  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Norwood",
      "address_1": "2 Edgewater Dr",
      "reason_for_recall": "When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25¿C (77¿F).",
      "address_2": "",
      "product_quantity": "4642",
      "code_info": "Carton/Bottle Lot Number:  C0099091C  P0099091C  C0100101F  P0100101F  C9101022I                       P9101022I  C0103042D  P0103042D  C0106062E  P0106062E",
      "center_classification_date": "20130104",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and internationally to Canada,  Aruba, Brazil, Denmark,  India, Mexico, New Zealand, Singapore,Thailand, Bahamas, South Wales, and Hong Kong,",
      "state": "MA",
      "product_description": "Siemens Chek-Stix Combo Pak  (Distributed OUS)  Catalog Number: 1364E    Chek-Stix¿ Positive and Negative Control Strips for Urine Chemistry: When used with Siemens reagent strips and tablets for urinalysis, CHEK-STIX¿ positive and negative control strips provide defined results for glucose, bilirubin, ketone (acetoacetic acid), specific gravity, blood, pH, protein, urobilinogin, nitrite, and leukocytes",
      "report_date": "20130116",
      "classification": "Class III",
      "openfda": {},
      "recalling_firm": "Siemens Healthcare Diagnostics Inc",
      "recall_number": "Z-0641-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "63809",
      "termination_date": "20141106",
      "more_code_info": "",
      "recall_initiation_date": "20121121",
      "postal_code": "02062-4637",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}