      "status": "Terminated",
      "city": "Waltham",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72808",
      "recalling_firm": "Perkinelmer",
      "address_1": "940 Winter Street",
      "address_2": "N/A",
      "postal_code": "02451",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide and the countries of   ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BOLIVIA, BRAZIL, CANADA, CHINA, DENMARK, FINLAND, FRANCE, GERMANY, GREAT BRITAIN, GREECE, HONG KONG, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, KOREA, KUWAIT, MYANMAR, NETHERLANDS, NEW ZEALAND, NORWAY, POLAND, PORTUGAL, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, and THAILAND.",
      "recall_number": "Z-0634-2016",
      "product_description": "Perkin Elmer ClWIZARD2 5-detector, 550 samples  Product Code: 2470-0050, 3470-0050.  The Wizard2 gamma counter is intended to detect and count gamma radiation emitted by clinical samples.",
      "product_quantity": "62 units",
      "reason_for_recall": "The content of the second #023 barcode ID label in the Barcode ID Label Binder is incorrect.  The second barcode ID label #023 symbol incorrectly identifies as #024 when scanned. If measurement protocols have been programmed for both ID #023 and #024, the protocol associated with barcode ID #024 is executed.  If the error is undetected, the Gamma Counter may produce erroneous results.",
      "recall_initiation_date": "20151208",
      "center_classification_date": "20160112",
      "termination_date": "20170804",
      "report_date": "20160120",
      "code_info": "Model: 2470-0050  Serial Numbers:  DG05106329  DG02117493  DG11107166  DG09106782  DG07118133  DG07118141  DG09106791  DG04117727  DG06118036  DG12107192  DG02106101  DG04106223  DG01128843  DG12107272  DG07118124  DG09106892  DG12118684  DG12118673  DG11107126  DG04117704  DG04117763  DG01117363  DG06106484  DG12107240  DG01128871  DG05117880  DG12118685  DG09118357  DG03106204  DG05106386  DG11107137  DG03106172  DG03106206  DG02106113  DG09106818  DG02106065  DG04106309  DG09106882  DG11118590  DG06106483  DG01117316  DG08118221  DG06117898  DG12107263  DG08118250  DG12118711  DG04117710  DG06106448  DG06106453  DG06106463  DG06117912  DG03106135  DG01117373  DG05117837  DG06106517  DG01128805  DG06117970  DG04106310  DG02117521  DG06106538  Model: 3470-0050 (2 units)  Serial Numbers:  DG12118660  DG10118464"
    }
  ]
}