Zero Footprint / ZFP-01980810 (01)00840682102988(10)6.0SP72093796001D  GPX458-Centricity Universal Viewer Zero Footprint / ZFP-01980811 (01)00840682102988(10)6.0SP72093796001D  GEC218-Centricity Universal Viewer Zero Footprint / ZFP-01457080 (01)00840682102988(10)6.0SP72093796001D  GPX459-Centricity Universal Viewer Zero Footprint / ZFP-01980471 (01)00840682102988(10)6.0SP92093796001D  DEN272-Centricity Universal Viewer Zero Footprint / ZFP-01370312 (01)00840682102988(10)6.0SP72093796001D  DPN729-Centricity Universal Viewer Zero Footprint / ZFP-01983752 (01)00840682102988(10)6.0SP72093796001D  DPN729-Centricity Universal Viewer Zero Footprint / ZFP-01983752 (01)00840682102988(10)6.0SP72093796001D  DPN672-Centricity Universal Viewer Zero Footprint / ZFP-01637671 (01)00840682102988(10)6.0SP72093796001D  GPX461-Centricity Universal Viewer Zero Footprint / ZFP-01980671 (01)00840682102988(10)6.0SP72093796001D  GPX461-Centricity Universal Viewer Zero Footprint / ZFP-01980674 (01)00840682102988(10)6.0SP72093796001D  GPX461-Centricity Universal Viewer Zero Footprint / ZFP-01980678 (01)00840682102988(10)6.0SP72093796001D  GPX461-Centricity Universal Viewer Zero Footprint / ZFP-01980680 (01)00840682102988(10)6.0SP72093796001D  GPX461-Centricity Universal Viewer Zero Footprint / ZFP-01980681 (01)00840682102988(10)6.0SP72093796001D  GPX461-Centricity Universal Viewer Zero Footprint / ZFP-01980685 (01)00840682102988(10)6.0SP72093796001D  GPX461-Centricity Universal Viewer Zero Footprint / ZFP-01980686 (01)00840682102988(10)6.0SP72093796001D  GPX461-Centricity Universal Viewer Zero Footprint / ZFP-01980687 (01)00840682102988(10)6.0SP72093796001D  GPX461-Centricity Universal Viewer Zero Footprint / ZFP-01980707 (01)00840682102988(10)6.0SP72093796001D  GPX461-Centricity Universal Viewer Zero Footprint / ZFP-01980708 (01)00840682102988(10)6.0SP72093796001D  GPX461-Centricity Universal Viewer Zero Footprint / ZFP-01980709 (01)00840682102988(10)6.0SP72093796001D  GPX461-Centricity Universal Viewer Zero Footprint / ZFP-01980730 (01)00840682102988(10)6.0SP72093796001D  GPX461-Centricity Universal Viewer Zero Footprint / ZFP-01980731 (01)00840682102988(10)6.0SP72093796001D  GPX461-Centricity Universal Viewer Zero Footprint / ZFP-01980732 (01)00840682102988(10)6.0SP72093796001D  GPX461-Centricity Universal Viewer Zero Footprint / ZFP-01980743 (01)00840682102988(10)6.0SP72093796001D  GPX461-Centricity Universal Viewer Zero Footprint / ZFP-01980744 (01)00840682102988(10)6.0SP72093796001D  GPX461-Centricity Universal Viewer Zero Footprint / ZFP-01980745 (01)00840682102988(10)6.0SP72093796001D  GPX461-Centricity Universal Viewer Zero Footprint / ZFP-01980746 (01)00840682102988(10)6.0SP72093796001D  GPX461-Centricity Universal Viewer Zero Footprint / ZFP-01980758 (01)00840682102988(10)6.0SP72093796001D  GPX461-Centricity Universal Viewer Zero Footprint / ZFP-01980761 (01)00840682102988(10)6.0SP72093796001D  GPX461-Centricity Universal Viewer Zero Footprint / ZFP-01980762 (01)00840682102988(10)6.0SP72093796001D  GPX461-Centricity Universal Viewer Zero Footprint / ZFP-01980763 (01)00840682102988(10)6.0SP72093796001D  UP285A-Centricity Universal Viewer Zero Footprint / ZFP-01815172 (01)00840682102988(10)6.0  GPX460-Centricity Universal Viewer Zero Footprint / ZFP-01980524 (01)00840682102988(10)6.0SP72093796001D  GPX460-Centricity Universal Viewer Zero Footprint / ZFP-01980529 (01)00840682102988(10)6.0SP72093796001D  GPX460-Centricity Universal Viewer Zero Footprint / ZFP-01980530 (01)00840682102988(10)6.0SP72093796001D  GPX460-Centricity Universal Viewer Zero Footprint / ZFP-01980531 (01)00840682102988(10)6.0SP72093796001D  GPX460-Centricity Universal Viewer Zero Footprint / ZFP-01980543 (01)00840682102988(10)6.0SP72093796001D  