  "results": [
    {
      "status": "Terminated",
      "city": "San Jose",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69838",
      "recalling_firm": "Stryker Endoscopy",
      "address_1": "5900 Optical Ct",
      "address_2": "N/A",
      "postal_code": "95138-1400",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.",
      "recall_number": "Z-0639-2015",
      "product_description": "PKG, PKG, HOOK SCISSORS, P/N 0250080260.   Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.",
      "product_quantity": "180,573 units total",
      "reason_for_recall": "The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods",
      "recall_initiation_date": "20141117",
      "center_classification_date": "20141217",
      "termination_date": "20150717",
      "report_date": "20141224",
      "code_info": "41490  44630  140583  142467  142469  143688  144270  246031  440310  443401  443405  443406  443407  542490  542498  542499  641965  641966  641968  645668  741519  741520  741521  741523  744992  744994  744995  747000  747001  747002  841991  842244  842246  842248  842259  842263  842267  842271  842876  941705  942762  942846  943012  943044  943107  943247  943417  943747  944089  944704  944838  944885  945128  945554  945555  945557  945558  945559  945562  945564  945565  945568  945569  1042239  1042474  1044885  1046102  1046734  1046735  1140692  1141608  1142179  1143096  1143097  1144503  1145104  1145105  1146377  1146378  1147161  1147162  1241404  1242145  1242261  1243810  1244133  1244439  1245093  1245235  1245831  1246051  1246185  1340227  1340932  1340933  1341825  1342200  1342201  1342202  1343515  1343516  1345798  1440785  1441076  1442552"
    }
  ]
}