  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Norwood",
      "address_1": "2 Edgewater Dr",
      "reason_for_recall": "When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25¿C (77¿F).",
      "address_2": "",
      "product_quantity": "7865",
      "code_info": "Carton/Bottle Lot Number:    C9093061B  P9093061B  C9094081B  P9094081B  C0097091B  P0097091B  C0100101C  P0100101C  C0103042F  P0103042F  C0106062B  P0106062B  C91 01022D  P91 01022D  CP9105052B  P9105052B  C9108092A  P9108092A",
      "center_classification_date": "20130104",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and internationally to Canada,  Aruba, Brazil, Denmark,  India, Mexico, New Zealand, Singapore,Thailand, Bahamas, South Wales, and Hong Kong,",
      "state": "MA",
      "product_description": "Siemens Chek-Stix Positive and Negative Control Strips for Urine Chemistry  Catalog Number:1360    Chek-Stix¿ Positive and Negative Control Strips for Urine Chemistry: When used with Siemens reagent strips and tablets for urinalysis, CHEK-STIX¿ positive and negative control strips provide defined results for glucose, bilirubin, ketone (acetoacetic acid), specific gravity, blood, pH, protein, urobilinogin, nitrite, and leukocytes",
      "report_date": "20130116",
      "classification": "Class III",
      "openfda": {},
      "recalling_firm": "Siemens Healthcare Diagnostics Inc",
      "recall_number": "Z-0635-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "63809",
      "termination_date": "20141106",
      "more_code_info": "",
      "recall_initiation_date": "20121121",
      "postal_code": "02062-4637",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}