GPX460-Centricity Universal Viewer Zero Footprint / ZFP-01980544 (01)00840682102988(10)6.0SP72093796001D  GPX460-Centricity Universal Viewer Zero Footprint / ZFP-01980545 (01)00840682102988(10)6.0SP72093796001D  GPX460-Centricity Universal Viewer Zero Footprint / ZFP-01980566 (01)00840682102988(10)6.0SP72093796001D  GPX460-Centricity Universal Viewer Zero Footprint / ZFP-01980567 (01)00840682102988(10)6.0SP72093796001D  GPX460-Centricity Universal Viewer Zero Footprint / ZFP-01980568 (01)00840682102988(10)6.0SP72093796001D  GPX460-Centricity Universal Viewer Zero Footprint / ZFP-01980569 (01)00840682102988(10)6.0SP72093796001D  NE2196-Centricity Universal Viewer Zero Footprint / ZFP-00038588 (01)00840682102988(10)6.0SP72093796001D  CPZ411-Centricity Universal Viewer Zero Footprint / ZFP-02042142 (01)00840682102988(10)6.0SP92093796001D  CPZ528-Centricity Universal Viewer Zero Footprint / ZFP-02042132 (01)00840682102988(10)6.0SP92093796001D  CPZ529-Centricity Universal Viewer Zero Footprint / ZFP-02042139 (01)00840682102988(10)6.0SP92093796001D  NEV440-Centricity Universal Viewer Zero Footprint / ZFP-00169834 (01)00840682102988(10)6.0SP9.0.12093796001D  O051CA01-Centricity Universal Viewer Zero Footprint / ZFP-01810930 (01)00840682102988(10)6.0SP72093796001D  O051CA01-Centricity Universal Viewer Zero Footprint / ZFP-01810930 (01)00840682102988(10)6.0SP72093796001D  O051CA01-Centricity Universal Viewer Zero Footprint / ZFP-01810930 (01)00840682102988(10)6.0SP72093796001D  O051CA01-Centricity Universal Viewer Zero Footprint / ZFP-01810930 (01)00840682102988(10)6.0SP7.0.22093796001D  O051CA01-Centricity Universal Viewer Zero Footprint / ZFP-01810930 (01)00840682102988(10)6.0SP92093796001D  SiteID-Centricity Universal Viewer Zero Footprint / ZFP-01270584 (01)00840682102988(10)6.0  529010-Centricity Universal Viewer Zero Footprint / ZFP-01824677 (01)00840682102988(10)6.0SP72093796001D  529010-Centricity Universal Viewer Zero Footprint / ZFP-01824677 (01)00840682102988(10)6.0SP72093796001D  529010-Centricity Universal Viewer Zero Footprint / ZFP-01978642 (01)00840682102988(10)6.0SP72093796001D  529010-Centricity Universal Viewer Zero Footprint / ZFP-01980592 (01)00840682102988(10)6.0SP72093796001D  Entitlement ID 845e-83ae-a73c-450f-89e2-0112-6c23-e63e (01)00840682102988(10)6.0SP9.0.12093796001D  437918-Centricity Universal Viewer Zero Footprint / ZFP-00163452 (01)00840682102988(10)6.0SP7.0.22093796001D  899010-Centricity Universal Viewer Zero Footprint / ZFP-00163425 (01)00840682102988(10)6.0  885020-Centricity Universal Viewer Zero Footprint / ZFP-00007543 (01)00840682102988(10)6.0SP7.0.22093796001D  529060-Centricity Universal Viewer Zero Footprint / ZFP-00389240 (01)00840682102988(10)6.0SP7.0.22093796001D  475290-Centricity Universal Viewer Zero Footprint / ZFP-00110992 (01)00840682102988(10)6.0SP7.0.22093796001D  482896-Centricity Universal Viewer Zero Footprint / ZFP-00160200 (01)00840682102988(10)6.0SP7.0.12093796001D  725030-Centricity Universal Viewer Zero Footprint / ZFP-00545600 (01)00840682102988(10)6.0SP7.0.22093796001D  730140-Centricity Universal Viewer Zero Footprint / ZFP-00038193 (01)00840682102988(10)6.0SP7.0.12093796001D  150080-Centricity Universal Viewer Zero Footprint / ZFP-01299636 (01)00840682102988(10)6.0SP7.0.22093796001D  0853550429-Centricity Universal Viewer Zero Footprint / ZFP-02041927 (01)00840682102988(10)6.0SP72093796001D  0853550675-Centricity Universal Viewer Zero Footprint / ZFP-01287925 (01)00840682102988(10)6.0SP7.0.22093796001D  0853550804-Centricity Universal Viewer Zero Footprint / ZFP-01286189 (01)00840682102988(10)6.0SP7.0.22093796001D  0853550993-Centricity Universal Viewer Zero Footprint / ZFP-02093449 (01)00840682102988(10)6.0SP7.0.12093796001D  