  "results": [
    {
      "status": "Terminated",
      "city": "Waltham",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72808",
      "recalling_firm": "Perkinelmer",
      "address_1": "940 Winter Street",
      "address_2": "N/A",
      "postal_code": "02451",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide and the countries of   ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BOLIVIA, BRAZIL, CANADA, CHINA, DENMARK, FINLAND, FRANCE, GERMANY, GREAT BRITAIN, GREECE, HONG KONG, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, KOREA, KUWAIT, MYANMAR, NETHERLANDS, NEW ZEALAND, NORWAY, POLAND, PORTUGAL, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, and THAILAND.",
      "recall_number": "Z-0635-2016",
      "product_description": "Perkin Elmer WIZARD2 10-detector, 550 samples Product Code: 2470-0100, 3470-0100.  The Wizard gamma counter is intended to detect and count gamma radiation emitted by clinical samples",
      "product_quantity": "47 units",
      "reason_for_recall": "The content of the second #023 barcode ID label in the Barcode ID Label Binder is incorrect.  The second barcode ID label #023 symbol incorrectly identifies as #024 when scanned. If measurement protocols have been programmed for both ID #023 and #024, the protocol associated with barcode ID #024 is executed.  If the error is undetected, the Gamma Counter may produce erroneous results.",
      "recall_initiation_date": "20151208",
      "center_classification_date": "20160112",
      "termination_date": "20170804",
      "report_date": "20160120",
      "code_info": "Model Number: 2470-0100  Serial Numbers:  DG06106436  DG08118240  DG02106050  DG06106502  DG11118609  DG01128795  DG02117461  DG12118702  DG09106861  DG05117833  DG05117834  DG10107006  DG06106452  DG09118367  DG03106154  DG04129081  DG06106559  DG06117903  DG07106599  DG09106774  DG05106387  DG06106524  DG10106949  DG09118284  DG05117787  DG04117755  DG09106770  DG03106196  DG04129159  DG08118192  DG06106469  DG08118189  DG04129075  DG12107252  DG12118677  DG06117923  DG08106741  DG03106186  DG12118759  DG01117444  DG02106112  DG12107236  DG11107031  DG12107249  DG11118612  DG05117781  Model: 3470-0100 (1 unit)  Serial Number: DG02128911"
    }
  ]
}