  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Center Valley",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89404",
      "recalling_firm": "Olympus Corporation of the Americas",
      "address_1": "3500 Corporate Pkwy",
      "address_2": "N/A",
      "postal_code": "18034-8229",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide",
      "recall_number": "Z-0638-2022",
      "product_description": "OER-Elite Endoscope Reprocessor-for use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes, their accessories, and endoscope reprocessor accessories",
      "product_quantity": "295 units",
      "reason_for_recall": "The detergent and alcohol tank tubing may have been incorrectly assembled in the OER-Elite. If the tubing is incorrectly assembled, alcohol will be dispensed  instead of detergent for the cleaning cycle and detergent will be dispensed instead of alcohol for the  alcohol cycle, resulting in a ineffective endoscope  reprocessing and patient exposure to residual detergent",
      "recall_initiation_date": "20211213",
      "center_classification_date": "20220211",
      "termination_date": "20231226",
      "report_date": "20220223",
      "code_info": "All serial numbers. UDI: 04953170404047"
    }
  ]
}