  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Suzhou",
      "state": "",
      "country": "China",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "93574",
      "recalling_firm": "Philips Healthcare (Suzhou) Co., Ltd.",
      "address_1": "Suzhou Industrial Park",
      "address_2": "No. 258 Zhongyuan Road; Pingjiang",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, ID, IL, IN, KS, KY, LA, MD, ME, MI, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, PA, RI, SC, TN, TX, UT, VT, WI, & Puerto Rico and the countries of Albania, Algeria, Argentina, Australia, Austria, Bhutan, Bolivia, Bosnia-Herz, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Ethiopia, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Laos, Latvia, Lebanon, Lithuania, Malaysia, Mali, Mauritius, Moldova, Morocco, Netherlands, New Zealand, Norway, Pakistan, Palestine, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russia, Rwanda, Saint Kitts, Saint Lucia, Saudi Arabia, Senegal, Serbia, Sint Maarten, Slovakia, South Africa, South Korea, Spain, St. Pier Miquel, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Kingdom, United Arab Emirates, Uzbekistan, Venezuela, & Vietnam.",
      "recall_number": "Z-0640-2024",
      "product_description": "Incisive CT, model 728143 & 728144 running Software Version 5.0.0.",
      "product_quantity": "828",
      "reason_for_recall": "Multiple software issues which have the potential to lead to misdiagnosis due to image artifacts or incorrect image orientation labels, or the need for a CT rescan.     These include: 1. Artifacts may occur on perfusion images, 2. Orientation label reversed in 2D images, 3. Auto voice does not work during the cardiac scan, 4. Auto voice and operator\u0019s voice are too quiet, patient can\u0019t hear, 5. Not all of the images are reconstructed in some cases, 6. No image displayed on the console monitor after enabling CCT Single Scan, 7. Ring artifact appears in 32x0.625 collimation of bone scan when soft tissue view is selected, 8. A design defect where the error couch motion signal of couch firmware causes couch to identify current position as target position in next protocol, and 9. calcium scoring study split into separate series after loading to ISP.",
      "recall_initiation_date": "20221223",
      "center_classification_date": "20240105",
      "report_date": "20240117",
      "code_info": "UDI-DI: 00884838085015 (Model 728143) UDI-DI: 00884838105508 (Model 728144)",
      "more_code_info": ""
    }
  ]
}