  "results": [
    {
      "country": "United States",
      "city": "Norwood",
      "address_1": "2 Edgewater Dr",
      "reason_for_recall": "When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25¿C (77¿F).",
      "address_2": "",
      "product_quantity": "6466",
      "code_info": "Carton/Bottle Lot Number:  S9093061A    P9093061A   S9094081A    P9094081A   S0097091C    P0097091C   S9093061A     P9093061A  S9094081A     P9094081A    S0097091C    P0097091C   S0100101B    P0100101B    S0103042A    P0103042A    S0103042H    P0103042H    S0106062A    P0106062A      S9101022A    P9101022A    S9101022B    P9101022B   S9105052A    P9105052A",
      "center_classification_date": "20130104",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and internationally to Canada,  Aruba, Brazil, Denmark,  India, Mexico, New Zealand, Singapore,Thailand, Bahamas, South Wales, and Hong Kong,",
      "state": "MA",
      "product_description": "Siemens CLINITEK Atlas Positive and Negative Control Strips for Urine  Chemistry  Catalog Number: 5019    When used with the CLINITEK ATLAS¿ Automated urine Chemistry Analyzer and CLINITEK ATLAS¿ 10 reagent Pak, CLINITEK ATLAS¿ Positive and Negative Control Strips provide defined results for color, clarity, glucose, bilirubin, ketone (acetoacetic acid), specific gravity, blood, pH, protein, urobilinogen, nitrite, and leukocytes. Each CLINITEK ATLAS Positive Control Strip is a firm plastic strip to which are affixed seven separate analyte areas.",
      "report_date": "20130116",
      "classification": "Class III",
      "openfda": {},
      "recalling_firm": "Siemens Healthcare Diagnostics Inc",
      "recall_number": "Z-0636-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "63809",
      "termination_date": "20141106",
      "more_code_info": "",
      "recall_initiation_date": "20121121",
      "postal_code": "02062-4637",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}