0853550185-Centricity Universal Viewer Zero Footprint / ZFP-02093535 (01)00840682102988(10)6.0SP7.0.22093796001D  06162-Centricity Universal Viewer Zero Footprint / ZFP-01270588 (01)00840682102988(10)6.0SP7.0.32093796001D  06051-Centricity Universal Viewer Zero Footprint / ZFP-00116181 (01)00840682102988(10)6.0SP7.0.22093796001D  16003-Centricity Universal Viewer Zero Footprint / ZFP-01824845 (01)00840682102988(10)6.0SP7.0.22093796001D  34340-Centricity Universal Viewer Zero Footprint / ZFP-01976214 (01)00840682102988(10)6.0SP72093796001D  SiteID-Centricity Universal Viewer Zero Footprint / ZFP-01816213 (01)00840682102988(10)6.0SP72093796001D  81001-Centricity Universal Viewer Zero Footprint / ZFP-00002610 (01)00840682102988(10)6.0SP7.0.22093796001D  80257.-Centricity Universal Viewer Zero Footprint / ZFP-00007681 (01)00840682102988(10)6.0SP7.0.22093796001D  RA901-Centricity Universal Viewer Zero Footprint / ZFP-01819168 (01)00840682102988(10)6.0SP7.0.22093796001D  RVVKC-Centricity Universal Viewer Zero Footprint / ZFP-01811138 (01)00840682102988(10)6.0SP7.0.22093796001D  RWF03-Centricity Universal Viewer Zero Footprint / ZFP-01453951 (01)00840682102988(10)6.0SP7.0.22093796001D  RVV09-Centricity Universal Viewer Zero Footprint / ZFP-00228542 (01)00840682102988(10)6.0SP7.0.22093796001D  81564-Centricity Universal Viewer Zero Footprint / ZFP-00436804 (01)00840682102988(10)6.0SP72093796001D  RRK02-Centricity Universal Viewer Zero Footprint / ZFP-01270599 (01)00840682102988(10)6.0  85059-Centricity Universal Viewer Zero Footprint / ZFP-01270589 (01)00840682102988(10)6.0SP7.0.22093796001D  60211-Centricity Universal Viewer Zero Footprint / ZFP-01987822 (01)00840682102988(10)6.0SP72093796001D  60211-Centricity Universal Viewer Zero Footprint / ZFP-01987822 (01)00840682102988(10)6.0SP72093796001D  85435-Centricity Universal Viewer Zero Footprint / ZFP-01270590 (01)00840682102988(10)6.0SP7.0.22093796001D  PRDC0-Centricity Universal Viewer Zero Footprint / ZFP-00163568 (01)00840682102988(10)6.0SP8.0.12093796001D  RPA02-Centricity Universal Viewer Zero Footprint / ZFP-01453953 (01)00840682102988(10)6.0SP7.0.22093796001D  Entitlement ID a6d0-bae8-15be-41cc-b098-7f30-8068-c73b (01)00840682102988(10)6.0SP9.0.12093796001D  RXQ50-Centricity Universal Viewer Zero Footprint / ZFP-00505094 (01)00840682102988(10)6.0SP7.0.22093796001D  RXQ50-Centricity Universal Viewer Zero Footprint / ZFP-01701456 (01)00840682102988(10)6.0SP72093796001D  RXQ50-Centricity Universal Viewer Zero Footprint / ZFP-00505094 (01)00840682102988(10)6.0SP9.0.12093796001D  RWEAA-Centricity Universal Viewer Zero Footprint / ZFP-00165095 (01)00840682102988(10)6.0SP9.0.12093796001D  Entitlement ID 6aa9-ddd2-7438-447a-ad6a-38b6-da28-9b6c (01)00840682102988(10)6.0SP9.0.12093796001D  80009-Centricity Universal Viewer Zero Footprint / ZFP-01270597 (01)00840682102988(10)6.0SP9.0.12093796001D  RRJ05-Centricity Universal Viewer Zero Footprint / ZFP-00006675 (01)00840682102988(10)6.0SP72093796001D  RRJ05-Centricity Universal Viewer Zero Footprint / ZFP-00006675 (01)00840682102988(10)6.0SP72093796001D  80674-Centricity Universal Viewer Zero Footprint / ZFP-01425411 (01)00840682102988(10)6.0SP82093796001D  80674-Centricity Universal Viewer Zero Footprint / ZFP-01425411 (01)00840682102988(10)6.0SP82093796001D  80674-Centricity Universal Viewer Zero Footprint / ZFP-01425411 (01)00840682102988(10)6.0SP9.0.12093796001D  DDC01-Centricity Universal Viewer Zero Footprint / ZFP-00162205 (01)00840682102988(10)6.0SP7.0.22093796001D  DDC01-Centricity Universal Viewer Zero Footprint / ZFP-00162205 (01)00840682102988(10)6.0SP9.0.12093796001D",
      "recall_initiation_date": "20181109",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Ongoing"
    }
  ]
}