{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukegan",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81579",
      "recalling_firm": "Cardinal Health 200, LLC",
      "address_1": "3651 Birchwood Dr",
      "address_2": "",
      "postal_code": "60085-8337",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (Nationwide distribution) to states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and to countries of: Austria, Belgium, Canada, Chile, Costa Rica, France, Germany, Italy, Luxembourg, Martinique, Monaco, Netherlands, Poland, Reunion, Spain, Sweden, Switzerland, Taiwan, and United Kingdom.",
      "recall_number": "Z-0633-2019",
      "product_description": "Protexis Neoprene Surgical Glove, Size 6.5    This powder-free Surgeon s glovessterile light brown colored surgeons gloves are a disposable device made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.",
      "product_quantity": "1,475,095 pairs total",
      "reason_for_recall": "Labeling issue in that the breakthrough (i.e., permeation) times for two chemotherapy drugs (Carmustine and Thiotepa) stated on the  wallet  (primary packaging) of the Protexis Neoprene surgical gloves are incorrect.  Additionally, breakthrough times for Mitomycin were added to the same label in error.",
      "recall_initiation_date": "20181031",
      "center_classification_date": "20181218",
      "termination_date": "20250812",
      "report_date": "20181226",
      "code_info": "Product code: 2D73DP65, Lot Numbers: TS18050409, TS18060149, TS18060240, TS18060312, TS18070036, TS18070106, TS18070306, TS18070371, TS18080026, TS18080348, TS18080445, TS18090113, TS18090159, TS18090212, and TS18090282.",
      "more_code_info": ""
    }
  ]
}