      "country": "United States",
      "city": "Norwood",
      "address_1": "2 Edgewater Dr",
      "reason_for_recall": "When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25¿C (77¿F).",
      "address_2": "",
      "product_quantity": "16711",
      "code_info": "Carton/Bottle Lot Number:  C9093061C  P9093061C  C9094081J  P9094081J  C0100101D  P0100101D  C0097091E  P0097091E  C9101022E  P9101022E  C0099091D  P0099091D  C0100101E  P0100101E  C9101022H  P9101022H  C9101022J  P9101022J  C0103042B  P0103042B  C0103042G  P0103042G  C0106062D  P0106062D  C9105052C  P9105052C",
      "center_classification_date": "20130104",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and internationally to Canada,  Aruba, Brazil, Denmark,  India, Mexico, New Zealand, Singapore,Thailand, Bahamas, South Wales, and Hong Kong,",
      "state": "MA",
      "product_description": "Siemens Chek-Stix Combo Pak   Catalog Number: 1364    Chek-Stix¿ Positive and Negative Control Strips for Urine Chemistry: When used with Siemens reagent strips and tablets for urinalysis, CHEK-STIX¿ positive and negative control strips provide defined results for glucose, bilirubin, ketone (acetoacetic acid), specific gravity, blood, pH, protein, urobilinogin, nitrite, and leukocytes",
      "report_date": "20130116",
      "classification": "Class III",
      "openfda": {},
      "recalling_firm": "Siemens Healthcare Diagnostics Inc",
      "recall_number": "Z-0637-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "63809",
      "termination_date": "20141106",
      "more_code_info": "",
      "recall_initiation_date": "20121121",
      "postal_code": "02062-4637",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}