  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waltham",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72808",
      "recalling_firm": "Perkinelmer",
      "address_1": "940 Winter Street",
      "address_2": "N/A",
      "postal_code": "02451",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide and the countries of   ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BOLIVIA, BRAZIL, CANADA, CHINA, DENMARK, FINLAND, FRANCE, GERMANY, GREAT BRITAIN, GREECE, HONG KONG, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, KOREA, KUWAIT, MYANMAR, NETHERLANDS, NEW ZEALAND, NORWAY, POLAND, PORTUGAL, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, and THAILAND.",
      "recall_number": "Z-0633-2016",
      "product_description": "Perkin Elmer WIZARD2 2-detector, 550 samples Product Code: 2470-0020  The Wizard gamma counter is intended to detect and count gamma radiation emitted by clinical samples",
      "product_quantity": "60 units",
      "reason_for_recall": "The content of the second #023 barcode ID label in the Barcode ID Label Binder is incorrect.  The second barcode ID label #023 symbol incorrectly identifies as #024 when scanned. If measurement protocols have been programmed for both ID #023 and #024, the protocol associated with barcode ID #024 is executed.  If the error is undetected, the Gamma Counter may produce erroneous results.",
      "recall_initiation_date": "20151208",
      "center_classification_date": "20160112",
      "termination_date": "20170804",
      "report_date": "20160120",
      "code_info": "Serial Numbers:  DG02117524  DG09118363  DG08118156  DG08118195  DG02128922  DG10118375  DG12107235  DG08118222  DG02117537  DG08118184  DG09106872  DG11107116  DG03117626  DG03117611  DG01117425  DG12107293  DG09118332  DG06106503  DG06118006  DG01128820  DG12118707  DG05106427  DG03106166  DG11118630  DG01117434  DG09106779  DG02117500  DG12118742  DG06106477  DG03106208  DG09106800  DG10106903  DG05117853  DG03117671  DG06106473  DG04117754  DG04117698  DG11118538  DG11118617  DG01117392  DG03117644  DG05117882  DG03117655  DG04117683  DG02106106  DG05106347  DG06118043  DG10118385  DG08106671  DG01128869  DG06118023  DG05106328  DG02117516  DG11107142  DG05106354  DG08118254  DG03106181  DG03106197  DG02117505  DG11107105"
    }
  ]
}