{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Torrance",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86737",
      "recalling_firm": "Kowa Optimed Inc",
      "address_1": "20001 S Vermont Ave",
      "address_2": "N/A",
      "postal_code": "90502-1326",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution including in the states of NY, WA, TX, CT, MA, ID, IL, MI.",
      "recall_number": "Z-0636-2021",
      "product_description": "Kowa VK-2s software Version 10.0.4.0, Product code 04987646809500 installed on Kowa nonmyd WX Retinal Cameras - Product Usage: is indicated for acquiring, displaying and storing image data from KOWA s mydriatic and non-mydriatic ophthalmic cameras.",
      "product_quantity": "N/A",
      "reason_for_recall": "It has been determined that the software distributed with the retinal camera may delete patient information. Specifically, if multiple IDs are delated at the same time, their is a potential for all patient information to be deleted. This could result in delay of treatment.",
      "recall_initiation_date": "20201103",
      "center_classification_date": "20201217",
      "termination_date": "20221206",
      "report_date": "20201223",
      "code_info": "Serial Numbers: 32168900007, 32168900008, 32168900009, 32168900010, 32168900011,  32168900019, 32168900020, 32168900021, 32168900022, 32168900025,  32168900035, 32168900036, 32168900037, 32168900044, 32168900045,  32169000052, 32169000144, 32169000145, 32169000152, 32169000153,  32169000166, 32169000167, 32169000168"
    }
  ]
}