  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waltham",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72808",
      "recalling_firm": "Perkinelmer",
      "address_1": "940 Winter Street",
      "address_2": "N/A",
      "postal_code": "02451",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide and the countries of   ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BOLIVIA, BRAZIL, CANADA, CHINA, DENMARK, FINLAND, FRANCE, GERMANY, GREAT BRITAIN, GREECE, HONG KONG, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, KOREA, KUWAIT, MYANMAR, NETHERLANDS, NEW ZEALAND, NORWAY, POLAND, PORTUGAL, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, and THAILAND.",
      "recall_number": "Z-0636-2016",
      "product_description": "Perkin Elmer WIZARD2 5-detector, 1000 samples  Product Code: 2470-0150, 3470-0150  The Wizard gamma counter is intended to detect and count gamma radiation emitted by clinical samples",
      "product_quantity": "4 units",
      "reason_for_recall": "The content of the second #023 barcode ID label in the Barcode ID Label Binder is incorrect.  The second barcode ID label #023 symbol incorrectly identifies as #024 when scanned. If measurement protocols have been programmed for both ID #023 and #024, the protocol associated with barcode ID #024 is executed.  If the error is undetected, the Gamma Counter may produce erroneous results.",
      "recall_initiation_date": "20151208",
      "center_classification_date": "20160112",
      "termination_date": "20170804",
      "report_date": "20160120",
      "code_info": "Serial Numbers:  DG04129084  DG12118776  DG04117773  DG05106408"
    }
  ]
}