  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waltham",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72808",
      "recalling_firm": "Perkinelmer",
      "address_1": "940 Winter Street",
      "address_2": "N/A",
      "postal_code": "02451",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide and the countries of   ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BOLIVIA, BRAZIL, CANADA, CHINA, DENMARK, FINLAND, FRANCE, GERMANY, GREAT BRITAIN, GREECE, HONG KONG, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, KOREA, KUWAIT, MYANMAR, NETHERLANDS, NEW ZEALAND, NORWAY, POLAND, PORTUGAL, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, and THAILAND.",
      "recall_number": "Z-0632-2016",
      "product_description": "Perkin Elmer WIZARD2 1-detector, 550 samples Product Code: 2470-0010  The Wizard gamma counter is intended to detect and count gamma radiation emitted by clinical samples",
      "product_quantity": "47 units",
      "reason_for_recall": "The content of the second #023 barcode ID label in the Barcode ID Label Binder is incorrect.  The second barcode ID label #023 symbol incorrectly identifies as #024 when scanned. If measurement protocols have been programmed for both ID #023 and #024, the protocol associated with barcode ID #024 is executed.  If the error is undetected, the Gamma Counter may produce erroneous results.",
      "recall_initiation_date": "20151208",
      "center_classification_date": "20160112",
      "termination_date": "20170804",
      "report_date": "20160120",
      "code_info": "Serial Numbers:  DG06106539  DG12118720  DG07106646  DG12107250  DG12107281  DG03117619  DG11118605  DG01117400  DG12107202  DG01117401  DG06106449  DG03117666  DG05117791  DG06106454  DG03106180  DG11118637  DG07118121  DG07106598  DG11118633  DG06106520  DG03106137  DG07118111  DG02128888  DG08118247  DG06118014  DG01128790  DG12107207  DG07106637  DG01128876  DG04117764  DG01106035  DG12118780  DG05106342  DG01117399  DG06117909  DG06118011  DG09106846  DG11107146  DG12118694  DG08118217  DG02106084  DG04117690  DG05117838  DG09106799  DG04117735  DG03117548  DG01117362"
    }
  ]
}