  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Norwood",
      "address_1": "2 Technology Way",
      "reason_for_recall": "RBC Reagent failure. High fluorescent counts during RBC Reagent + Diluent QC check that will result in failed QC check, preventing user from testing patient samples for RBC.",
      "address_2": "",
      "product_quantity": "",
      "code_info": "Part number 111011, Lot # R1513",
      "center_classification_date": "20181219",
      "distribution_pattern": "US Nationwide distribution in the states of IN, WA, NC, CA, GA, CO.",
      "state": "MA",
      "product_description": "GloCyte RBC Reagent - Product Usage: The GloCyte Automated Cell Counter System is intended for use by trained healthcare professionals in clinical laboratories to provide quantitative determination of fluorescence labeled total nucleated cells and erythrocytes in cerebrospinal fluid collected from adult and pediatric patients. The GloCyte Low and High Level Controls are assayed hematology controls designed to monitor the performance of the GloCyte Automated Cell Counter System. Assayed parameters include total nucleated cells and erythrocytes.",
      "report_date": "20181226",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Advanced Instruments, LLC",
      "recall_number": "Z-0640-2019",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "79584",
      "termination_date": "20190924",
      "more_code_info": "",
      "recall_initiation_date": "20180207",
      "postal_code": "02062-2633",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}